Lung Nodule Assessment in Computed Tomography: Precision of Attenuation Measurement Based on Computer-Aided Volumetry

被引:6
作者
Knoess, N. [1 ]
Hoffmann, B. [1 ]
Fabel, M. [1 ]
Wiese, C. [1 ]
Jochens, A. [2 ]
Bolte, H. [1 ]
Heller, M. [1 ]
Biederer, J. [1 ]
机构
[1] Univ Klinikum Schleswig Holstein Campus Kiel, Radiol Diagnost Klin, D-24105 Kiel, Germany
[2] Univ Klinikum Schleswig Holstein Campus Kiel, Inst Med Informat & Stat, D-24105 Kiel, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2009年 / 181卷 / 12期
关键词
CT; volumetry; pulmonary nodule; SOLITARY PULMONARY NODULES; INTEROBSERVER-VARIABILITY; CT SCANS; ACCURACY; ENHANCEMENT;
D O I
10.1055/s-0028-1109785
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare the reproducibility (r) of CT value measurement of pulmonary nodules using volumetry software (LungCare, LC) and manual ROIs (mROI). Materials and Methods: 54 artificial nodules in a chest phantom were scanned three times with CT. CT values were measured with LC and mROI. The intrascan-r was assessed with three measurements in the first scan, and the interscan-r with measurements in three consecutive scans (one observer). Intrascan-r und interobserver-r (two obs.) were assessed in the first scan and in contrast-enhanced CT of 51 nodules from 15 patients (kernels b50f and b80f). Intrascan-r and Interscan-r were described as the mean range and interobserver-r as the mean difference of CT values. The significance of differences was tested using t-test and sign test. Results: Reproducibility was significantly higher for volumetry-based measurements in both artificial and patient nodules (range 0.11 vs. 6.16 HU for intrascan-r, 2.22 vs. 7.03 HU for interscan-r, difference 0.11 vs. 18.42 HU for interobserver-r; patients: 1.78 vs. 13.19 HU (b50f-Kernel) and 1.88 vs. 27.4 HU (b80f-Kernel) for intrascan-r, 3.71 vs. 22.43 HU for interobserver-r). Absolute CT values differed significantly between convolution kernels (pat./mROI: 29.3 [b50f] and 151.9 HU [b80f] pat./LC: 5 [b50f] and 147 HU [b80f]). Conclusion: The reproducibility of volumetry-based measurements of CT values in pulmonary nodules is significantly higher and should therefore be recommended, e.g. in dynamic chest CT protocols. Reproducibility does not depend on absolute CT values.
引用
收藏
页码:1151 / 1156
页数:6
相关论文
共 15 条
[1]   Artificial thorax for MR imaging studies in porcine heart-lung preparations [J].
Biederer, J ;
Heller, M .
RADIOLOGY, 2003, 226 (01) :250-255
[2]   Interobserver-variability of lung nodule volumetry considering different segmentation algorithms and observer training levels [J].
Bolte, H. ;
Jahnke, T. ;
Schaefer, F. K. W. ;
Wenke, R. ;
Hoffmann, B. ;
Freitag-Wolf, S. ;
Dicken, V. ;
Kuhnigk, J. M. ;
Lohmann, J. ;
Voss, S. ;
Knoess, N. ;
Heller, M. ;
Biederer, J. .
EUROPEAN JOURNAL OF RADIOLOGY, 2007, 64 (02) :285-295
[3]   Computed tomography-based lung nodule volumetry -: Do optimized reconstructions of routine protocols achieve similar accuracy, reproducibility and interobserver variability to that of special volumetry protocols? [J].
Bolte, H. ;
Riedel, C. ;
Knoess, N. ;
Freitag, S. ;
Hoffmann, B. ;
Heller, M. ;
Biederer, J. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2007, 179 (03) :276-281
[4]   Influence of convolution filtering on coronary plaque attenuation values: observations in an ex vivo model of multislice computed tomography coronary angiography [J].
Cademartiri, Filippo ;
La Grutta, Ludovico ;
Runza, Giuseppe ;
Palumbo, Alessandro ;
Maffei, Erica ;
Mollet, Nico R. A. ;
Bartolotta, Tommaso Vincenzo ;
Somers, Pamela ;
Knaapen, Michiel ;
Verheye, Stefan ;
Midiri, Massimo ;
Hamers, Ronald ;
Bruining, Nico .
EUROPEAN RADIOLOGY, 2007, 17 (07) :1842-1849
[5]   Solitary Pulmonary Nodule on Helical Dynamic CT Scans: Analysis of the Enhancement Patterns Using a Computer-Aided Diagnosis (CAD) System [J].
Choi, Eun Jung ;
Jin, Gong Yong ;
Han, Young Min ;
Lee, Young Sun ;
Kweon, Keun Sang .
KOREAN JOURNAL OF RADIOLOGY, 2008, 9 (05) :401-408
[6]   Accuracy of automated volumetry of pulmonary nodules across different multislice CT scanners [J].
Das, Marco ;
Ley-Zaporozhan, Julia ;
Gietema, H. A. ;
Czech, Andre ;
Muehlenbruch, Georg ;
Mahnken, Andreas H. ;
Katoh, Markus ;
Bakai, Annemarie ;
Salganicoff, Marcos ;
Diederich, Stefan ;
Prokop, Mathias ;
Kauczor, Hans-Ulrich ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
EUROPEAN RADIOLOGY, 2007, 17 (08) :1979-1984
[7]   Accuracy of the CT numbers of simulated lung nodules imaged with multi-detector CT scanners [J].
Goodsitt, Mitchell M. ;
Chan, Heang-Ping ;
Way, Ted W. ;
Larson, Sandra C. ;
Christodoulou, Emmanuel G. ;
Kim, Jeomsoon .
MEDICAL PHYSICS, 2006, 33 (08) :3006-3017
[8]   Linear and Volume Measurements of Pulmonary Nodules at Different CT Dose Levels - Intrascan and Interscan Analysis [J].
Hein, P. A. ;
Romano, V. C. ;
Rogalla, P. ;
Klessen, C. ;
Lembcke, A. ;
Dickenz, V. ;
Bornemannz, L. ;
Bauknecht, H. -C. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (01) :24-31
[9]   CT screening for lung cancer: Suspiciousness of nodules according to size on baseline scans [J].
Henschke, CI ;
Yankelevitz, DF ;
Naidich, DP ;
McCauley, DI ;
McGuinness, G ;
Libby, DM ;
Smith, JP ;
Pasmantier, MW ;
Miettinen, OS .
RADIOLOGY, 2004, 231 (01) :164-168
[10]   Solid or partly solid solitary pulmonary nodules - Their characterization using contrast wash-in and morphologic features at helical CT [J].
Lee, Kyung Soo ;
Yi, Chin A. ;
Jeong, Sun Young ;
Jeong, Yeon Joo ;
Kim, Seonwoo ;
Chung, Myung Jin ;
Kim, Ha Young ;
Kim, Yoon Kyung ;
Lee, Kwang Hwi .
CHEST, 2007, 131 (05) :1516-1525