Quality assurance of volumetric modulated arc therapy using Elekta Synergy

被引:41
作者
Haga, Akihiro [1 ]
Nakagawa, Keiichi [1 ]
Shiraishi, Kenshiro [1 ]
Itoh, Saori [1 ]
Terahara, Atsuro [1 ]
Yamashita, Hideomi [1 ]
Ohtomo, Kuni [1 ]
Saegusa, Shigeki [1 ]
Imae, Toshikazu [1 ]
Yoda, Kiyoshi [2 ]
Pellegrini, Roberto [3 ]
机构
[1] Tokyo Univ Hosp, Dept Radiol, Bunkyo Ku, Tokyo 1138655, Japan
[2] Elekta KK, Kobe, Hyogo, Japan
[3] 3D Line, Elekta, Milan, Italy
关键词
D O I
10.3109/02841860903081905
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose. Recently, Elekta has supplied volumetric modulated arc therapy (VMAT) in which multi-leaf collimator (MLC) shape, jaw position, collimator angle, and gantry speed vary continuously during gantry rotation. A quality assurance procedure for VMAT delivery is described. Methods and materials. A single-arc VMAT plan with 73 control points (CPs) and 5-degree gantry angle spacing for a prostate cancer patient has been created by ERGO++ treatment planning system (TPS), where MLC shapes are given by anatomic relationship between a target and organs at risk and the monitor unit for each CP is optimized based on given dose prescriptions. Actual leaf and jaw positions, gantry angles and dose rates during prostate VMAT delivery were recorded in every 0.25 seconds, and the errors between planned and actual values were evaluated. The dose re-calculation using these recorded data has been performed and compared with the original TPS plan using the gamma index. Results. Typical peak errors of gantry angles, leaf positions, and jaw positions were 3 degrees, 0.6 mm, and 1 mm, respectively. The dose distribution obtained by the TPS plan and the recalculated one agreed well under 2%-2 mm gamma index criteria. Conclusions. Quality assurance for prostate VMAT delivery has been performed with a satisfied result.
引用
收藏
页码:1193 / 1197
页数:5
相关论文
共 10 条
  • [1] Treatment of lung cancer using volumetric modulated arc therapy and image guidance: A case study
    Bedford, James L.
    Hansen, Vibeke Nordmark
    Mcnair, Helen A.
    Aitken, Alexandra H.
    Brock, Juliet E. C.
    Warrington, Alan P.
    Brada, Michael
    [J]. ACTA ONCOLOGICA, 2008, 47 (07) : 1438 - 1443
  • [2] COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT)
    Bedford, James L.
    Warrington, Alan P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02): : 537 - 545
  • [3] Inverse planning for intensity-modulated arc therapy using direct aperture optimization
    Earl, MA
    Shepard, DM
    Naqvi, S
    Li, XA
    Yu, CX
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (08) : 1075 - 1089
  • [4] EARL MA, 2002, INT J RADIAT ONCOL, V53, P222
  • [5] Dosimetric verification of RapidArc treatment delivery
    Korreman, Stine
    Medin, Joakim
    Kjaer-Kristoffersen, Flemming
    [J]. ACTA ONCOLOGICA, 2009, 48 (02) : 185 - 191
  • [6] Volumetric modulated arc therapy: IMRT in a single gantry arc
    Otto, Karl
    [J]. MEDICAL PHYSICS, 2008, 35 (01) : 310 - 317
  • [7] An arc-sequencing algorithm for intensity modulated arc therapy
    Shepard, D. M.
    Cao, D.
    Afghan, M. K. N.
    Earl, M. A.
    [J]. MEDICAL PHYSICS, 2007, 34 (02) : 464 - 470
  • [8] Takahashi S., 1965, Acta Radiol Diagn Suppl, V242, P1
  • [9] INTENSITY-MODULATED ARC THERAPY WITH DYNAMIC MULTILEAF COLLIMATION - AN ALTERNATIVE TO TOMOTHERAPY
    YU, CX
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (09) : 1435 - 1449
  • [10] Clinical implementation of intensity-modulated arc therapy
    Yu, CX
    Li, XA
    Ma, LJ
    Chen, DJ
    Naqvi, S
    Shepard, D
    Sarfaraz, M
    Holmes, TW
    Suntharalingam, M
    Mansfield, CM
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (02): : 453 - 463