Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging

被引:57
作者
Bateman, Timothy M. [1 ,2 ]
Heller, Gary V. [3 ]
McGhie, A. Iain [1 ]
Courter, Staci A. [2 ]
Golub, Robert A. [4 ]
Case, James A. [2 ]
Cullom, S. James [2 ]
机构
[1] Univ Missouri, Mid Amer Heart Inst, Kansas City, MO 64110 USA
[2] Cardiovasc Imaging Technol, Kansas City, MO USA
[3] Hartford Hosp, Hartford, CT 06115 USA
[4] Cardiol Associates Cent Connecticut, Wallingford, CT USA
关键词
Radionuclides; single photon emission computed tomography; myocardial perfusion imaging; attenuation correction; stress testing; coronary artery disease; MYOCARDIAL-PERFUSION SPECT; CORONARY-ARTERY-DISEASE; CARDIOLOGY; NUCLEAR; SCAN; CT;
D O I
10.1007/s12350-009-9109-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
New iterative algorithms for scatter compensation (SC), noise suppression, and depth-dependent collimator resolution (RR) can shorten rest and stress SPECT acquisitions by 50% while maintaining quality and accuracy equivalent to conventional scans. Full-time stress-only myocardial perfusion SPECT is accurate and efficient when combined with line-source attenuation correction (LSAC). We investigated the potential for half-time stress-only LSAC-SPECT by comparing this to conventional rest/stress SPECT in patients imaged for suspected CAD at three different centers. One hundred and ten patients (58% men, 53% exercise) had 64 projection rest/stress Tc-99m ECG-gated SPECT with simultaneous Gd-153 LSAC: 18 had a parts per thousand currency sign5% CAD likelihood and 92 had coronary angiography. The stress scans were retrospectively 'stripped' to create equally spaced 32 projection "half-time" (HT) scans for the emission and transmission (TX) projections. Astonish (Philips, Milpitas, CA) processing with AC, SC, and RR was applied to the HT data with the HT TX maps reconstructed using a Bayesian iterative method. The conventional rest/stress image sets processed using filtered back projection and without AC and the HT-AC stress-only images were interpreted in random sequence by consensus of two readers blinded to clinical information in separate reading sessions. Comparing rest/stress FBP and HT-LSAC, stress perfusion quality was excellent/good in 82 and 89% (P = .13); interpretive certainty (definitely normal or abnormal) was 88 and 95% (P = .14); sensitivity was 77 and 83% (P = .38); specificity was 67 and 71% (P = .65); normalcy was 94 and 100% (P = 1.0); SSS for CAD pts was 7.4 vs 7.8 and for non-CAD pts was 0.7 vs 0 (P = .44 and .16, respectively). Mean stress LVEF was 60% in both groups. Stress-only imaging with HT-LSAC using the Astonish acquisition/processing method provides results equivalent to conventional rest/stress scanning. This new approach has the potential to significantly improve operational efficiency without sacrificing accuracy.
引用
收藏
页码:726 / 735
页数:10
相关论文
共 23 条
[1]   Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies [J].
Borges-Neto, Salvador ;
Pagnanelli, Robert A. ;
Shaw, Linda K. ;
Honeycutt, Emily ;
Shwartz, Shuli C. ;
Adams, George L. ;
Coleman, Ralph Edward .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (04) :555-565
[2]   A Bayesian iterative transmission gradient reconstruction algorithm for cardiac SPECT attenuation correction [J].
Case, James A. ;
Hsu, Bai Ling ;
Bateman, Timothy M. ;
Cullom, S. James .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (03) :324-333
[3]   Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart - A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association [J].
Cerqueira, MD ;
Weissman, NJ ;
Dilsizian, V ;
Jacobs, AK ;
Kaul, S ;
Laskey, WK ;
Pennell, DJ ;
Rumberger, JA ;
Ryan, T ;
Verani, MS .
CIRCULATION, 2002, 105 (04) :539-542
[4]   ANALYSIS OF PROBABILITY AS AN AID IN THE CLINICAL-DIAGNOSIS OF CORONARY-ARTERY DISEASE [J].
DIAMOND, GA ;
FORRESTER, JS .
NEW ENGLAND JOURNAL OF MEDICINE, 1979, 300 (24) :1350-1358
[5]   Radiation dose to patients from cardiac diagnostic imaging [J].
Einstein, Andrew J. ;
Moser, Kevin W. ;
Thompson, Randall C. ;
Cerqueira, Manuel D. ;
Henzlova, Milena J. .
CIRCULATION, 2007, 116 (11) :1290-1305
[6]  
Fricke H, 2004, J NUCL MED, V45, P1619
[7]   Low event rate for stress-only perfusion imaging in patients evaluated for chest pain [J].
Gibson, PB ;
Demus, D ;
Noto, R ;
Hudson, W ;
Johnson, LL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (06) :999-1004
[8]   Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT [J].
Goetze, Sibyll ;
Wahl, Richard L. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (02) :200-206
[9]   Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging [J].
Heller, GV ;
Bateman, TM ;
Johnson, LL ;
Cullom, SJ ;
Case, JA ;
Galt, JR ;
Garcia, EV ;
Haddock, K ;
Moutray, KL ;
Poston, C ;
Botvinick, EH ;
Fish, MB ;
Follansbee, WP ;
Hayes, S ;
Iskandrian, AE ;
Mahmarian, JJ ;
Vandecker, W .
JOURNAL OF NUCLEAR CARDIOLOGY, 2004, 11 (03) :273-281
[10]  
KOKKILARA A, 2007, J NUCL CARDIOL, V14, pS102