Implementation of a Rapid Whole Blood D-Dimer Test in the Emergency Department of an Urban Academic Medical Center Impact on ED Length of Stay and Ancillary Test Utilization

被引:33
作者
Lee-Lewandrowski, Elizabeth [1 ]
Nichols, John [1 ]
Van Cott, Elizabeth [1 ]
Grisson, Ricky [1 ]
Louissaint, Abner [1 ]
Benzer, Theodore [2 ]
Lewandrowski, Kent [1 ]
机构
[1] Massachusetts Gen Hosp, Dept Pathol, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Dept Emergency Med, Boston, MA 02114 USA
关键词
Rapid whole blood D-dimer test; Emergency department; Urban academic medical center; Length of stay; Ancillary test utilization; Turnaround time; Point-of-care testing; Venous thromboembolism; PULMONARY-EMBOLISM; VENOUS THROMBOSIS; POINT; ASSAY; EXCLUSION; TRIAL;
D O I
10.1309/AJCP6US3ILGEAREE
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Overcrowding and prolonged patient length-of-stay (LOS) in emergency departments (EDs) are growing problems. We evaluated the impact of implementing a rapid whole blood quantitative D-dimer test (Biosite Triage, Biosite Diagnostics, San Diego, CA) in our ED satellite laboratory on 252 patients before vs 211 patients after implementation. All patients also underwent testing with the existing central laboratory method (VIDAS D-dimer, bioMerieux, Durham, NC). D-dimer turnaround time (from blood draw to result) decreased approximately 79% (similar to 2 hours vs 25 minutes). The mean ED LOS declined from 8.46 to 7.14 hours (P = .016). Hospital admissions decreased 13.8%, ED discharges increased 7.3%, and the number of patients admitted for observation increased 6.4% (P = .005). No difference in the utilization of radiologic studies was observed (P = .86). At 3 months' follow-up, none of the after-implementation patients with negative D-dimer results were admitted for subsequent venous thromboembolic disease. The rapid D-dimer test was associated with a shorter ED LOS and fewer hospital admissions.
引用
收藏
页码:326 / 331
页数:6
相关论文
共 17 条
[1]  
Blick KE, 2005, POINT CARE, V4, P11
[2]  
Bradley VM, 2005, NURS ECON, V23, P14
[3]   The accuracy of the enzyme-linked immunosorbent assay D-dimer test in the diagnosis of pulmonary embolism: A meta-analysis [J].
Brown, MD ;
Rowe, BH ;
Reeves, MJ ;
Bermingham, JM ;
Goldhaber, SZ .
ANNALS OF EMERGENCY MEDICINE, 2002, 40 (02) :133-144
[4]  
Dempfle CE, 2001, THROMB HAEMOSTASIS, V85, P671
[5]  
GOLDHABER S, 2005, HARRISONS PRINCIPLES, pCH244
[6]   D-dimer testing for deep venous thrombosis: A metaanalysis [J].
Heim, SW ;
Schectman, JM ;
Siadaty, MS ;
Philbrick, JT .
CLINICAL CHEMISTRY, 2004, 50 (07) :1136-1147
[7]  
Kendall J, 1998, BRIT MED J, V316, P1052
[8]  
LANDRO L, 2001, WALL STREET J 0713, pB1
[9]  
Lee-Lewandrowski E, 2003, ARCH PATHOL LAB MED, V127, P456
[10]   Evaluation of the Biosite Triage (R) Quantitative Whole Blood D-dimer Assay and Comparison With the bioMerieux VIDAS (R) D-dimer Exclusion Test Validation and Utility For Use in the Central Laboratory and at the Point of Care [J].
Lee-Lewandrowski, Elizabeth ;
Van Cott, Elizabeth M. .
POINT OF CARE, 2005, 4 (03) :133-137