Cement augmentation of vertebral screws enhances the interface strength between interbody device and vertebral body

被引:49
作者
Tan, Juay-Seng
Bailey, Christopher S.
Dvorak, Marcel F.
Fisher, Charles G.
Cripton, Peter A.
Oxland, Thomas R.
机构
[1] Univ British Columbia, Dept Mech Engn, Vancouver, BC V6T 1W5, Canada
[2] Univ British Columbia, Dept Orthopaed, Vancouver, BC V6T 1W5, Canada
[3] Vancouver Gen Hosp, Vancouver, BC V6T 1W5, Canada
[4] Vancouver Coastal Hlth Res Inst, Vancouver, BC V6T 1W5, Canada
关键词
biomechanics; spinal fusion; internal fixators; lumbar vertebrae; thoracic vertebrae; bone cement;
D O I
10.1097/01.brs.0000253645.24141.21
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. An in vitro cadaveric study comparing cage - vertebra interface strengths for 3 different screw cement configurations. Objectives. To determine the effects of cement augmentation of pedicle screws on cage - vertebra interface failure properties for 2 interbody device shapes ( elliptical or cloverleaf); and to compare between pedicle and anterior vertebral body screws with cement augmentation. Summary of Background Data. Pedicle or anterior screw fixation is commonly used with interbody device fixation. Cement has recently been shown to augment screw fixation in the osteoporotic spine by improving the screw bone interface strength. The effect of cement augmentation of pedicle or anterior screws on cage - vertebra interface properties has not been previously studied or compared. Methods. An elliptical or a cloverleaf-shaped indentor covering 40% of the endplate was axially compressed against the superior endplate of 48 thoracolumbar vertebrae. Each vertebra had polymethylmethacrylate cement augmentation of 1) anterior screws, 2) pedicle screws, or 3) pedicle screws without cement. Compressive load was applied through a mechanism that allowed unconstrained rotation of the indentors. Results. Cement augmentation of pedicle screws resulted in significantly higher failure loads ( 54%) and failure strength ( 69%) for both shaped indentors when compared with uncemented pedicle screws. There was no significant difference in failure load and failure strength between pedicle and anterior screws with cement augmentation. Indentor shape was not a significant factor on failure load or failure strength. Conclusions. Cage - vertebra interface properties were improved when cement was used to augment vertebral and pedicle screws. Cement augmentation of pedicle or anterior screws may reduce interbody device subsidence.
引用
收藏
页码:334 / 341
页数:8
相关论文
共 52 条
[1]   Population aging: A comparison among industrialized countries [J].
Anderson, GF ;
Hussey, PS .
HEALTH AFFAIRS, 2000, 19 (03) :191-203
[2]   The use of an injectable, biodegradable calcium phosphate bone substitute for the prophylactic augmentation of osteoporotic vertebrae and the management of vertebral compression fractures [J].
Bai, B ;
Jazrawi, LM ;
Kummer, FJ ;
Spivak, JM .
SPINE, 1999, 24 (15) :1521-1526
[3]   Augmentation of anterior vertebral body screw fixation by an injectable, biodegradable calcium phosphate bone substitute [J].
Bai, B ;
Kummer, FJ ;
Spivak, J .
SPINE, 2001, 26 (24) :2679-2683
[4]   Percutaneous vertebroplasty for pain relief and spinal stabilization [J].
Barr, JD ;
Barr, MS ;
Lemley, TJ ;
McCann, RM .
SPINE, 2000, 25 (08) :923-928
[5]   An ex vivo biomechanical evaluation of a hydroxyapatite cement for use with vertebroplasty [J].
Belkoff, SM ;
Mathis, JM ;
Jasper, LE ;
Deramond, H .
SPINE, 2001, 26 (14) :1542-1546
[6]   An ex vivo biomechanical evaluation of an inflatable bone tamp used in the treatment of compression fracture [J].
Belkoff, SM ;
Mathis, JM ;
Fenton, DC ;
Scribner, RM ;
Reiley, ME ;
Talmadge, K .
SPINE, 2001, 26 (02) :151-156
[7]   Biomechanical evaluation of a new bone cement for use in vertebroplasty [J].
Belkoff, SM ;
Mathis, JM ;
Erbe, EM ;
Fenton, DC .
SPINE, 2000, 25 (09) :1061-1064
[8]   An in vitro biomechanical evaluation of bone cements used in percutaneous vertebroplasty [J].
Belkoff, SM ;
Maroney, M ;
Fenton, DC ;
Mathis, JM .
BONE, 1999, 25 (02) :23S-26S
[9]   INFLUENCE OF BONE-MINERAL DENSITY ON THE FIXATION OF THORACOLUMBAR IMPLANTS - A COMPARATIVE-STUDY OF TRANSPEDICULAR SCREWS, LAMINAR HOOKS, AND SPINOUS PROCESS WIRES [J].
COE, JD ;
WARDEN, KE ;
HERZIG, MA ;
MCAFEE, PC .
SPINE, 1990, 15 (09) :902-907
[10]  
Cook Stephen D, 2004, Spine J, V4, P402, DOI 10.1016/j.spinee.2003.11.010