Signal-to-Noise Contribution of Principal Component Loads in Reconstructed Near-Infrared Raman Tissue Spectra

被引:13
|
作者
Grimbergen, M. C. M. [1 ,2 ]
van Swol, C. F. P. [3 ]
Kendall, C. [4 ]
Verdaasdonk, R. M. [2 ]
Stone, N. [4 ]
Bosch, J. L. H. R. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Urol, Utrecht, Netherlands
[2] Univ Med Ctr, Dept Med Technol & Clin Phys, Utrecht, Netherlands
[3] St Antonius Hosp, Dept Med Phys, Nieuwegein, Netherlands
[4] Gloucestershire Hosp NHS Fdn Trust, Biophoton Res Unit, Gloucester, England
关键词
Signal-to-noise ratio; SNR; Principal component analysis; PCA; Raman spectroscopy; Bladder; IN-VIVO; SPECTROSCOPY; DIAGNOSIS; CANCER;
D O I
10.1366/000370210790572052
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The overall quality of Raman spectra in the near-infrared region, where biological samples are often studied, has benefited from various improvements to optical instrumentation over the past decade. However, obtaining ample spectral quality for analysis is still challenging due to device requirements and short integration times required for (in vivo) clinical applications of Raman spectroscopy. Multivariate analytical methods, such as principal component analysis (PCA) and linear discriminant analysis (LDA), are routinely applied to Raman spectral datasets to develop classification models. Data compression is necessary prior to discriminant analysis to prevent or decrease the degree of over-fitting. The logical threshold for the selection of principal components (PCs) to be used in discriminant analysis is likely to be at a point before the PCs begin to introduce equivalent signal and noise and, hence, include no additional value. Assessment of the signal-to-noise ratio (SNR) at a certain peak or over a specific spectral region will depend on the sample measured. Therefore, the mean SNR over the whole spectral region (SNRmsr) is determined in the original spectrum as well as for spectra reconstructed from an increasing number of principal components. This paper introduces a method of assessing the influence of signal and noise from individual PC loads and indicates a method of selection of PCs for LDA. To evaluate this method, two data sets with different SNRs were used. The sets were obtained with the same Raman system and the same measurement parameters on bladder tissue collected during white light cystoscopy (set A) and fluorescence-guided cystoscopy (set B). This method shows that the mean SNR over the spectral range in the original Raman spectra of these two data sets is related to the signal and noise contribution of principal component loads. The difference in mean SNR over the spectral range can also be appreciated since fewer principal components can reliably be used in the low SNR data set (set B) compared to the high SNR data set (set A). Despite the fact that no definitive threshold could be found, this method may help to determine the cutoff for the number of principal components used in discriminant analysis. Future analysis of a selection of spectral databases using this technique will allow optimum thresholds to be selected for different applications and spectral data quality levels.
引用
收藏
页码:8 / 14
页数:7
相关论文
共 50 条
  • [21] Ex vivo tissue discrimination by visible and near-infrared spectra with chemometrics
    Sakudo, Akikazu
    Tsenkova, Roumiana
    Tei, Kyoko
    Morita, Hiroyuki
    Ikuta, Kazuyoshi
    Onodera, Takashi
    JOURNAL OF VETERINARY MEDICAL SCIENCE, 2006, 68 (12) : 1375 - 1378
  • [22] Application of principal component analysis in calibration sample selection of near-infrared quantitative model
    Li Yan-Zhou
    Min Shun-Geng
    Liu Xia
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2007, 35 (09) : 1331 - 1334
  • [23] NEAR-INFRARED FT-RAMAN SPECTRA OF THE RAT-BRAIN TISSUES
    MIZUNO, A
    HAYASHI, T
    TASHIBU, K
    MARAISHI, S
    KAWAUCHI, K
    OZAKI, Y
    NEUROSCIENCE LETTERS, 1992, 141 (01) : 47 - 52
  • [24] SELECTION OF NEAR-INFRARED WAVELENGTHS FOR MONITORING MILK COAGULATION USING PRINCIPAL COMPONENT ANALYSIS
    SAPUTRA, D
    PAYNE, FA
    LODDER, RA
    SHEARER, SA
    TRANSACTIONS OF THE ASAE, 1992, 35 (05): : 1597 - 1605
  • [25] Principal Component Analysis and Near-Infrared Spectroscopy as Noninvasive Blood Glucose Assay Methods
    Barati, Hadi
    Madani, Arian Mousavi
    Shadzinavaz, Sorena
    Fardmanesh, Mehdi
    APPLIED SPECTROSCOPY, 2024,
  • [26] Further classification of eucalypt pulpwoods using principal components analysis of near-infrared spectra
    Michell, AJ
    Schimleck, LR
    APPITA JOURNAL, 1998, 51 (02): : 127 - 131
  • [27] Study on baseline correction methods or the Fourier transform infrared spectra with different signal-to-noise ratios
    Shen, Xianchun
    Ye, Shubin
    Xu, Liang
    Hu, Rong
    Jin, Ling
    Xu, Hanyang
    Liu, Jianguo
    Liu, Wenqing
    APPLIED OPTICS, 2018, 57 (20) : 5794 - 5799
  • [28] Data-driven signal-to-noise enhancement in scattering near-field infrared microscopy
    Baiz, Carlos R.
    Kanevche, Katerina
    Kozuch, Jacek
    Heberle, Joachim
    JOURNAL OF CHEMICAL PHYSICS, 2025, 162 (05)
  • [29] Peak detection algorithm of Raman spectra based on multi-scale local signal-to-noise ratio
    Sun, Q. (sunq@ciomp.ac.cn), 1600, Chinese Optical Society (34):
  • [30] An Entropy-based Method of Background and Noise Removal for Analysis of Near-Infrared Spectra
    Peng, Dan
    He, Yaqiang
    Dong, Kaina
    Li, Xia
    2010 4TH INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING (ICBBE 2010), 2010,