Architecture Design of the High Integrated System-on-Chip for Biomedical Applications

被引:0
作者
Rozanowski, Krzysztof [1 ]
Sondej, Tadeusz [2 ]
机构
[1] Mil Inst Aviat Med, Aviat Bioengn Dept, Warsaw, Poland
[2] Mil Univ Technol, Dept Elect, Warsaw, Poland
来源
MIXED DESIGN OF INTEGRATED CIRCUITS AND SYSTEMS, MIXDES 2013 | 2013年
关键词
biosensor; biomedical monitoring; body-area sensors; health monitoring; analog front-end; ECG; EMG; system-on-chip (SoC); signal acquisition; ASIC; SOC;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The article describes a new microchip measuring technology known as "BioSystem-on-Chip" (BioSoC). The solution will consist of an application-specific integrated circuit containing analog/digital signal processing circuits and a microcontroller for signal processing and analysis. Dedicated interfaces will enable measuring psycho-physiological signals immediately after connecting external measurement sensors, without the need for any calibration. The integrated microcontroller core will ensure necessary control capabilities and the processing and analysis of measurands. The BioSoC's structure will enable designing a complete monitoring and diagnostic system using a minimum number of discrete components. This approach will make the BioSoC integrable with new generation smartphones, without limitation, allowing for adding innovative human physiology monitoring functionalities. The proposed BioSoC architecture contains analog front-end circuits (with alternative designs available), an analog/digital processing circuit supported by DMA channels, a dedicated autonomous sampling engine, a 32-bit processor with integral SRAM, and required peripherals.
引用
收藏
页码:529 / 533
页数:5
相关论文
共 29 条
[1]  
Analog Devices Data Sheet, 2013, ADAS1000 63W P3W 3 E
[2]  
[Anonymous], BIOINFORMATICS BIOME, DOI DOI 10.1109/TBCAS.2009.2031627
[3]  
Chia-Ching Chou, 2011, Proceedings of the 1st IEEE First International Conference on Consumer Electronics - Berlin (IEEE ICCE-Berlin 2011), P172, DOI 10.1109/ICCE-Berlin.2011.6031850
[4]   Mechanically Flexible Wireless Multisensor Platform for Human Physical Activity and Vitals Monitoring [J].
Chuo, Yindar ;
Marzencki, Marcin ;
Hung, Benny ;
Jaggernauth, Camille ;
Tavakolian, Kouhyar ;
Lin, Philip ;
Kaminska, Bozena .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2010, 4 (05) :281-294
[5]   Universal application-specific integrated circuit for bioelectric data acquisition [J].
Fuchs, B ;
Vogel, S ;
Schroeder, D .
MEDICAL ENGINEERING & PHYSICS, 2002, 24 (10) :695-701
[6]  
Intan Technologies Data Sheet, 2012, RHA2000 FULL INT MUL
[7]  
Jocke SC, 2009, SYMP VLSI CIRCUITS, P60
[8]  
Kmon P., 2010, P 17 INT C MIX DES I, P163
[9]   An Ultra-Low Power ECG Acquisition and Monitoring ASIC System for WBAN Applications [J].
Liu, Xin ;
Zheng, Yuanjin ;
Phyu, Myint Wai ;
Endru, F. N. ;
Navaneethan, V. ;
Zhao, Bin .
IEEE JOURNAL ON EMERGING AND SELECTED TOPICS IN CIRCUITS AND SYSTEMS, 2012, 2 (01) :60-70
[10]  
Marcinek K, 2012, IEEE INT SYMP DESIGN, P26, DOI 10.1109/DDECS.2012.6219018