Patellofemoral interactions in walking, stair ascent, and stair descent using a virtual patella model

被引:7
作者
Bischoff, Jeffrey E. [1 ]
Hertzler, Justin S. [1 ]
Mason, James J. [1 ]
机构
[1] Zimmer Inc, Warsaw, IN 46581 USA
关键词
Total knee replacement; UHMWPE; Kinematic modeling; Articulation; TOTAL KNEE ARTHROPLASTY; FINITE-ELEMENT-ANALYSIS; TIBIO-FEMORAL CONTACT; CROSS-LINKED UHMWPE; JOINT; TRACKING; POLYETHYLENE; FORCES; REPLACEMENT; STRESSES;
D O I
10.1016/j.jbiomech.2009.04.018
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Restoration of normal patella kinematics is an important clinical outcome of total knee arthroplasty. Failure of the patella within total knee systems has been documented and, upon occurrence, often necessitates revision surgery. It is thus important to understand patella mechanics following implantation, subject to load states that are typically realized during walking and other gaits. Here, a computational model of the patella is developed and used to examine the effects of walking, stair ascent, and stair descent on the development of stress and contact pressure in the patella throughout the gait cycle. Motion of the patella was governed by a combination of kinematic and force control, based on knee flexion and patellofemoral joint reaction force data from the literature. Unlike most previous analyses of full gait, quasi-static equilibrium was enforced throughout the cycle. Results indicate that, though peak forces vary greatly between the three gaits, maximum contact pressure and von Mises stress are roughly equivalent. However, contact area is larger in stair ascent and descent than walking, as patellofemoral loading, implant geometry, and polyethylene yield increase conformity between the femoral component and patella. Additionally, maximum contact pressure does not coincide with maximum load except for the case of walking. Though specific to the implant design considered here, this result has important ramifications for patella testing and emphasizes the need to characterize patella mechanics throughout gait. (c) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1678 / 1684
页数:7
相关论文
共 37 条
[1]   In vitro measurement of the tracking pattern of the human patella [J].
Ahmed, AM ;
Duncan, NA ;
Tanzer, M .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1999, 121 (02) :222-228
[2]   Finite element analysis of human knee joint in varus-valgus [J].
Bendjaballah, MZ ;
ShiraziAdl, A ;
Zukor, DJ .
CLINICAL BIOMECHANICS, 1997, 12 (03) :139-148
[3]   An augmented hybrid constitutive model for simulation of unloading and cyclic loading behavior of conventional and highly crosslinked UHMWPE [J].
Bergström, JS ;
Rimnac, CM ;
Kurtz, SM .
BIOMATERIALS, 2004, 25 (11) :2171-2178
[4]   Impact of time-dependency on long-term material testing and modeling of polyethylene [J].
Bischoff, Jeffrey E. .
MECHANICS OF TIME-DEPENDENT MATERIALS, 2008, 12 (03) :189-203
[5]   Patellofemoral joint stress during stair ascent and descent in persons with and without patellofemoral pain [J].
Brechter, JH ;
Powers, CM .
GAIT & POSTURE, 2002, 16 (02) :115-123
[6]   Comparison of conventional and highly crosslinked UHMWPE patellae evaluated by a new in vitro patellofemoral joint simulator [J].
Burroughs, Brian R. ;
Rubash, Harry E. ;
Estok, Daniel ;
Jasty, Murali ;
Krevolin, Janet ;
Muratoglu, Orhun K. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2006, 79B (02) :268-274
[7]  
D'Lima D D., 2001, Clin Orthop Relat Res, V392, P232
[8]   Impact of patellofemoral design on patellofemoral forces and polyethylene stresses [J].
D'Lima, DD ;
Chen, PC ;
Kester, MA ;
Colwell, CW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A :85-93
[9]   A finite element model of the human knee joint for the study of tibio-femoral contact [J].
Donahue, TLH ;
Hull, ML ;
Rashid, MM ;
Jacobs, CR .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2002, 124 (03) :273-280
[10]   How the stiffness of meniscal attachments and meniscal material properties affect tibio-femoral contact pressure computed using a validated finite element model of the human knee joint [J].
Donahue, TLH ;
Hull, ML ;
Rashid, MM ;
Jacobs, CR .
JOURNAL OF BIOMECHANICS, 2003, 36 (01) :19-34