摘要
Objective: To evaluate the clinical efficacy of a new digital drainage system, the DigiVent (TM) Chest Drainage System that gives accurate measurements of air leakage and pleural. pressures and can display those measurements over time. Methods: The DigiVent (TM) Chest Drainage System was tested in three steps: Step 1; first clinical use in five patients, Step 2; management and acceptance in further 15 patients and Step 3; reliability in routine use in 50 patients. Result: The results from Steps 1 and 2 showed that the system was good enough for extended use. The results from Step 3 showed excellent clinical performance, however, we experienced device malfunctions in four cases without any consequences for the patients. The cause of the malfunctions was identified and steps taken to prevent their further occurrence. Conclusion: The ability of the DigiVent (TM) to measure airflow and present mean values for 1, 3 and 6 h proved to be very practical. The ability to save data and present curves for the entire course of treatment will be an advantage for research in the field of lung surgery. (c) 2007 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:209 / 213
页数:5
相关论文
共 26 条
[1]   Prolonged air leak following radical upper lobectomy - An analysis of incidence and possible risk factors [J].
Abolhoda, A ;
Liu, D ;
Brooks, A ;
Burt, M .
CHEST, 1998, 113 (06) :1507-1510
[2]   AIRFIX®:: the first digital postoperative chest tube airflowmetry -: a novel method to quantify air leakage after lung resection [J].
Anegg, Udo ;
Lindenmann, Jorg ;
Matzi, Veronika ;
Mujkic, Dzenana ;
Maier, Alfred ;
Fritz, Lukas ;
Smolle-Juettner, Freyja Maria .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2006, 29 (06) :867-872
[3]  
Antanavicius G, 2005, AM SURGEON, V71, P416
[4]   Comparison of function of commercially available pleural drainage units and catheters [J].
Baumann, AH ;
Patel, PB ;
Roney, CW ;
Petrini, MF .
CHEST, 2003, 123 (06) :1878-1886
[5]  
Baumann M H, 2000, Curr Opin Pulm Med, V6, P275, DOI 10.1097/00063198-200007000-00004
[6]   The clinician's perspective on pneumothorax management [J].
Baumann, MH ;
Strange, C .
CHEST, 1997, 112 (03) :822-828
[7]   The effect of autologous fibrin sealant (Vivostat®) on morbidity after pulmonary lobectomy:: a prospective randomised, blinded study [J].
Belboul, A ;
Dernevik, L ;
Aljassim, O ;
Skrbic, B ;
Radberg, G ;
Roberts, D .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 26 (06) :1187-1191
[8]   A prospective algorithm for the management of air leaks after pulmonary resection [J].
Cerfolio, RJ ;
Tummala, RP ;
Holman, WL ;
Zorn, GL ;
Kirklin, JK ;
McGiffin, DC ;
Naftel, DC ;
Pacifico, AD .
ANNALS OF THORACIC SURGERY, 1998, 66 (05) :1726-1730
[9]   Prospective randomized trial compares suction versus water seal for air leaks [J].
Cerfolio, RJ ;
Bass, C ;
Katholi, CR .
ANNALS OF THORACIC SURGERY, 2001, 71 (05) :1613-1617
[10]   Predictors and treatment of persistent air leaks [J].
Cerfolio, RJ ;
Bass, CS ;
Pask, AH ;
Katholi, CR .
ANNALS OF THORACIC SURGERY, 2002, 73 (06) :1727-1730