Quantification of the heterogeneity in breast cancer cell lines using whole-cell impedance spectroscopy

被引:170
作者
Han, Arum
Yang, Lily
Frazier, A. Bruno
机构
[1] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30332 USA
[2] Texas A&M Univ, Dept Elect & Comp Engn, College Stn, TX USA
[3] Emory Univ, Dept Surg, Atlanta, GA 30322 USA
[4] Emory Univ, Winship Canc Inst, Atlanta, GA 30322 USA
关键词
D O I
10.1158/1078-0432.CCR-06-1346
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Quantification of the heterogeneity of tumor cell populations is of interest for many diagnostic and therapeutic applications, including determining the cancerous stage of tumors. We attempted to differentiate human breast cancer cell lines from different pathologic stages and compare that with a normal human breast tissue cell line by characterizing the impedance properties of each cell line. Experimental Design: A microelectrical impedance spectroscopy system has been developed that can trap a single cell into an analysis cavity and measure the electrical impedance of the captured cell over a frequency range from 100 Hz to 3.0 MHz. Normal human breast tissue cell line MCF-10A, early-stage breast cancer cell line MCF-7, invasive human breast cancer cell line MDA-MB-231, and metastasized human breast cancer cell line MDA-MB-435 were used. Results: The whole-cell impedance signatures show a clear difference between each cell line in both magnitude and phase of the electrical impedance. The membrane capacitance calculated from the impedance data was 1.94 +/- 0.14, 1.86 +/- 0.11, 1.63 +/- 0.17, and 1.57 +/- 0.12 mu F/cm(2) at 100 kHz for MCF-10A, MCF-7, MDA-MB-231, and MDA-MB-435, respectively. The calculated resistance for each cancer cell line at 100 kHz was 24.8 +/- 1.05, 24.8 +/- 0.93, 24.9 +/- 1.12, and 26.2 +/- 1.07 MOhm, respectively. The decrease in capacitances of the cancer cell lines compared with that of the normal cell line MCF-10A was 4.1%, 16.0%, and 19.1%, respectively, at 100 kHz. Conclusions: These findings suggest that microelectrical impedance spectroscopy might find application as a method for quantifying progression of cancer cells without the need for tagging or modifying the sampled cells.
引用
收藏
页码:139 / 143
页数:5
相关论文
共 24 条
  • [1] DIELECTRIC MEASUREMENT OF A SINGLE SUBMILLIMETER SIZE MICROCAPSULE
    ASAMI, K
    ZHAO, KS
    [J]. COLLOID AND POLYMER SCIENCE, 1994, 272 (01) : 64 - 71
  • [2] DIELECTRIC-DISPERSION OF A SINGLE SPHERICAL BILAYER-MEMBRANE IN SUSPENSION
    ASAMI, K
    IRIMAJIRI, A
    [J]. BIOCHIMICA ET BIOPHYSICA ACTA, 1984, 769 (02) : 370 - 376
  • [3] Impedance spectra of tumour tissue in comparison with normal tissue; A possible clinical application for electrical impedance tomography
    Blad, B
    Baldetorp, B
    [J]. PHYSIOLOGICAL MEASUREMENT, 1996, 17 : A105 - A115
  • [4] DIELECTRIC-PROPERTIES OF FEMALE HUMAN BREAST-TISSUE MEASURED INVITRO AT 3.2 GHZ
    CAMPBELL, AM
    LAND, DV
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (01) : 193 - 210
  • [5] Circulating tumor cells, disease progression, and survival in metastatic breast cancer
    Cristofanilli, M
    Budd, GT
    Ellis, MJ
    Stopeck, A
    Matera, J
    Miller, MC
    Reuben, JM
    Doyle, GV
    Allard, WJ
    Terstappen, LWMM
    Hayes, DF
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2004, 351 (08) : 781 - 791
  • [6] Rapid prototyping of microfluidic systems in poly(dimethylsiloxane)
    Duffy, DC
    McDonald, JC
    Schueller, OJA
    Whitesides, GM
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (23) : 4974 - 4984
  • [7] Ten-year risk of false positive screening mammograms and clinical breast examinations
    Elmore, G
    Barton, MB
    Moceri, VM
    Polk, S
    Arena, PJ
    Fletcher, SW
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (16) : 1089 - 1096
  • [8] FOSTER KR, 1981, J MICROWAVE POWER EE, V16, P107
  • [9] Dielectric spectroscopy in a micromachined flow cytometer: theoretical and practical considerations
    Gawad, S
    Cheung, K
    Seger, U
    Bertsch, A
    Renaud, P
    [J]. LAB ON A CHIP, 2004, 4 (03) : 241 - 251
  • [10] GAWAD S, 2001, MICRO TOTAL ANAL SYS, P253