Computer-Aided Pulmonary Nodule Detection - Performance of Two CAD Systems at Different CT Dose Levels

被引:20
作者
Hein, P. A. [1 ]
Rogalla, P. [1 ]
Klessen, C. [1 ]
Lembcke, A. [1 ]
Romano, V. C. [1 ]
机构
[1] Charite, Inst Radiol, Charite Campus Mitte, D-10117 Berlin, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2009年 / 181卷 / 11期
关键词
thorax; neoplasms; CT spiral; computer-aided detection; radiation dose; EARLY LUNG-CANCER; SPIRAL CT; CHEST CT; PRELIMINARY EXPERIENCE; AUTOMATED DETECTION; ASSISTED DETECTION; SCREENING-PROGRAM; MULTIDETECTOR CT; SLICE THICKNESS; BASE-LINE;
D O I
10.1055/s-0028-1109394
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the impact of dose reduction on the performance of computer-aided lung nodule detection systems (CAD) of two manufacturers by comparing respective CAD results on ultra-low-close computed tomography (ULD-CT) and standard dose CT (SD-CT). Materials and Methods: Multi-slice Computed tomography (MSCT) data sets of 26 patients (13 male and 13 female, patients 31 - 74 years old) were retrospectively selected for CAD analysis. Indication for CT examination was staging of a known primary malignancy or suspected pulmonary malignancy. CT images were consecutively acquired at 5 mAs (ULD-CT) and 75 mAs (SD-CT) with 120 kV tube voltage (I mm slice thickness). The standard of reference was determined by three experienced readers in consensus. CAD reading algorithms (pre-commercial CAD system, Philips, Netherlands: CAD-1; LungCARE, Siemens, Germany: CAD-2) were applied to the CT data sets. Results: Consensus reading identified 253 nodules on SD-CT and ULD-CT. Nodules ranged in diameter between 2 and 41 mm (mean diameter 4.8 mm). Detection rates were recorded with 72% and 62% (CAD-1 vs. CAD-2) for SD-CT and with 73% and 56% for ULD-CT. Median false positive rates per patient were calculated with 6 and 5 (CAD-1 vs. CAD-2) for SD-CT and with 8 and 3 for ULD-CT. After separate statistical analysis of nodules with diameters of 5 mm and greater, the detection rates increased to 83% and 61% for SD-CT and to 89% and 67% for ULD-CT (CAD-1 vs. CAD-2). For both CAD systems there were no significant differences between the detection rates for standard and ultra-low-dose data sets (p > 0.05). Conclusion: Dose reduction of the underlying CT scan did not significantly influence nodule detection performance of the tested CAD systems.
引用
收藏
页码:1056 / 1064
页数:9
相关论文
共 40 条
[31]   Pulmonary nodules on multi-detector row CT scans: Performance comparison of radiologists and computer-aided detection [J].
Rubin, GD ;
Lyo, JK ;
Paik, DS ;
Sherbondy, AJ ;
Chow, LC ;
Leung, AN ;
Mindelzun, R ;
Schraedley-Desmond, PK ;
Zinck, SE ;
Naidich, DP ;
Napel, S .
RADIOLOGY, 2005, 234 (01) :274-283
[32]   Computer-aided detection of pulmonary nodules in computed tomography: Analysis and review of the literature [J].
Saba, Luca ;
Caddeo, Giancarlo ;
Mallarini, Giorgio .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2007, 31 (04) :611-619
[33]   New imaging techniques in the treatment guidelines for lung cancer [J].
Schaefer-Prokop, C ;
Prokop, M .
EUROPEAN RESPIRATORY JOURNAL, 2002, 19 :71S-83S
[34]   SPIRAL CT OF THE CHEST - COMPARISON OF CINE AND FILM-BASED VIEWING [J].
SELTZER, SE ;
JUDY, PF ;
ADAMS, DF ;
JACOBSON, FL ;
STARK, P ;
KIKINIS, R ;
SWENSSON, RG ;
HOOTON, S ;
HEAD, B ;
FELDMAN, U .
RADIOLOGY, 1995, 197 (01) :73-78
[35]   CT-Expo - a novel program for dose evaluation in CT [J].
Stamm, G ;
Nagel, HD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (12) :1570-1576
[36]   BENEFIT OF INDEPENDENT DOUBLE READING IN A POPULATION-BASED MAMMOGRAPHY SCREENING-PROGRAM [J].
THURFJELL, EL ;
LERNEVALL, KA ;
TAUBE, AAS .
RADIOLOGY, 1994, 191 (01) :241-244
[37]   Automated pulmonary nodule volumetry with an optimized algorithm.: Accuracy at different slice thicknesses compared to unidimensional and bidimensional measurements [J].
Vogel, M. N. ;
Vonthein, R. ;
Schmuecker, S. ;
Maksimovich, O. ;
Bethge, W. ;
Dicken, V. ;
Claussen, C. D. ;
Horger, M. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2008, 180 (09) :791-797
[38]   Detection of pulmonary nodules at multirow-detector CT: effectiveness of double reading to improve sensitivity at standard-dose and low-dose chest CT [J].
Wormanns, D ;
Ludwig, K ;
Beyer, F ;
Heindel, W ;
Diederich, S .
EUROPEAN RADIOLOGY, 2005, 15 (01) :14-22
[39]   Automatic detection of pulmonary nodules at spiral CT: clinical application of a computer-aided diagnosis system [J].
Wormanns, D ;
Fiebich, M ;
Saidi, M ;
Diederich, S ;
Heindel, W .
EUROPEAN RADIOLOGY, 2002, 12 (05) :1052-1057
[40]   Computer-aided detection in screening CT for pulmonary nodules [J].
Yuan, R ;
Vos, PM ;
Cooperberg, PL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (05) :1280-1287