The DEPFET pixel BIOSCOPE

被引:20
作者
Neeser, W
Böcker, M
Buchholz, P
Fischer, P
Holl, P
Kemmer, J
Klein, P
Koch, H
Löcker, M
Lutz, G
Matthäy, H
Strüder, L
Trimpl, M
Ulrici, J
Wermes, N
机构
[1] Univ Bonn, Inst Phys, D-53115 Bonn, Germany
[2] Univ Dortmund, Expt Phys E Vb, D-44221 Dortmund, Germany
[3] Ketek GMBH, D-85764 Oberschleissheim, Germany
[4] Ruhr Univ Bochum, D-44780 Bochum, Germany
[5] Max Planck Inst Phys, D-80805 Munich, Germany
[6] Max Planck Inst Extraterr Phys, D-85740 Garching, Germany
关键词
D O I
10.1109/23.856581
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The DEPFET pixel Bioscope system based on a DEPFET pixel matrix provides good spatial and energy resolution in real-time digital autoradiography. Due to its very low noise and a thin entrance window the online detection of tritium without vacuum or cooling is demonstrated for the first time. To achieve this milestone a p-channel junction field effect transistor on a fully depleted high ohmic silicon substrate (DEPFET) is used as unit cell for pixel detectors providing an excellent noise performance of 158 eV FWHM at 6 keV (12 e ENC) at room temperature (300 K). 64 x 64 DEPFET pixel matrices with square (50 mu m x 50 mu m) or hexagonal (50 pm x 42 mu m) pixels have been developed and operated successfully. First measurements result in a homogeneous charge collection efficiency on the whole matrix area and a good linearity in the X-ray range from 5 keV to 60 keV. First images taken with the square and the hexagonal pixel matrices using an Fe-55-source as well as a 'proof of principle' experiment for online H-3-detection are presented and discussed.
引用
收藏
页码:1246 / 1250
页数:5
相关论文
共 10 条
[1]   New pixel detector concepts based on junction field effect transistors on high resistivity silicon [J].
Cesura, G ;
Findeis, N ;
Hauff, D ;
Hornel, N ;
Kemmer, J ;
Klein, P ;
Lechner, P ;
Lutz, G ;
Richter, RH ;
Seitz, H .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 377 (2-3) :521-528
[2]  
FISCHER P, 2000, UNPUB NUCL INSTR M A
[3]  
FISCHER P, 1999, UNPUB NUCL INSTR M A
[4]   SEMICONDUCTOR DRIFT CHAMBER - AN APPLICATION OF A NOVEL CHARGE TRANSPORT SCHEME [J].
GATTI, E ;
REHAK, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 225 (03) :608-614
[5]  
HOLL P, 1999, IN PRESS IEEE T NUCL
[6]   NEW DETECTOR CONCEPTS [J].
KEMMER, J ;
LUTZ, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 253 (03) :365-377
[7]   Study of a DEPJFET pixel matrix with continuous clear mechanism [J].
Klein, P ;
Cesura, G ;
Fischer, P ;
Lutz, G ;
Neeser, W ;
Richter, RH ;
Wermes, N .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3) :254-259
[8]  
NEESER W, 1999, IN PRESS NUCL INST A
[9]  
OVERDICK M, 1999, DIGITAL AUTORADIOGRA
[10]   An ISPA camera for beta radiography [J].
Puertolas, D ;
Piedigrossi, D ;
Leutz, H ;
Gys, T ;
DAmbrosio, C .