Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams

被引:36
作者
Bednarz, G
Huq, MS
Rosenow, UF
机构
[1] Thomas Jefferson Univ, Jefferson Med Coll, Kimmel Canc Ctr, Dept Radiat Oncol, Philadelphia, PA 19107 USA
[2] Univ Gottingen, Fac Med, Dept Radiat Biol & Radiat Phys, D-3400 Gottingen, Germany
关键词
D O I
10.1088/0031-9155/47/20/306
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents the results of measurements of output factors (OFs) for a model U Gamma Knife collimator, with special emphasis on the accurate determination of the OF for the 4 mm collimator (OF4). In the past, the OF4 was set to 0.800 relative to the 18 mm collimator. Recently, the manufacturer has recommended a new value of 0.870 for OF4. However, most centres still use the old value of the OF4. In the present study, the Gamma Knife OFs were measured using a commercially available miniature diamond detector and a miniature 0.006 cc ion chamber, which was especially designed for the task. The measured OF4 were corrected for spatial averaging effects by measuring dose profiles for the 4 mm collimator with the same detectors and deconvolving their response from the measured profiles. A Gaussian kernel was used to describe the detector response. The relative OFs measured with the diamond detector/ion chamber were 0.986/0.982,0.953/0.935 and 0.812/0.765 for the 14, 8 and 4 mm collimators, respectively, as compared with the manufacturer's values of 0.984, 0.956 and 0.87. The corrected OF4 was 0.881 +/- 0.012 for the diamond detector and 0.851 +/- 0.012 for the ion chamber, supporting the manufacturer's revised value for this collimator.
引用
收藏
页码:3643 / 3649
页数:7
相关论文
共 11 条
[1]  
[Anonymous], 1989, NUMERICAL RECIPES FO
[2]  
ARNDT J, 1999, MED PHYS, V26, P1059
[3]   A film technique for the determination of output factors and end effect times for the Leksell Gamma Knife [J].
Ekstrand, KE ;
Bourland, JD .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03) :703-706
[4]  
*EL INSTR AB, 1998, LEKS GAMM KNIF PUBL
[5]   Experimental determination of the convolution kernel for the study of the spatial response of a detector [J].
Garcia-Vicente, F ;
Delgado, JM ;
Peraza, C .
MEDICAL PHYSICS, 1998, 25 (02) :202-207
[6]  
Goetsch S, 1999, MED PHYS, V26, P1131
[7]   An efficient method of measuring the 4 mm helmet output factor for the Gamma Knife [J].
Ma, LJ ;
Li, XA ;
Yu, CX .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03) :729-733
[8]   On absorbed dose in narrow 60Co gamma-ray beams and dosimetry of the Gamma Knife [J].
Nizin, PS .
MEDICAL PHYSICS, 1998, 25 (12) :2347-2351
[9]   Dosimetric characterization of radiosurgical beams with a diamond detector [J].
Rustgi, SN ;
Frye, DMD .
MEDICAL PHYSICS, 1995, 22 (12) :2117-2121
[10]   INFLUENCE OF DETECTOR SIZE IN PHOTON-BEAM PROFILE MEASUREMENTS [J].
SIBATA, CH ;
MOTA, HC ;
BEDDAR, AS ;
HIGGINS, PD ;
SHIN, KH .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (05) :621-631