Combined motor and somatosensory evoked potentials for intraoperative monitoring: intra- and postoperative data in a series of 69 operations

被引:65
作者
Weinzierl, M. R.
Reinacher, P.
Gilsbach, J. M.
Rohde, V.
机构
[1] Rhein Westfal TH Aachen, Dept Neurosurg, D-52057 Aachen, Germany
[2] Univ Aachen, Dept Neurosurg, D-5100 Aachen, Germany
[3] Univ Gottingen, Dept Neurosurg, D-3400 Gottingen, Germany
关键词
motor evoked potentials; somatosensory evoked potentials; intraoperative monitoring;
D O I
10.1007/s10143-006-0061-5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The primary objective of neurophysiologic monitoring during surgery is to avoid permanent neurological injury resulting from surgical manipulation. To prevent motor deficits, either somatosensory (SSEP) or transcranial motor evoked potentials (MEP) are applied. This prospective study was conducted to evaluate if the combined use of SSEP and MEP might be beneficial. Combined SSEP/MEP monitoring was attempted in 100 consecutive procedures, including intracranial and spinal operations. Repetitive transcranial electric motor cortex stimulation was used to elicit MEP from muscles of the upper and lower limb. Stimulation of the tibial and median nerves was performed to record SSEP. Critical SSEP/MEP changes were defined as decreases in amplitude of more than 50% or increases in latency of more than 10% of baseline values. The operation was paused or the surgical strategy was modified in every case of SSEP/MEP changes. Combined SSEP/MEP monitoring was possible in 69 out of 100 operations. In 49 of the 69 operations (71%), SSEP/ MEP were stable, and the patients remained neurologically intact. Critical SSEP/ MEP changes were seen in six operations. Critical MEP changes with stable SSEP occurred in 12 operations. Overall, critical MEP changes were recorded in 18 operations (26%). In 12 of the 18 operations, MEP recovered to some extent after modification of the surgical strategy, and the patients either showed no (n =10) or only a transient motor deficit (n= 2). In the remaining six operations, MEP did not recover and the patients either had a transient (n =3) or a permanent (n = 3) motor deficit. Critical SSEP changes with stable MEP were observed in two operations; both patients did not show a new motor deficit. Our data again confirm that MEP monitoring is superior to SSEP monitoring in detecting impending impairment of the functional integrity of cerebral and spinal cord motor pathways during surgery. Detection of MEP changes and adjustment of the surgical strategy might allow to prevent irreversible pyramidal tract damage. Stable SSEP/MEP recordings reassure the surgeon that motor function is still intact and surgery can be continued safely. The combined SSEP/ MEP monitoring becomes advantageous, if one modality is not recordable.
引用
收藏
页码:109 / 116
页数:8
相关论文
共 27 条
[1]  
BROWNING JL, 1992, ANESTH ANALG, V74, P643
[2]   Somatosensory evoked potential phase reversal and direct motor cortex stimulation during surgery in and around the central region [J].
Cedzich, C ;
Taniguchi, M ;
Schafer, S ;
Schramm, J .
NEUROSURGERY, 1996, 38 (05) :962-970
[3]  
Deletis V, 2001, ANN NY ACAD SCI, V939, P137
[4]   Transcranial electrical motor evoked potential monitoring for brain tumor resection [J].
Deletis, V ;
Camargo, AB .
NEUROSURGERY, 2001, 49 (06) :1488-1489
[5]  
Deletis V, 1993, Adv Neurol, V63, P201
[6]   MONITORING OF SOMATOSENSORY EVOKED-POTENTIALS DURING SURGERY FOR MIDDLE CEREBRAL-ARTERY ANEURYSMS [J].
FRIEDMAN, WA ;
CHADWICK, GM ;
VERHOEVEN, FJS ;
MAHLA, M ;
DAY, AL .
NEUROSURGERY, 1991, 29 (01) :83-88
[7]   POSTOPERATIVE PARAPLEGIA WITH PRESERVED INTRAOPERATIVE SOMATOSENSORY EVOKED-POTENTIALS [J].
GINSBURG, HH ;
SHETTER, AG ;
RAUDZENS, PA .
JOURNAL OF NEUROSURGERY, 1985, 63 (02) :296-300
[8]   SOMATOSENSORY EVOKED-POTENTIALS [J].
GUGINO, V ;
CHABOT, RJ .
INTERNATIONAL ANESTHESIOLOGY CLINICS, 1990, 28 (03) :154-164
[9]   Subcortical strokes from intracranial aneurysm surgery: Implications for intraoperative neuromonitoring [J].
Holland, NR .
JOURNAL OF CLINICAL NEUROPHYSIOLOGY, 1998, 15 (05) :439-446
[10]   Monitoring of intraoperative motor evoked potentials to increase the safety of surgery in and around the motor cortex [J].
Kombos, T ;
Suess, O ;
Ciklatekerlio, Ö ;
Brock, M .
JOURNAL OF NEUROSURGERY, 2001, 95 (04) :608-614