An improved electrical impedance myography (EIM) tongue array for use in clinical trials

被引:15
作者
McIlduff, Courtney [1 ]
Yim, Sung [1 ]
Pacheck, Adam [1 ]
Geisbush, Tom [1 ]
Mijailovic, Aleksandar [1 ]
Rutkove, Seward B. [1 ]
机构
[1] Harvard Univ, Sch Med, Beth Israel Deaconess Med Ctr, Dept Neurol, Boston, MA 02215 USA
基金
美国国家卫生研究院;
关键词
Electrical impedance myography; Tongue; Bulbar; Reliability; Amyotrophic lateral sclerosis; Biomarker; AMYOTROPHIC-LATERAL-SCLEROSIS; ORAL PERFORMANCE INSTRUMENT; HAND STRENGTH; RELIABILITY; BIOMARKERS; DISEASE; ADULTS; PAIN;
D O I
10.1016/j.clinph.2015.06.021
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objectives: Electrical impedance myography (EIM) measurements of the tongue could provide valuable information about bulbar dysfunction in amyotrophic lateral sclerosis (ALS). A prototype tongue depressor EIM array produced gag reflexes. The objectives of this study were to determine the reliability, mean phase values, and tolerability of tongue EIM measurements using a smaller electrode array. Methods: Tongue EIM measurements were performed in a total of 31 healthy individuals and four neuromuscular patients with lingual abnormalities. Reliability was assessed by calculating the intraclass correlation coefficient (ICC) and percent difference in addition to performing Bland-Altman analyses. Standard descriptive statistics, including results of a Mann-Whitney test, were also determined. Results: At the 50 kHz frequency, the ICCs for intra- and inter-rater reliability were 0.76 with 5.17% difference and 0.78 with 5.34% difference respectively. The mean EIM phase values of healthy participants (11.61 degrees +/- 1.00 degrees) and patients (9.87 degrees +/- 1.28 degrees) were significantly different (p = 0.0051). None of the participants experienced gag reflexes or discomfort. Conclusions: The small tongue array provided good inter- and intra-rater reliability, could preliminarily distinguish between healthy and diseased muscle, and was well-tolerated. Significance: Biomarker information about tongue health could be more comfortably obtained with a smaller EIM array. (C) 2015 International Federation of Clinical Neurophysiology. Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:932 / 935
页数:4
相关论文
共 16 条
[1]   Reliability of measurements of tongue and hand strength and endurance using the Iowa Oral Performance Instrument with elderly adults [J].
Adams, Valerie ;
Mathisen, Bernice ;
Baines, Surinder ;
Lazarus, Cathy ;
Callister, Robin .
DISABILITY AND REHABILITATION, 2015, 37 (05) :389-395
[2]   Reliability of Measurements of Tongue and Hand Strength and Endurance Using the Iowa Oral Performance Instrument with Healthy Adults [J].
Adams, Valerie ;
Mathisen, Bernice ;
Baines, Surinder ;
Lazarus, Cathy ;
Callister, Robin .
DYSPHAGIA, 2014, 29 (01) :83-95
[3]   Assessing neuromuscular disease with multifrequency electrical impedance myography [J].
Esper, Gregory J. ;
Shiffman, Carl A. ;
Aaron, Ronald ;
Lee, Kyungmouk S. ;
Rutkove, Seward B. .
MUSCLE & NERVE, 2006, 34 (05) :595-602
[4]  
GANS BM, 1977, ARCH PHYS MED REHAB, V58, P13
[5]   NATURAL-HISTORY OF AMYOTROPHIC-LATERAL-SCLEROSIS IN A DATABASE POPULATION - VALIDATION OF A SCORING SYSTEM AND A MODEL FOR SURVIVAL PREDICTION [J].
HAVERKAMP, LJ ;
APPEL, V ;
APPEL, SH .
BRAIN, 1995, 118 :707-719
[6]   EMG related anxiety and pain: A prospective study [J].
Jan, MMS ;
Schwartz, M ;
Benstead, TJ .
CANADIAN JOURNAL OF NEUROLOGICAL SCIENCES, 1999, 26 (04) :294-297
[7]   The electromyographic diagnosis of amyotrophic lateral sclerosis: Does the evidence support the El Escorial criteria? [J].
Makki, Achraf A. ;
Benatar, Michael .
MUSCLE & NERVE, 2007, 35 (05) :614-619
[8]   Electrical impedance myography to assess outcome in amyotrophic lateral sclerosis clinical trials [J].
Rutkove, Seward B. ;
Zhang, Hui ;
Schoenfeld, David A. ;
Raynor, Elizabeth M. ;
Shefner, Jeremy M. ;
Cudkowicz, Merit E. ;
Chin, Anne B. ;
Aaron, Ronald ;
Shiffman, Carl A. .
CLINICAL NEUROPHYSIOLOGY, 2007, 118 (11) :2413-2418
[9]   ELECTRICAL IMPEDANCE MYOGRAPHY CORRELATES WITH STANDARD MEASURES OF ALS SEVERITY [J].
Rutkove, Seward B. ;
Caress, James B. ;
Cartwright, Michael S. ;
Burns, Ted M. ;
Warder, Judy ;
David, William S. ;
Goyal, Namita ;
Maragakis, Nicholas J. ;
Benatar, Michael ;
Sharma, Khema R. ;
Narayanaswami, Pushpa ;
Raynor, Elizabeth M. ;
Watson, Mary Lou ;
Shefner, Jeremy M. .
MUSCLE & NERVE, 2014, 49 (03) :441-443
[10]   ELECTRICAL IMPEDANCE MYOGRAPHY: BACKGROUND, CURRENT STATE, AND FUTURE DIRECTIONS [J].
Rutkove, Seward B. .
MUSCLE & NERVE, 2009, 40 (06) :936-946