Development of Microelectrode Arrays for Artificial Retinal Implants Using Liquid Crystal Polymers

被引:59
作者
Lee, Seung Woo [1 ,3 ,4 ]
Seo, Jong-Mo [1 ,3 ,4 ]
Ha, Seungmin [2 ,3 ,4 ]
Kim, Eui Tae [1 ,3 ,4 ]
Chung, Hum [2 ,3 ,4 ]
Kim, Sung June [1 ,3 ,4 ]
机构
[1] Seoul Natl Univ, Sch Elect Engn & Comp Sci, Coll Med, Seoul 151742, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Ophthalmol, Seoul 151742, South Korea
[3] Seoul Natl Univ, Seoul Artificial Eye Ctr, Seoul 151742, South Korea
[4] Seoul Natl Univ, Nano Bioelect & Syst Res Ctr, Seoul 151742, South Korea
关键词
ELECTRICAL-STIMULATION; PROSTHESIS; ELECTRODE; LCP; FABRICATION;
D O I
10.1167/iovs.09-3743
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To develop a liquid crystal polymer (LCP)-based, long-term implantable, retinal stimulation microelectrode array using a novel fabrication method. METHODS. The fabrication process used laser micromachining and customized thermal-press bonding to produce LCP-based microelectrode arrays. To evaluate the fabrication process and the resultant electrode arrays, in vitro reliability tests and in vivo animal experiments were performed. The in vitro tests consisted of electrode site impedance recording and electrode interlayer adhesion monitoring during accelerated soak tests. For in vivo testing, the fabricated electrode arrays were implanted in the suprachoroidal space of rabbit eyes. Optical coherence tomography (OCT) and electrically evoked cortical potentials (EECPs) were used to determine long-term biocompatibility and functionality of the implant. RESULTS. The fabricated structure had a smooth, rounded edge profile and exhibited moderate flexibility, which are advantageous features for safe implantation without guide tools. After accelerated soak tests at 75 degrees C in phosphate-buffered saline, the electrode sites showed no degradation, and the interlayer adhesion of the structure showed acceptable stability for more than 2 months. The electrode arrays were safely implanted in the suprachoroidal space of rabbit eyes, and EECP waveforms were recorded. Over a 3-month postoperative period, no chorioretinal inflammation or structural deformities were observed by OCT and histologic examination. CONCLUSIONS. LCP-based flexible microelectrode arrays can be successfully applied as retinal prostheses. The results demonstrate that such electrode arrays are safe, biocompatible, and mechanically stable and that they can be effective as part of a chronic retinal implant system. (Invest Ophthalmol Vis Sci. 2009; 50: 5859-5866) DOI:10.1167/iovs.09-3743
引用
收藏
页码:5859 / 5866
页数:8
相关论文
共 23 条
[1]   In vivo studies of polypyrrole/peptide coated neural probes [J].
Cui, XY ;
Wiler, J ;
Dzaman, M ;
Altschuler, RA ;
Martin, DC .
BIOMATERIALS, 2003, 24 (05) :777-787
[2]  
CULBERTSON EC, 1995, EL COMP TECHN C MAY, P520
[3]   Realization of ultra fine pitch traces on LCP substrates [J].
Dean, Robert Neal, Jr. ;
Weller, Jennifer ;
Bozack, Mike J. M. J. ;
Rodekohr, Chad L. ;
Farrell, Brian ;
Jauniskis, Linas ;
Ting, Joseph ;
Edell, David J. ;
Hetke, Jamille F. .
IEEE TRANSACTIONS ON COMPONENTS AND PACKAGING TECHNOLOGIES, 2008, 31 (02) :315-321
[4]   AQUEOUS DEGRADATION OF POLYIMIDES [J].
DEIASI, R ;
RUSSELL, J .
JOURNAL OF APPLIED POLYMER SCIENCE, 1971, 15 (12) :2965-&
[5]   LONG-TERM MEASURES OF ELECTRODE IMPEDANCE AND AUDITORY-THRESHOLDS FOR THE INERAID COCHLEAR IMPLANT [J].
DORMAN, MF ;
SMITH, LM ;
DANKOWSKI, K ;
MCCANDLESS, G ;
PARKIN, JL .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1992, 35 (05) :1126-1130
[6]   Visual perception in a blind subject with a chronic microelectronic retinal prosthesis [J].
Humayun, MS ;
Weiland, JD ;
Fujii, GY ;
Greenberg, R ;
Williamson, R ;
Little, J ;
Mech, B ;
Cimmarusti, V ;
Van Boemel, G ;
Dagnelie, G ;
de Juan, E .
VISION RESEARCH, 2003, 43 (24) :2573-2581
[7]  
Jayaraj K, 1998, ADV MICROELECTRON, V25, P15
[8]  
KEESARA VV, 2006, MAT RES SOC S P E, V926
[9]   Fabrication of pillar shaped electrode arrays for artificial retinal implants [J].
Kim, Eui Tae ;
Seo, Jong-Mo ;
Woo, Se Joon ;
Zhou, Jing Ai ;
Chung, Hum ;
Kim, Sung June .
SENSORS, 2008, 8 (09) :5845-5856
[10]   Neural signal recording using microelectrode arrays fabricated on liquid crystal polymer material [J].
Lee, CJ ;
Oh, SJ ;
Song, JK ;
Kim, SJ .
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC AND SUPRAMOLECULAR SYSTEMS, 2004, 24 (1-2) :265-268