A physiologically realistic in vitro model of microvascular networks

被引:70
作者
Rosano, Jenna M. [1 ]
Tousi, Nazanin [1 ]
Scott, Robert C. [1 ]
Krynska, Barbara [2 ]
Rizzo, Victor [3 ]
Prabhakarpandian, Balabhaskar [4 ]
Pant, Kapil [4 ]
Sundaram, Shivshankar [4 ]
Kiani, Mohammad F. [1 ,5 ]
机构
[1] Temple Univ, Dept Mech Engn, Philadelphia, PA 19122 USA
[2] Temple Univ, Dept Neurol, Philadelphia, PA 19122 USA
[3] Temple Univ, Dept Anat & Cell Biol, Philadelphia, PA 19122 USA
[4] CFD Res Corp, Biomed Technol Dept, Huntsville, AL USA
[5] Temple Univ, Dept Radiat Oncol, Philadelphia, PA 19122 USA
关键词
Microfluidic device; Microvascular network; Flow chamber; Adhesion molecules;
D O I
10.1007/s10544-009-9322-8
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Existing microfluidic devices, e.g. parallel plate flow chambers, do not accurately depict the geometry of microvascular networks in vivo. We have developed a synthetic microvascular network (SMN) on a polydimethalsiloxane (PDMS) chip that can serve as an in vitro model of the bifurcations, tortuosities, and cross-sectional changes found in microvascular networks in vivo. Microvascular networks from a cremaster muscle were mapped using a modified Geographical Information System, and then used to manufacture the SMNs on a PDMS chip. The networks were cultured with bovine aortic endothelial cells (BAEC), which reached confluency 3-4 days after seeding. Propidium iodide staining indicated viable and healthy cells showing normal behavior in these networks. Anti-ICAM-1 conjugated 2-mu m microspheres adhered to BAEC cells activated with TNF-alpha in significantly larger numbers compared to control IgG conjugated microspheres. This preferential adhesion suggests that cultured cells retain an intact cytokine response in the SMN. This microfluidic system can provide novel insight into characterization of drug delivery particles and dynamic flow conditions in microvascular networks.
引用
收藏
页码:1051 / 1057
页数:7
相关论文
共 31 条
  • [1] CAMP JP, 2007, BIOMED MICRODEVICES, V10, P2
  • [2] Adhesive dynamics simulations of the shear threshold effect for leukocytes
    Caputo, Kelly E.
    Lee, Dooyoung
    King, Michael R.
    Hammer, Daniel A.
    [J]. BIOPHYSICAL JOURNAL, 2007, 92 (03) : 787 - 797
  • [3] FABRICATION OF IN-VITRO MICROVASCULAR BLOOD-FLOW SYSTEMS BY PHOTOLITHOGRAPHY
    COKELET, GR
    SOAVE, R
    PUGH, G
    RATHBUN, L
    [J]. MICROVASCULAR RESEARCH, 1993, 46 (03) : 394 - 400
  • [4] Design maps for nanoparticles targeting the diseased microvasculature
    Decuzzi, Paolo
    Ferrari, Mauro
    [J]. BIOMATERIALS, 2008, 29 (03) : 377 - 384
  • [5] Cells on chips
    El-Ali, Jamil
    Sorger, Peter K.
    Jensen, Klavs F.
    [J]. NATURE, 2006, 442 (7101) : 403 - 411
  • [6] Endothelialized microvasculature based on a biodegradable elastomer
    Fidkowski, C
    Kaazempur-Mofrad, MR
    Borenstein, J
    Vacanti, JP
    Langer, R
    Wang, YD
    [J]. TISSUE ENGINEERING, 2005, 11 (1-2): : 302 - 309
  • [7] Targeted drug delivery by in vivo coupling to endogenous albumin: an albumin-binding prodrug of methotrexate (MTX) is better than MTX in the treatment of murine collagen-induced arthritis
    Fiehn, C.
    Kratz, F.
    Sass, G.
    Mueller-Ladner, U.
    Neumann, E.
    [J]. ANNALS OF THE RHEUMATIC DISEASES, 2008, 67 (08) : 1188 - 1191
  • [8] Frame Mary D. S., 1995, Microcirculation (New York), V2, P377, DOI 10.3109/10739689509148282
  • [9] GAEHTGENS P, 1992, INT J MICROCIRC CLIN, V11, P2
  • [10] KELLY KJ, 2003, AM J PHYSIOL-CELL PH, V284, P2