Double potential step chronoamperometry at microdisk electrodes: simulating the case of unequal diffusion coefficients

被引:83
|
作者
Klymenko, OV
Evans, RG
Hardacre, C
Svir, IB
Compton, RG
机构
[1] Univ Oxford, Phys & Theoret Chem Lab, Oxford OX1 3QZ, England
[2] Kharkov Natl Univ Radioelect, Math & Comp Modelling Lab, UA-61166 Kharkov, Ukraine
[3] Queens Univ Belfast, Sch Chem, Belfast BT9 5AG, Antrim, North Ireland
基金
英国工程与自然科学研究理事会;
关键词
double potential step chronoamperometry; microdisk electrode; simulation; unequal diffusion coefficients;
D O I
10.1016/j.jelechem.2004.05.012
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An efficient approach to the simulation of the double potential step chronoamperometry at a microdisk electrode based on an exponentially expanding time grid and conformal mapping of the space is presented. The dimensionless second potential step flux data are included as a function of the first potential step duration and the ratio of the diffusion coefficients of the reacting species allowing instant analysis of the experimental double potential step chronoamperograms without a need for simulation. The values of the diffusion coefficients are determined for several test systems and found to be in good agreement with existing literature data. (C) 2004 Elsevier B.V. All rights reserved.