Experimental Study on Flame Temperature Measurement by Double Line of Atomic Emission Spectroscopy

被引:5
作者
Chen Xiao bin [1 ]
Cai Xiao shu [1 ]
Fan Xue liang [1 ]
Shen Jia-qi [1 ]
机构
[1] Shanghai Univ Sci & Technol, Sch Energy & Power Engn, Shanghai 200093, Peoples R China
关键词
Flame temperature; Atomic emission spectroscopy; Combustion; Measurement of temperature;
D O I
10.3964/j.issn.1000-0593(2009)12-3177-04
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The flame temperature was measured by the double line of atomic emission spectroscopy according to the spectra of K (766.5 and 769.9 nm) whose relative intensity was obtained by fiber spectrometer. The principles, methods and experiment system were described. The temperature measured by the double line of atomic emission spectroscopy was compared to the results measured by thermal couple under the condition of thermal equilibrium of blackbody furnace. The comparison indicated a good coherence between these two measurement methods. The method was demonstrated on coal powder and timber, and the temperature measured corresponded to reality.
引用
收藏
页码:3177 / 3180
页数:4
相关论文
共 8 条
[1]  
[程智海 Cheng Zhihai], 2004, [工程热物理学报, Journal of Engineering Thermophysics], V25, P519
[2]   A new synthesis method for signals for testing of flame-detection algorithms [J].
Fliess, T ;
Jentschel, HJ ;
Lenkheit, K .
FIRE SAFETY JOURNAL, 2002, 37 (02) :151-164
[3]  
MANTA V, 2000, 2 D SOOT DIAGNOSTIC
[4]  
NGENDAKUMANA P, 1998, 20 TASK LEAD M IEA C
[5]  
QIU DR, 2002, ATOMIC SPECTRAL ANAL
[6]  
ROMERO C, 2003, LEHIGH ENERGY UPDATE, V21, P1
[7]  
YANG JF, 1990, ATOMIC PHYS, V4
[8]  
ZHOU XT, 2002, SPECTROSC SPECT ANAL, V22, P407