MS/MS of Ions in a Low Pressure Linear Ion Trap using a Pulsed Gas

被引:26
作者
Collings, Bruce A. [1 ]
Romaschin, Matthew A. [1 ]
机构
[1] MDS Analyt Technol, Concord, ON L4K 4V8, Canada
关键词
OCTOPOLE FIELD; FRAGMENTATION; EXCITATION; EJECTION;
D O I
10.1016/j.jasms.2009.05.005
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A pulsed valve was used to increase the pressure within the trapping region of a low-pressure linear ion trap by situating the pulsed valve close to the ion trapping region. The pressure was estimated to increase from a background pressure of 3.5e-5 Torr of nitrogen to 0.49 mTorr at the center of the trap. The increased pressure allowed excitation periods to be reduced from 100 to 25 ms without suffering losses in MS/MS efficiency during dipolar excitation. The reduction in excitation period translates into an increase in the overall duty cycle of the scan. (J Am Soc Mass Spectrom 2009, 20, 1714-1717) (C) 2009 American Society for Mass Spectrometry
引用
收藏
页码:1714 / 1717
页数:4
相关论文
共 10 条
[1]  
[Anonymous], METHODS OF EXPERIMEN
[2]  
ASHKENAS H, 1966, RAREFIED GAS DYN, V2, P84
[3]   PULSED VALVE ADDITION OF COLLISION AND REAGENT GASES IN FOURIER-TRANSFORM MASS-SPECTROMETRY [J].
CARLIN, TJ ;
FREISER, BS .
ANALYTICAL CHEMISTRY, 1983, 55 (03) :571-574
[4]   COMPETITION BETWEEN RESONANCE EJECTION AND ION DISSOCIATION DURING RESONANT EXCITATION IN A QUADRUPOLE ION-TRAP [J].
CHARLES, MJ ;
MCLUCKEY, SA ;
GLISH, GL .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (12) :1031-1041
[5]   Increased fragmentation efficiency of ions in a low pressure linear ion trap with an added dc octopole field [J].
Collings, BA .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2005, 16 (08) :1342-1352
[6]   Fragmentation of ions in a low pressure linear ion trap [J].
Collings, Bruce A. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (08) :1459-1466
[7]   INTERNAL PULSED VALVE SAMPLE INTRODUCTION ON A QUADRUPOLE ION-TRAP MASS-SPECTROMETER [J].
COOPERSMITH, BI ;
YOST, RA .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1995, 6 (10) :976-980
[8]   Novel Linac II electrode geometry for creating an axial field in a multipole ion guide [J].
Loboda, A ;
Krutchinsky, A ;
Loboda, O ;
McNabb, J ;
Spicer, V ;
Ens, W ;
Standing, K .
EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2000, 6 (06) :531-536
[9]   Mass selective axial ion ejection from a linear quadrupole ion trap [J].
Londry, FA ;
Hager, JW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2003, 14 (10) :1130-1147
[10]   Ion excitation in a linear quadrupole ion trap with an added octopole field [J].
Michaud, AL ;
Frank, AJ ;
Ding, C ;
Zhao, XZ ;
Douglas, DJ .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2005, 16 (06) :835-849