Development of an enzymatic reactor applying spontaneously adsorbed trypsin on the surface of a PDMS microfluidic device

被引:19
作者
Kecskemeti, Adam [1 ]
Bako, Jozsef [2 ]
Csarnovics, Istvan [3 ]
Csosz, Eva [4 ]
Gaspar, Attila [1 ]
机构
[1] Univ Debrecen, Dept Inorgan & Analyt Chem, Egyet Ter 1, H-4032 Debrecen, Hungary
[2] Univ Debrecen, Dept Biomat & Prosthet Dent, Nagyerdei Krt 98, H-4032 Debrecen, Hungary
[3] Univ Debrecen, Dept Expt Phys, Bem Ter 18-A, H-4026 Debrecen, Hungary
[4] Univ Debrecen, Dept Biochem & Mol Biol, Egyet Ter 1, H-4032 Debrecen, Hungary
关键词
Immobilization; Trypsin; Digestion; Peptide mapping; Enzyme reactor; Microfluidic device; CAPILLARY ZONE ELECTROPHORESIS; EFFICIENT PROTEIN DIGESTION; IMMOBILIZED TRYPSIN; MASS-SPECTROMETRY; PEPTIDE SEPARATION; IDENTIFICATION; MICROREACTOR; PROTEOLYSIS; ADSORPTION; MICROCHIP;
D O I
10.1007/s00216-017-0295-9
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Herein, a microfluidic device (MD) containing immobilized trypsin for rapid and efficient proteolysis was described. Trypsin was immobilized via non-specific protein adsorption onto the hydrophobic poly(dimethylsiloxane) (PDMS) channel wall of the MD. Peptide mapping of bovine serum albumin (BSA) samples was carried out to estimate the stability of trypsin adsorbed on PDMS surface. Peptide maps of BSA samples were obtained by capillary zone electrophoresis (CZE), the RSD% for migration times were under 1%. Several proteins (hemoglobin, myoglobin, lysozyme, and BSA) in a wide molecular size range (15-70 kDa) were digested efficiently with similar to 50 s contact time. The number of separated peaks correlated well with the expected number of peptides formed in the complete tryptic digestion of the proteins. Peptide mass fingerprinting of BSA and human serum was carried out. Trypsin retained its activity for 2 h; within this period, the MD can be used for multiple digestions. The main properties of this device are simple channel pattern, simple immobilization procedure, regenerability, and disposability; all these features make this MD one of the simplest yet applicable enzymatic microreactors.
引用
收藏
页码:3573 / 3585
页数:13
相关论文
共 56 条
  • [1] TRYPSIN-MODIFIED FUSED-SILICA CAPILLARY MICROREACTOR FOR PEPTIDE-MAPPING BY CAPILLARY ZONE ELECTROPHORESIS
    AMANKWA, LN
    KUHR, WG
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (14) : 1610 - 1613
  • [2] Reproducibility of a solid-phase trypsin microreactor for peptide mapping by capillary electrophoresis
    Bonneil, E
    Mercier, M
    Waldron, KC
    [J]. ANALYTICA CHIMICA ACTA, 2000, 404 (01) : 29 - 45
  • [3] Investigation of fundamental physical properties of a polydimethylsiloxane (PDMS) membrane using a proton transfer reaction mass spectrometer (PTRMS)
    Boscaini, E
    Alexander, ML
    Prazeller, P
    Märk, TD
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2004, 239 (2-3) : 179 - 186
  • [4] Integrated microfluidic system enabling (bio)chemical reactions with on-line MALDI-TOF mass spectrometry
    Brivio, M
    Fokkens, RH
    Verboom, W
    Reinhoudt, DN
    Tas, NR
    Goedbloed, M
    van den Berg, A
    [J]. ANALYTICAL CHEMISTRY, 2002, 74 (16) : 3972 - 3976
  • [5] Development of a bioreactor based on trypsin immobilized on monolithic support for the on-line digestion and identification of proteins
    Calleri, E
    Temporini, C
    Perani, E
    Stella, C
    Rudaz, S
    Lubda, D
    Mellerio, G
    Veuthey, JL
    Caccialanza, G
    Massolini, G
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2004, 1045 (1-2) : 99 - 109
  • [6] Nanostructured microfluidic digestion system for rapid high-performance proteolysis
    Cheng, Gong
    Hao, Si-Jie
    Yu, Xu
    Zheng, Si-Yang
    [J]. LAB ON A CHIP, 2015, 15 (03) : 650 - 654
  • [7] Construction of a high-performance magnetic enzyme nanosystem for rapid tryptic digestion
    Cheng, Gong
    Zheng, Si-Yang
    [J]. SCIENTIFIC REPORTS, 2014, 4
  • [8] Immobilization of trypsin onto multifunctional meso-fmacroporous core-shell microspheres: A new platform for rapid enzymatic digestion
    Cheng, Gong
    Chen, Ping
    Wang, Zhi-Gang
    Sui, Xiao-Jing
    Zhang, Ji-Lin
    Ni, Jia-Zuan
    [J]. ANALYTICA CHIMICA ACTA, 2014, 812 : 65 - 73
  • [9] Adsorption of proteins to thin-films of PDMS and its effect on the adhesion of human endothelial cells
    Chumbimuni-Torres, Karin Y.
    Coronado, Ramon E.
    Mfuh, Adelphe M.
    Castro-Guerrero, Carlos
    Silva, Maria Fernanda
    Negrete, George R.
    Bizios, Rena
    Garcia, Carlos D.
    [J]. RSC ADVANCES, 2011, 1 (04) : 706 - 714
  • [10] HIGH-SENSITIVITY PEPTIDE-MAPPING BY CAPILLARY ZONE ELECTROPHORESIS AND MICROCOLUMN LIQUID-CHROMATOGRAPHY, USING IMMOBILIZED TRYPSIN FOR PROTEIN DIGESTION
    COBB, KA
    NOVOTNY, M
    [J]. ANALYTICAL CHEMISTRY, 1989, 61 (20) : 2226 - 2231