Titanium-alloy enhances bone-pedicle screw fixation:: mechanical and histomorphometrical results of titanium-alloy versus stainless steel

被引:101
作者
Christensen, FB
Dalstra, M
Sejling, F
Overgaard, S
Bünger, C
机构
[1] Aarhus Univ Hosp, Orthopaed Res Lab, DK-8000 Aarhus C, Denmark
[2] Aarhus Univ Hosp, Inst Expt Clin Res, DK-8000 Aarhus, Denmark
关键词
pedicle screw; titanium; stainless steel; pull-out test; torsion; histomorphometry;
D O I
10.1007/s005860050218
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Several types of pedicle screw systems have been utilized to augment lumbar spine fusion. The majority of these systems are made of stainless steel (Ss), but titanium-alloy (Ti-alloy) devices have recently been available on the market. Ti-alloy implants have several potential advantages over Ss ones. High bioactivity and more flexibility may improve bone ingrowth and mechanical fixation, and the material also offers superior magnetic resonance imaging (MRI) and computed tomography (CT) resolution and significantly less signal interference. However, no data are available from loaded spinal constructs regarding bony ingrowth and mechanical fixation. The aim of this study was to analyse the effect of Ti-alloy versus Ss pedicle screws on mechanical fixation and bone ingrowth in a loaded mini-pig model. Eighteen adult mini-pigs underwent total laminectomy and posterolateral spinal fusion at L3-L4, and were randomly selected to receive either Ss (n = 9) or Ti (n = 9) pedicle screw devices. In both groups, the device used was compact Cotrel-Dubousset instrumentation (Sofamore Danek) of an identical size and shape. The postoperative observation time was 3 months. Screws from LJ were used for histomorphometric studies. Mechanical testing (torsional tests and pull-out tests) was performed on the screws from L4. The Ti screws had a higher maximum torque (P < 0.05) and angular stiffness (P < 0.07), measured by torsional testing. In the pull-out tests, no differences were found between the two groups with respect to the maximum load, stiffness and energy to failure. No correlation between removal torque and the pull-out strength was found (r = 0.1). Bone ongrowth on Ti was increased by 33% compared with Ss (P < 0.04), whereas no differences in bone volume around the screws were shown. Mechanical binding at the bone-screw interface was significantly greater for Ti pedicle screws than for Ss, which was explained by the fact that Ti screws had a superior bone ongrowth. There was no correlation between the screw removal torque and the pull-out strength, which indicates that the peripheral bone structure around the screw was unaffected by the choice of metal.
引用
收藏
页码:97 / 103
页数:7
相关论文
共 38 条
[1]   AN ULTRASTRUCTURAL CHARACTERIZATION OF THE INTERFACE BETWEEN BONE AND SPUTTERED TITANIUM OR STAINLESS-STEEL SURFACES [J].
ALBREKTSSON, T ;
HANSSON, HA .
BIOMATERIALS, 1986, 7 (03) :201-205
[2]  
BRADDELEY AJ, 1986, J MICROSCO, V142, P259
[3]  
CARLSON GD, 1992, SPINE S, V17, P196
[4]  
Chang K W, 1988, J Spinal Disord, V1, P257
[5]   THE EFFECTS OF IMPLANT STIFFNESS ON THE BYPASSED BONE-MINERAL DENSITY AND FACET FUSION STIFFNESS OF THE CANINE SPINE [J].
CRAVEN, TG ;
CARSON, WL ;
ASHER, MA ;
ROBINSON, RG .
SPINE, 1994, 19 (15) :1664-1673
[6]   STATIC AND CYCLICAL BIOMECHANICAL ANALYSIS OF PEDICLE SCREW SPINAL CONSTRUCTS [J].
CUNNINGHAM, BW ;
SEFTER, JC ;
SHONO, Y ;
MCAFEE, PC .
SPINE, 1993, 18 (12) :1677-1688
[7]   CORRELATIONS BETWEEN SCREW HOLE PREPARATION, TORQUE OF INSERTION, AND PULLOUT STRENGTH FOR SPINAL SCREWS [J].
DAFTARI, TK ;
HORTON, WC ;
HUTTON, WC .
JOURNAL OF SPINAL DISORDERS, 1994, 7 (02) :139-145
[8]  
EBRAHEIM NA, 1994, J SPINAL DISORD, V7, P478
[9]  
EBRAHEIM NA, 1992, CLIN ORTHOP RELAT R, P133
[10]  
EULENBERGER J, 1990, UNFALLCHIRURG, V93, P96