Influence of Pixel Size on Quantification of Airway Wall Thickness in Computed Tomography

被引:11
作者
Achenbach, Tobias [1 ]
Weinheimer, Oliver [1 ]
Dueber, Christoph [1 ]
Heussel, Claus Peter [2 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Diagnost Intervent Radiol, D-55101 Mainz, Germany
[2] Univ Heidelberg Hosp, Thoraxklin, Heidelberg, Germany
关键词
computed tomography; airway; quantification; MULTIDETECTOR ROW CT; FIELD-OF-VIEW; IMAGE-RECONSTRUCTION;
D O I
10.1097/RCT.0b013e318190699a
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The purpose of this study was to determine the point where a further decrease in voxel size does not result in better automatic quantification of the bronchial wall thickness by using 2 different assessment techniques. Materials and Methods: The results from the commonly used full-width-at-half-maximum (FWHM) principle and a new technique (integral-based method [IBM]) were compared for thin-section multi-detector computed tomography (MDCT) data sets from an airway phantom containing 10 different tubular airway phantoms and in a human subsegmental bronchus in vivo. Correlation with the actual wall thickness and comparison of the wall thicknesses assessed for different voxel sizes were performed, and the image resolutions were also compared subjectively. Results: The relative error ranged from 0% (biggest phantom) to 330% (smallest phantom, biggest field of view, smaller matrix, and FWHM). Using IBM, the maximum relative error was 10% in the same setting. For FWHM, the improvement was marginal for most settings with a pixel spacing less than 0.195 x 0.195 x 0.8 mm; however, it still decreases the relative error from 290% to 273.6% for a wall thickness of 0.3 mm and a pixel spacing of 0.076 x 0.076 x 0.8 mm. Conclusions: (1) Using a special technique such as IBM to account for computed tomography's blurring effect in assessing airway wall thickness had the greatest impact on correct quantification. (2),The visual impression and the automatic quantification using the FWHM technique improved marginally by decreasing the voxel size to less than 0.195 x 0.195 x 0.8 mm. (3) The FWHM technique as a model for visual quantification is not reliable for airway wall thicknesses less than 1.5 mm.
引用
收藏
页码:725 / 730
页数:6
相关论文
共 9 条
  • [1] Measurement of thickness and density of thin structures by computed tomography: A simulation study
    Dougherty, G
    Newman, D
    [J]. MEDICAL PHYSICS, 1999, 26 (07) : 1341 - 1348
  • [2] Volumetric measurement of synthetic lung nodules with multi-detector row CT: Effect of various image reconstruction parameters and segmentation thresholds on measurement accuracy
    Goo, JM
    Tongdee, T
    Tongdee, R
    Yeo, K
    Hildebolt, CF
    Bae, KT
    [J]. RADIOLOGY, 2005, 235 (03) : 850 - 856
  • [3] Does two-segment image reconstruction at 64-section CT coronary angiography improve image quality and diagnostic accuracy?
    Herzog, Christopher
    Nguyen, Shaun A.
    Savino, Giancarlo
    Zwerner, Peter L.
    Doll, Josh
    Nielsen, Christopher D.
    Flohr, Thomas G.
    Vogl, Thomas J.
    Costello, Philip
    Schoepf, U. Joseph
    [J]. RADIOLOGY, 2007, 244 (01) : 121 - 129
  • [4] Computer-assisted lung nodule volumetry from multi-detector row CT: Influence of image reconstruction parameters
    Honda, Osamu
    Sumikawa, Hiromitsu
    Johkoh, Takeshi
    Tomiyama, Noriyuki
    Mihara, Naoki
    Inoue, Atsuo
    Tsubamoto, Mitsuko
    Natsag, Javzandulam
    Hamada, Seiki
    Nakamura, Hironobu
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2007, 62 (01) : 106 - 113
  • [5] Coronary artery calcification: Effect of size of field of view on multi-detector row CT measurements
    Hong, C
    Pilgram, TK
    Zhu, F
    Bae, KT
    [J]. RADIOLOGY, 2004, 233 (01) : 281 - 285
  • [6] Influence of a small field-of-view size on the detection of coronary artery calcifications with MSCT:: in vitro and in vivo study
    Mahnken, AH
    Mühlenbruch, G
    Koos, R
    Das, M
    Pohl, S
    Stanzel, S
    Günther, RW
    Wildberger, JE
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (02) : 358 - 364
  • [7] Maximizing quantitative accuracy of lung airway lumen and wall measures obtained from X-ray CT imaging
    Saba, OI
    Hoffman, EA
    Reinhardt, JM
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 2003, 95 (03) : 1063 - 1075
  • [8] Usefulness of morphological characteristics for the differentiation of benign from malignant solitary pulmonary lesions using HRCT
    Seemann, MD
    Staebler, A
    Beinert, T
    Dienemann, H
    Obst, B
    Matzko, M
    Pistitsch, C
    Reiser, MF
    [J]. EUROPEAN RADIOLOGY, 1999, 9 (03) : 409 - 417
  • [9] About objective 3-D analysis of airway geometry in computerized tomography
    Weinheimer, Oliver
    Achenbach, Tobias
    Bletz, Carsten
    Dueber, Christoph
    Kauczor, Hans-Ulrich
    Heussel, Claus Peter
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (01) : 64 - 74