Factors Associated with Failure of the Artificial Bowel Sphincter: A Study of Over 50 Cases from Cleveland Clinic Florida

被引:39
作者
Wexner, Steven D. [1 ]
Jin, Hei Ying [1 ]
Weiss, Eric G. [1 ]
Nogueras, Juan J. [1 ]
Li, Vicky Ka Ming [1 ]
机构
[1] Cleveland Clin Florida, Dept Colorectal Surg, Weston, FL 33331 USA
关键词
Artificial bowel sphincter; Failure; Complications; Morbidity; FECAL INCONTINENCE; ANAL-SPHINCTER; IMPLANTATION; NEOSPHINCTER; COMPLICATIONS; SURGERY; SAFETY; TRIAL;
D O I
10.1007/DCR.0b013e3181af62f8
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
PURPOSE: This study investigated the risk factors related to artificial bowel sphincter infection, complications, and failure. METHOD: Complications may occur at any time after artificial bowel sphincter implantation. Early-stage complication is defined as any complications that occurred before artificial bowel sphincter activation, whereas late-stage complications are defined as any complications that occurred after device activation. Assessment of the outcomes of all artificial bowel sphincter operations included evaluation of factors related to patient demographics, operative procedures, and postoperative events. RESULT: From January 1998 to May 2007, 51 artificial bowel sphincter implantations were performed in 47 patients (43; 84.3% female) with a mean age of 48.8 +/- 12.5 (range, 19-79) years and a mean incontinence score of 18 +/- 1.4 (range, 0-20). In 24 patients (54.5%), the etiology of incontinence was secondary to imperforate anus; 15 (24.2%) patients had obstetric injury or anorectal trauma. Twenty-three (41.2%) artificial bowel sphincter implantations became infected, 18 (35.3%) of which developed early-stage infection, whereas 5 (5.9%) had late-stage infection. One patient in the latter group had associated erosion, and two patient had fistula formation. Late-stage complications continued to increase with time. Multivariate analysis revealed that the time between artificial bowel sphincter implantation and first bowel movement and a history of perineal sepsis were independent risk factors for early-stage artificial bowel sphincter infection. CONCLUSION: The time from implantation to first bowel movement and history of perineal infection were risk factors for early-stage artificial bowel sphincter infection and failure. Late-stage failures were more often the result of device malfunction and indicated the need for mechanical refinement.
引用
收藏
页码:1550 / 1557
页数:8
相关论文
共 20 条
[1]   Multicentre retrospective analysis of the outcome of artificial anal sphincter implantation for severe faecal incontinence [J].
Altomare, DF ;
Dodi, G ;
La Torre, F ;
Romano, G ;
Melega, E ;
Rinaldi, M .
BRITISH JOURNAL OF SURGERY, 2001, 88 (11) :1481-1486
[2]   Neosphincter surgery for fecal incontinence:: A critical and unbiased review of the relevant literature [J].
Belyaev, O ;
Müller, C ;
Uhl, W .
SURGERY TODAY, 2006, 36 (04) :295-303
[3]   Clinical examination, endosonography, and MR imaging in preoperative assessment of fistula in ano: Comparison with outcome-based reference standard [J].
Buchanan, GN ;
Halligan, S ;
Bartram, CI ;
Williams, AB ;
Tarroni, D ;
Cohen, CRG .
RADIOLOGY, 2004, 233 (03) :674-681
[4]  
CHRISTIANSEN J, 1987, LANCET, V2, P244
[5]   New surgical options for fecal incontinence in patients with imperforate anus [J].
da Silva, GM ;
Jorge, JMN ;
Belin, B ;
Nogueras, JJ ;
Weiss, EG ;
Vernava, AM ;
Habr-Gama, A ;
Wexner, SD .
DISEASES OF THE COLON & RECTUM, 2004, 47 (02) :204-209
[6]   Artificial anal sphincter -: Complications and functional results of a large personal series [J].
Devesa, JM ;
Rey, A ;
Hervas, PL ;
Halawa, KS ;
Larrañaga, I ;
Svidler, L ;
Abraira, V ;
Muriel, A .
DISEASES OF THE COLON & RECTUM, 2002, 45 (09) :1154-1163
[7]   Comparison of quality of life and anorectal function after artificial sphincter implantation [J].
Lehur, PA ;
Zerbib, F ;
Neunlist, M ;
Glemain, P ;
des Varannes, SB .
DISEASES OF THE COLON & RECTUM, 2002, 45 (04) :508-513
[8]  
Madoff RD, 2000, DIS COLON RECTUM, V43, P1635, DOI 10.1007/BF02236756
[9]  
MARCIO J, 1993, DIS COLON RECTUM, V36, P77
[10]   Systematic review of safety and effectiveness of an artificial bowel sphincter for faecal incontinence [J].
Mundy, L ;
Merlin, TL ;
Maddern, GJ ;
Hiller, JE .
BRITISH JOURNAL OF SURGERY, 2004, 91 (06) :665-672