Highly sensitive electrochemical sensor for tyrosine detection using a sub-millimeter electrode

被引:37
作者
Deng, Peihong [1 ]
Xiao, Jingyun [2 ]
Feng, Jinxia [2 ]
Tian, Yaling [2 ]
Wu, Yiyong [2 ]
Li, Junhua [1 ]
He, Quanguo [2 ]
机构
[1] Hengyang Normal Univ, Key Lab Funct Organometall Mat Hunan Prov Univ, Dept Chem & Mat Sci, Key Lab Funct Met Organ Cpds Hunan Prov, Hengyang 421008, Peoples R China
[2] Hunan Univ Technol, Sch Life Sci & Chem, Zhuzhou 412007, Peoples R China
基金
中国国家自然科学基金;
关键词
Acetylene black; Submillimeter electrode; Electrochemical oxidation; Tyrosine sensor; Reduced graphene oxide; BLACK PASTE ELECTRODE; SELECTIVE DETERMINATION; ACETYLENE BLACK; GRAPHENE OXIDE; VOLTAMMETRIC DETERMINATION; MASS-SPECTROMETRY; CARBON NANOTUBES; ASCORBIC-ACID; L-TRYPTOPHAN; FILM;
D O I
10.1016/j.microc.2021.106106
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In recent years, small-sized electrodes have drawn widespread attraction in electroanalysis due to high sensitivity, rapid response and low cost. Hereby, a sub-millimeter paste electrode with inner diameter of 0.8 mm was prepared by combining the excellent properties of reduced graphene oxide (rGO) and acetylene black (AB), and the electrochemical behavior of tyrosine (Tyr) on the modified electrode (denoted as rGO/ABPE) was investigated in detail. The electrochemical responses obtained on the sub-millimeter rGO/ABPE were compared with that obtained on a conventional size electrode with diameter of 3 mm. The results showed that the 0.8 mm one has better performance than that of 3 mm. Compared with the paste electrode made of graphite powder, the oxidation peak current of Tyr increased significantly due to the synergistic effect of rGO and AB. Under the optimized experimental conditions, the calibration curve of Tyr showed a linear relationship between 0.1 and 100 ?M (R = 0.9995) with good sensitivity of 3.314 ?A ?M-1 cm-2, and the detection limit was 0.06 ?M. By comparison with conventional size electrode, the detection of Tyr on the sub-millimeter electrode shows higher sensitivity, wider linear range and lower detection limit. In addition, the excellent reproducibility, stability and selectivity of the sub-millimeter rGO/ABPE make it suitable for Tyr analysis in both food and biological samples.
引用
收藏
页数:11
相关论文
共 46 条
  • [41] Construction of effective electrochemical sensor for the determination of quinoline yellow based on different morphologies of manganese dioxide functionalized graphene
    Wu, Yiyong
    Deng, Peihong
    Tian, Yaling
    Magesa, Felista
    Liu, Jun
    Li, Guangli
    He, Quanguo
    [J]. JOURNAL OF FOOD COMPOSITION AND ANALYSIS, 2019, 84
  • [42] Rapid recognition and determination of tryptophan by carbon nanotubes and molecularly imprinted polymer-modified glassy carbon electrode
    Wu, Yiyong
    Deng, Peihong
    Tian, Yaling
    Ding, Ziyu
    Li, Guangli
    Liu, Jun
    Zuberi, Zavuga
    He, Quanguo
    [J]. BIOELECTROCHEMISTRY, 2020, 131
  • [43] Electrocatalytic oxidation and direct determination of L-tyrosine by square wave voltammetry at multi-wall carbon nanotubes modified glassy carbon electrodes
    Xu, Q
    Wang, SF
    [J]. MICROCHIMICA ACTA, 2005, 151 (1-2) : 47 - 52
  • [44] A sensitive molecular imprinted electrochemical sensor based on gold nanoparticles decorated graphene oxide: Application to selective determination of tyrosine in milk
    Yola, Mehmet Lutfi
    Eren, Tanju
    Atar, Necip
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2015, 210 : 149 - 157
  • [45] Application of L-DOPA modified carbon nanotubes as a bifunctional electrocatalyst for simultaneous determination of ascorbic acid, adrenaline, acetaminophen and tyrosine
    Zare, Hamid R.
    Moradiyan, Behnaz
    Shekari, Zahra
    Benvidi, Ali
    [J]. MEASUREMENT, 2016, 90 : 510 - 518
  • [46] Determination of L-tyrosine by β-cyclodextrin sensitized fluorescence quenching method
    Zhu, Xiashi
    Xu, Suqin
    [J]. SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2010, 77 (03) : 566 - 571