Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy

被引:7
作者
Murazaki, Hiroo [1 ]
Fukunaga, Junichi [1 ]
Hirose, Taka-aki [1 ]
Funatsu, Naomi [1 ]
Matsumoto, Ryoji [1 ]
Hidaka, Kyohei [1 ]
Nagamine, Shuji [1 ]
Nakanishi, Daiki [1 ]
Kato, Toyoyuki [1 ]
机构
[1] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, 3-1-1 Maidashi, Fukuoka, Fukuoka 8128582, Japan
关键词
Metal artifact reduction; Single-energy metal artifact reduction; Radiation treatment planning; Dose distribution; ANISOTROPIC ANALYTICAL ALGORITHM; ACUROS XB; DOSE CALCULATION; CT; QUALITY; IMPACT; SIMULATION; AAA;
D O I
10.1007/s12194-019-00517-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aimed to evaluate the performance of a single-energy metal artifact reduction (SEMAR) algorithm for radiation therapy treatment using phantom cases with metal inserts, assess improvements in computed tomography (CT) number accuracy, and investigate its effects on treatment planning dosimetry. A standard electron density phantom was scanned with and without metal inserts. The numbers of tissue-equivalent materials on both uncorrected and SEMAR-corrected CT images were compared. Treatment planning accuracy was evaluated by comparing dose distributions computed using true density images (without metal inserts), uncorrected images (with metal inserts), and SEMAR-corrected images (with metal inserts) using three-dimensional gamma analysis. The numbers of the true density and uncorrected and SEMAR-corrected CT images in a muscle plug with unilateral inserts were 25.9 HU, - 281.8 HU, and 26.1 HU, respectively. A similar tendency was obtained for other tissue-equivalent materials, and the numbers on CT images were improved with the SEMAR algorithm. In cases involving 1 portal irradiation, 10-MV X-ray, and the Acuros XB algorithm, the pass ratio between the true density and uncorrected images was 89.89%, while that between the true density and SEMAR-corrected images was 95.03%. Improvements in dose distribution were evident using the SEMAR algorithm. Similar trends were found for different irradiation methods and dose calculation algorithms. The SEMAR algorithm can significantly reduce metal artifacts on CT images used for radiation treatment planning. This aspect influenced dosimetry in the region of the artifact and dose distribution was significantly improved with use of the SEMAR-corrected images.
引用
收藏
页码:268 / 276
页数:9
相关论文
共 50 条
  • [31] Frequency split metal artifact reduction (FSMAR) in computed tomography
    Meyer, Esther
    Raupach, Rainer
    Lell, Michael
    Schmidt, Bernhard
    Kachelriess, Marc
    MEDICAL PHYSICS, 2012, 39 (04) : 1904 - 1916
  • [32] Value of virtual monochromatic spectral images with metal artifact reduction algorithm in dual-energy computed tomography-guided microcoil localization of pulmonary nodules
    Liu, Zhuo
    Zhang, Zhuolu
    Chen, Chen
    Hong, Nan
    MEDICINE, 2018, 97 (29)
  • [33] Single-Energy Retrospective Metal Artifact Reduction Using Adaptive Thresholding for Metal Implants in the Abdomen and Pelvis
    Dommaraju, Sujithraj
    Nakhaei, Masoud
    Zhang, Da
    Camacho, Andres
    Boos, Johannes
    Appel, Elisabeth
    Brook, Olga R.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2020, 44 (03) : 443 - 449
  • [34] Combined application of single-energy metal artifact reduction and reconstruction techniques in patients with Cochlear implants
    Wei, Fanqin
    Li, Jiahui
    Zhou, Chunxiang
    Li, Yun
    Wang, Xianren
    Huang, Bixue
    Sun, Qiyang
    Xiong, Guanxia
    JOURNAL OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2020, 49 (01)
  • [35] Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation
    Bamberg, Fabian
    Dierks, Alexander
    Nikolaou, Konstantin
    Reiser, Maximilian F.
    Becker, Christoph R.
    Johnson, Thorsten R. C.
    EUROPEAN RADIOLOGY, 2011, 21 (07) : 1424 - 1429
  • [36] Single-energy metal artefact reduction with CT for carbon-ion radiation therapy treatment planning
    Miki, Kentaro
    Mori, Shinichiro
    Hasegawa, Azusa
    Naganawa, Kensuke
    Koto, Masashi
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1062)
  • [37] Effectiveness of single-energy metal artifact reduction algorithm in CT: application to contrast-enhanced CT with EVAR and coil embolization
    Mochizuki, Junji
    Oyama-Manabe, Noriko
    Kato, Fumi
    Takahashi, Hideki
    Manabe, Osamu
    Sawada, Akihiro
    Okamura, Homare
    Nakano, Mitsunori
    Yamaguchi, Atsushi
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2025, 41 (02) : 369 - 376
  • [38] Combined Assessment of Pulmonary Ventilation and Perfusion with Single-Energy Computed Tomography and Image Processing
    Fujita, Yukio
    Kent, Michael
    Wisner, Erik
    Johnson, Lynelle
    Stern, Joshua
    Qi, Lihong
    Boone, John
    Yamamoto, Tokihiro
    ACADEMIC RADIOLOGY, 2021, 28 (05) : 636 - 646
  • [39] Gradient image smoothing for metal artifact reduction (GISMAR) in computed tomography
    Rodriguez-Gallo, Yakdiel
    Orozco-Morales, Ruben
    Perez-Diaz, Marlen
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (03):
  • [40] Model Image-Based Metal Artifact Reduction for Computed Tomography
    Luzhbin, Dmytro
    Wu, Jay
    JOURNAL OF DIGITAL IMAGING, 2020, 33 (01) : 71 - 82