Signal-to-Noise Contribution of Principal Component Loads in Reconstructed Near-Infrared Raman Tissue Spectra

被引:13
|
作者
Grimbergen, M. C. M. [1 ,2 ]
van Swol, C. F. P. [3 ]
Kendall, C. [4 ]
Verdaasdonk, R. M. [2 ]
Stone, N. [4 ]
Bosch, J. L. H. R. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Urol, Utrecht, Netherlands
[2] Univ Med Ctr, Dept Med Technol & Clin Phys, Utrecht, Netherlands
[3] St Antonius Hosp, Dept Med Phys, Nieuwegein, Netherlands
[4] Gloucestershire Hosp NHS Fdn Trust, Biophoton Res Unit, Gloucester, England
关键词
Signal-to-noise ratio; SNR; Principal component analysis; PCA; Raman spectroscopy; Bladder; IN-VIVO; SPECTROSCOPY; DIAGNOSIS; CANCER;
D O I
10.1366/000370210790572052
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The overall quality of Raman spectra in the near-infrared region, where biological samples are often studied, has benefited from various improvements to optical instrumentation over the past decade. However, obtaining ample spectral quality for analysis is still challenging due to device requirements and short integration times required for (in vivo) clinical applications of Raman spectroscopy. Multivariate analytical methods, such as principal component analysis (PCA) and linear discriminant analysis (LDA), are routinely applied to Raman spectral datasets to develop classification models. Data compression is necessary prior to discriminant analysis to prevent or decrease the degree of over-fitting. The logical threshold for the selection of principal components (PCs) to be used in discriminant analysis is likely to be at a point before the PCs begin to introduce equivalent signal and noise and, hence, include no additional value. Assessment of the signal-to-noise ratio (SNR) at a certain peak or over a specific spectral region will depend on the sample measured. Therefore, the mean SNR over the whole spectral region (SNRmsr) is determined in the original spectrum as well as for spectra reconstructed from an increasing number of principal components. This paper introduces a method of assessing the influence of signal and noise from individual PC loads and indicates a method of selection of PCs for LDA. To evaluate this method, two data sets with different SNRs were used. The sets were obtained with the same Raman system and the same measurement parameters on bladder tissue collected during white light cystoscopy (set A) and fluorescence-guided cystoscopy (set B). This method shows that the mean SNR over the spectral range in the original Raman spectra of these two data sets is related to the signal and noise contribution of principal component loads. The difference in mean SNR over the spectral range can also be appreciated since fewer principal components can reliably be used in the low SNR data set (set B) compared to the high SNR data set (set A). Despite the fact that no definitive threshold could be found, this method may help to determine the cutoff for the number of principal components used in discriminant analysis. Future analysis of a selection of spectral databases using this technique will allow optimum thresholds to be selected for different applications and spectral data quality levels.
引用
收藏
页码:8 / 14
页数:7
相关论文
共 50 条
  • [11] Near-Infrared Spectra in Buccal Tissue as a Marker for Detection of Hypoxia
    Amini, Maryam
    Hisdal, Jonny
    Gjovaag, Terje
    Kapetanovic, Nihad
    Strand, Trond-Eirik
    Owe, Jan Ove
    Horthe, John Ragnar
    Mirtaheri, Peyman
    AEROSPACE MEDICINE AND HUMAN PERFORMANCE, 2016, 87 (05) : 498 - 504
  • [12] Principal component analysis of visible and near-infrared multispectral images of works of art
    Baronti, S
    Casini, A
    Lotti, F
    Porcinai, S
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1997, 39 (01) : 103 - 114
  • [13] Segregating wood wastes by repetitive principal component analysis of near infrared spectra
    Kobori, Hikaru
    Higa, Sakura
    Tsuchikawa, Satoru
    Kojima, Yoichi
    Suzuki, Shigehiko
    JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2017, 25 (03) : 180 - 187
  • [14] Denoising method for Raman spectra with low signal-to-noise ratio based on feature extraction
    Zhao, X. Y.
    Liu, G. Y.
    Sui, Y. T.
    Xu, M.
    Tong, L.
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2021, 250
  • [15] Principal Component Analysis Based Interconversion Between Infrared and Near-Infrared Spectra for the Study of Thermal-Induced Weak Interaction Changes of Poly(N-Isopropylacrylamide)
    Zhang, Liping
    Noda, Isao
    Wu, Yuqing
    APPLIED SPECTROSCOPY, 2009, 63 (06) : 694 - 699
  • [16] APPLICATION OF PRINCIPAL-COMPONENT ANALYSIS ON NEAR-INFRARED SPECTROSCOPIC DATA OF VEGETABLE-OILS FOR THEIR CLASSIFICATION
    SATO, T
    JOURNAL OF THE AMERICAN OIL CHEMISTS SOCIETY, 1994, 71 (03) : 293 - 298
  • [17] GOES-16 Advanced Baseline Imager visible near-infrared channel low-light signal-to-noise ratio
    Iacovazzi, Robbie
    Wu, Xiangqian
    JOURNAL OF APPLIED REMOTE SENSING, 2020, 14 (02):
  • [18] Detection of tea seed oil adulteration based on near-infrared and Raman spectra information fusion
    Zan, Jiajun
    Li, Hanlin
    Cai, Lexiao
    Wu, Chengzhao
    Fan, Zhengxin
    Sun, Tong
    LWT-FOOD SCIENCE AND TECHNOLOGY, 2024, 213
  • [19] Comparison of principal component analysis and generalized two-dimensional correlation spectroscopy: Spectral analysis of synthetic model system and near-infrared spectra of milk
    Sasic, S
    Ozaki, Y
    APPLIED SPECTROSCOPY, 2001, 55 (01) : 29 - 38
  • [20] Temperature-dependent near-infrared spectra of bovine serum albumin in aqueous solutions: Spectral analysis by principal component analysis and evolving factor analysis
    Yuan, B
    Murayama, K
    Wu, YQ
    Tsenkova, R
    Dou, XM
    Era, S
    Ozaki, Y
    APPLIED SPECTROSCOPY, 2003, 57 (10) : 1223 - 1229