Speech perception in nucleus C124M cochlear implant users with processor settings based on electrically evoked compound action potential thresholds

被引:105
作者
Smoorenburg, GF [1 ]
Willeboer, C [1 ]
van Dijk, JE [1 ]
机构
[1] Univ Utrecht, Med Ctr, Dept Otorhinolaryngol, NL-3508 GA Utrecht, Netherlands
关键词
cochlear implants; compound action potentials; processor fitting; speech perception;
D O I
10.1159/000066154
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Adjusting the speech processor of a cochlear implant, per electrode, to the individual's response is a laborious task that may interfere with a user-friendly start of implant-mediated hearing, particularly in children. This research concerns the possibility of processor adjustment based on a profile derived from measurements of the electrically evoked compound action potential (ECAP) thresholds across the electrode array, followed by adjustment of the overall level of the profile to the hearing threshold and maximum comfortable loudness level using live voice. The results for CVC word lists show that speech perception is quite insensitive to the threshold setting of the speech processor. On average, the speech score does not decrease by more than 10% when, with the new method, the threshold setting comes out so much lower that the dynamic range has doubled. In contrast, the speech score appears to be sensitive to an increase of the maximum high-frequency stimulation settings for the basal electrodes, resulting in lower scores at these higher settings. The correlation between the overall ECAP thresholds and conventionally measured subjective thresholds is weak (r = 0.64). However, the correlation between the slopes of these threshold curves is satisfactory (r = 0.82). The correlation between the ECAP thresholds and the maximum stimulation levels is poor, both with respect to overall level and slope (r = 0.39 and 0.36, respectively). Applicability of the ECAP threshold in processor adjustment could not be demonstrated in this study. Prediction of the most critical factor in speech perception, the slope of the maximum stimulation curve, from the ECAP thresholds is poor. However, considering habituation to the initial processor setting of at least 6 months, the small decrease in the CVC scores with the new setting suggests that a more user-friendly adjustment procedure can be developed. Copyright (C) 2002 S. Karger AG, Basel.
引用
收藏
页码:335 / 347
页数:13
相关论文
共 13 条
[1]   Summary of results using the nucleus CI24M implant to record the electrically evoked compound action potential [J].
Abbas, PJ ;
Brown, CJ ;
Shallop, JK ;
Firszt, JB ;
Hughes, ML ;
Hong, SH ;
Staller, SJ .
EAR AND HEARING, 1999, 20 (01) :45-59
[2]  
BOSMAN AJ, 1989, SPEECH PERCEPTION HE, pCH3
[3]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - PARAMETRIC DATA FROM THE CAT [J].
BROWN, CJ ;
ABBAS, PJ .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (05) :2205-2210
[4]   ELECTRICALLY EVOKED WHOLE-NERVE ACTION-POTENTIALS - DATA FROM HUMAN COCHLEAR IMPLANT USERS [J].
BROWN, CJ ;
ABBAS, PJ ;
GANTZ, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1990, 88 (03) :1385-1391
[5]   Electrically evoked whole nerve action potentials in ineraid cochlear implant users: Responses to different stimulating electrode configurations and comparison to psychophysical responses [J].
Brown, CJ ;
Abbas, PJ ;
Borland, J ;
Bertschy, MR .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1996, 39 (03) :453-467
[6]  
Brown CJ, 1998, AM J OTOL, V19, P320
[7]   The relationship between EAP and EABR thresholds and levels used to program the nucleus 24 speech processor: Data from adults [J].
Brown, CJ ;
Hughes, ML ;
Luk, B ;
Abbas, PJ ;
Wolaver, A ;
Gervais, J .
EAR AND HEARING, 2000, 21 (02) :151-163
[8]   Effects of dynamic range and amplitude mapping on phoneme recognition in nucleus-22 cochlear implant users [J].
Fu, QJ ;
Shannon, RV .
EAR AND HEARING, 2000, 21 (03) :227-235
[9]   Effects of amplitude nonlinearity on phoneme recognition by cochlear implant users and normal-hearing listeners [J].
Fu, QJ ;
Shannon, RV .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 104 (05) :2570-2577
[10]   Comparison of EAP thresholds with MAP levels in the nucleus 24 cochlear implant: Data from children [J].
Hughes, ML ;
Brown, CJ ;
Abbas, PJ ;
Wolaver, AA ;
Gervais, JP .
EAR AND HEARING, 2000, 21 (02) :164-174