Development and Validation of a GATE Simulation Model for the LabPET Scanner

被引:14
作者
Rechka, Sanae [1 ]
Fontaine, Rejean [1 ]
Rafecas, Magdalena [2 ]
Lecomte, Roger [3 ]
机构
[1] Univ Sherbrooke, Dept Elect Engn & Comp Engn, Sherbrooke, PQ J1K 2R1, Canada
[2] Univ Valencia, Inst Fis Corpuscular, CSIC, Valencia, Spain
[3] Univ Sherbrooke, Dept Med Nucl & Radiobiol, Sherbrooke, PQ J1K 2R1, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
GATE; LabPET; LYSO/LGSO phoswich; positron emission tomography; simulation; PET; ARCHITECTURE;
D O I
10.1109/TNS.2009.2032091
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A three-dimensional (3D) model of the LabPET scanner has been newly developed by means of GATE (Geant4 application for tomographic emission) with the purpose of carrying out a detailed study of the system's performance. The accuracy of this model was examined by comparing simulated, theoretical and experimental results obtained with LabPET for different performance features including detection efficiency, sensitivity, count rates, photoelectric probability, phoswich detector energy resolution and intrinsic spatial resolution. An agreement of 95% for singles count rates, 96% for detection efficiency and 90% for absolute sensitivity was obtained between simulations and measurements. Also, an agreement of >97% for photoelectric fraction, 96% for radial intrinsic resolution and >93% for axial intrinsic resolution, was obtained between simulations and the theoretical calculations. In conclusion, the achieved results support the accurate modeling of LabPET with GATE.
引用
收藏
页码:3672 / 3679
页数:8
相关论文
共 15 条
  • [1] [Anonymous], P IEEE NUCL SCI S C
  • [2] Performance Evaluation of the LabPET APD-Based Digital PET Scanner
    Bergeron, Melanie
    Cadorette, Jules
    Beaudoin, Jean-Francois
    Lepage, Martin D.
    Robert, Ghislain
    Selivanov, Vitali
    Tetrault, Marc-Andre
    Viscogliosi, Nicolas
    Norenberg, Jeffrey P.
    Fontaine, Rejean
    Lecomte, Roger
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (01) : 10 - 16
  • [3] Cherry S.R., 2012, Physics of nuclear medicine
  • [4] Timing improvement by low-pass filtering and linear interpolation for the LabPET scanner
    Fontaine, Rejean
    Lemieux, Francois
    Viscogliosi, Nicolas
    Tetrault, Marc-Andre
    Bergeron, Melanie
    Riendeau, Joel
    Berard, Philippe
    Cadorette, Jules
    Lecomte, Roger
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (01) : 34 - 39
  • [5] The Hardware and Signal Processing Architecture of LabPET™, a Small Animal APD-Based Digital PET Scanner
    Fontaine, Rejean
    Belanger, Francois
    Viscogliosi, Nicolas
    Semmaoui, Hicham
    Tetrault, Marc-Andre
    Michaud, Jean-Baptiste
    Pepin, Catherine
    Cadorette, Jules
    Lecomte, Roger
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (01) : 3 - 9
  • [6] Counting rates modeling for PET scanners with GATE
    Guez, David
    Bataille, Frederic
    Comtat, Claude
    Honore, Pierre-Francois
    Jan, Sebastien
    Kerhoas, Sophie
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (01) : 516 - 523
  • [7] Modeling electronic processing in GATE
    Kerhoas-Cavata, Sophie
    Guez, David
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02) : 330 - 334
  • [8] Full modelling of the MOSAIC animal PET system based on the GATE Monte Carlo simulation code
    Merheb, C.
    Petegnief, Y.
    Talbot, J. N.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (03) : 563 - 576
  • [9] *OPENGATE COLL, 2007, GATE US GUID VERS 3
  • [10] Physical characterization of the LabPET™ LGSO and LYSO scintillators
    Pepin, C. M.
    St-Pierre, C.
    Forgues, J. -C.
    Kurata, Y.
    Shimura, N.
    Usui, T.
    Takeyama, T.
    Ishibashi, H.
    Lecomte, R.
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2292 - +