Acoustic properties of butadiene and silicone elastomers at megahertz frequencies

被引:7
作者
Burke, M. P. [1 ]
Smith, J. D. [2 ]
Carroll, N. L. [3 ]
Townend, D. J. [4 ]
Porter, D. [4 ]
Hoskins, P. R. [5 ]
机构
[1] QinetiQ Farnborough, Farnborough GU14 0LX, Hants, England
[2] DSTL Porton Down, Salisbury SP4 0JQ, Wilts, England
[3] QinetiQ Winfrith, Winfrith Technol Ctr, Dorchester DT2 8XJ, England
[4] DERA Farnborough, Farnborough, Hants, England
[5] Univ Edinburgh, Edinburgh, Midlothian, Scotland
关键词
Butadiene rubber; Ultrasonics; Acoustics; Dynamic properties; MECHANICAL-PROPERTIES;
D O I
10.1179/146580109X12473409437020
中图分类号
TB33 [复合材料];
学科分类号
摘要
A tissue mimic material with a sound velocity of 1540 m s(-1) and an acoustic attenuation of 0.5 dB MHz(-1) cm(-1) is required for calibrating medical Doppler ultrasound transducers. These properties can be obtained with a solid material comprised of a gel containing a high water content. If solid elastomers are to be used to provide a more stable tissue mimic, modelling suggests that the materials would require low viscoelastic damping, with a low glass transition temperature, T(g). Castable polyurethanes have been made using a polybutadiene soft segment and by adding process oil, the acoustic loss has been reduced to 0.9 dB MHz(-1) cm(-1) with a sound velocity of 1459 m s(-1). Unfilled butadiene rubber ( with a high-cis microstructure) had an acoustic loss around 1 dB MHz(-1) cm(-1), the low loss being associated with its low T(g), with a sound velocity of 1567 m s(-1). Silicone elastomers made from unfilled poly dimethyl siloxane had an acoustic loss of 0.71 dB MHz(-1) cm(-1), but the sound velocity was too low at 1033 m s(-1). Filler particles were added to raise the sound velocity, but they seemed to act as scattering centres and the composite had high acoustic attenuation. None of the elastomers were able to reproduce the acoustic attenuation characteristics of the tissue mimic gel.
引用
收藏
页码:343 / 348
页数:6
相关论文
共 11 条
  • [1] Acoustic and mechanical properties of polyurethanes based on polybutadiene soft segment
    Burke, M
    Townend, DJ
    [J]. PLASTICS RUBBER AND COMPOSITES, 1999, 28 (05) : 185 - 190
  • [2] GABRIELLI J, 1955, NOUOVO CIMENTO, V1, P403
  • [3] PROPAGATION OF LOW FREQUENCY ULTRASONIC WAVES IN RUBBERS AND RUBBER-LIKE POLYMERS
    HATFIELD, P
    [J]. BRITISH JOURNAL OF APPLIED PHYSICS, 1950, 1 (OCT): : 252 - 256
  • [4] Ultrasonic characterization of elastomers and elastomeric composites
    Jayaraman, S
    Dick, J
    Craychee, T
    Du, JK
    Tittmann, B
    [J]. SMART NONDESTRUCTIVE EVALUATION FOR HEALTH MONITORING OF STRUCTURAL AND BIOLOGICAL SYSTEMS, 2002, 4702 : 397 - 403
  • [5] KAYE GWC, 1990, TABLE PHYS CHEM CONS
  • [6] *LANX, 2008, BUN CB23 PROD DAT SH
  • [7] DYNAMIC MECHANICAL-PROPERTIES OF POLYBUTADIENE RUBBERS
    MOHSIN, MA
    BERRY, JP
    TRELOAR, LRG
    [J]. POLYMER, 1985, 26 (10) : 1463 - 1468
  • [8] Porter D., 1995, Group interaction modelling of polymer properties
  • [9] A WALL-LESS VESSEL PHANTOM FOR DOPPLER ULTRASOUND STUDIES
    RICKEY, DW
    PICOT, PA
    CHRISTOPHER, DA
    FENSTER, A
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (09) : 1163 - 1176
  • [10] *SART, 2005, SART PROD UR EL HYDR