Endothelial cell density in donor corneas -: A comparison of automatic software programs with manual counting

被引:20
作者
Hirneiss, Christoph [1 ]
Schumann, Ricarda G. [1 ]
Grueterich, Martin [1 ]
Welge-Luessen, Ulrich C. [1 ]
Kampik, Anselm [1 ]
Neubauer, Aljoscha S. [1 ]
机构
[1] Univ Munich, Dept Ophthalmol, D-80336 Munich, Germany
关键词
endothelial cell density; organ culture; corneal endothelium; automated endothelium counting;
D O I
10.1097/ICO.0b013e31802be629
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Several automated systems have become available to determine endothelial cell density (ECD) of donor corneas. The purpose of this study was the comparison of 2 systems, the Endothelial Analysis System EAT VIA (Rhinetec, Duesseldorf, Germany) and the NAVIS Cell Count Version 3.4.1 (Nidek Technologies, Erlangen, Germany), against manual ECD counts. Methods: During organ culture, the endothelium of 50 human corneas was photographed on both conventional film and digitally. Manual fixed-frame counting was performed on 2 photographs by 3 experienced ophthalmologists. Three digital pictures of each cornea were analyzed semiautomatically on the Rhinetec EAT system V1.4. Images were imported and analyzed by the fully automatic mode of the NAVIS system V3.4.1 (which also offers a semiautomatic counting mode). In an additional experiment on 11 corneas, the NAVIS counts on the microscopy system previously used only for manual cell counts were directly compared with manual counts. Results: Mean endothelial cell density obtained by manual counting of all observers was 2679 +/- 313 (SD) cells/mm(2). Values obtained by digital analysis with EAT were a mean of 1.3% higher (2713 +/- 286 cells/mm2), which is not a significant difference. In contrast, analysis of the digital pictures with the fully automated NAVIS software mode showed a mean of 7.0% higher measurements (2867 +/- 111 cells/mm(2)), a significant difference (P = 0.0001). Altman-Bland analysis revealed that the fully automated NAVIS counts gave too low results for high ECDs and too high results for low ECDs. This result was confirmed by the direct comparison with manual counting. Conclusions: Fully automated analysis of ECD remains problematic. Validation of any computer-assisted system against manual counting is of importance.
引用
收藏
页码:80 / 83
页数:4
相关论文
共 18 条
[1]   Increased endothelial cell density in the paracentral and peripheral regions of the human cornea [J].
Amann, J ;
Holley, GP ;
Lee, SB ;
Edelhauser, HF .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2003, 135 (05) :584-590
[2]   Predicting endothelial cell loss and long-term corneal graft survival [J].
Armitage, WJ ;
Dick, AD ;
Bourne, WM .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (08) :3326-3331
[3]   AUTOMATED DIGITAL IMAGE-ANALYSIS OF ORGAN-CULTURE PRESERVED DONOR CORNEAS [J].
BARISANIASENBAUER, T ;
BAUMGARTNER, I ;
GRABNER, G ;
STUR, M .
OPHTHALMIC RESEARCH, 1993, 25 (02) :94-99
[4]   A NEW SYSTEM FOR STORING DONOR CORNEAS [J].
BOHNKE, M .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1991, 198 (02) :135-137
[5]   Cellular changes in transplanted human corneas [J].
Bourne, WM .
CORNEA, 2001, 20 (06) :560-569
[6]  
DELBOSC B, 1986, J FR OPHTALMOL, V9, P445
[7]   Automated tri-image analysis of stored corneal endothelium [J].
Gain, P ;
Thuret, G ;
Kodjikian, L ;
Gavet, Y ;
Turc, PH ;
Theillere, C ;
Acquart, S ;
Le petit, JC ;
Maugery, J ;
Campos, L .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2002, 86 (07) :801-808
[8]  
GAVET Y, 2004, 16 ANN M EUR EYE BAN
[9]   Comparison between manual count of the endothelial cells in donor corneas and computer-aided automatic count [J].
Kohlhof, JK ;
Ruprecht, KW .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2004, 221 (06) :473-478
[10]  
LESTER JM, 1981, INVEST OPHTH VIS SCI, V20, P407