Dosimetric assessment of a single-energy metal artifact reduction algorithm for computed tomography images in radiation therapy

被引:7
|
作者
Murazaki, Hiroo [1 ]
Fukunaga, Junichi [1 ]
Hirose, Taka-aki [1 ]
Funatsu, Naomi [1 ]
Matsumoto, Ryoji [1 ]
Hidaka, Kyohei [1 ]
Nagamine, Shuji [1 ]
Nakanishi, Daiki [1 ]
Kato, Toyoyuki [1 ]
机构
[1] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, 3-1-1 Maidashi, Fukuoka, Fukuoka 8128582, Japan
关键词
Metal artifact reduction; Single-energy metal artifact reduction; Radiation treatment planning; Dose distribution; ANISOTROPIC ANALYTICAL ALGORITHM; ACUROS XB; DOSE CALCULATION; CT; QUALITY; IMPACT; SIMULATION; AAA;
D O I
10.1007/s12194-019-00517-7
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aimed to evaluate the performance of a single-energy metal artifact reduction (SEMAR) algorithm for radiation therapy treatment using phantom cases with metal inserts, assess improvements in computed tomography (CT) number accuracy, and investigate its effects on treatment planning dosimetry. A standard electron density phantom was scanned with and without metal inserts. The numbers of tissue-equivalent materials on both uncorrected and SEMAR-corrected CT images were compared. Treatment planning accuracy was evaluated by comparing dose distributions computed using true density images (without metal inserts), uncorrected images (with metal inserts), and SEMAR-corrected images (with metal inserts) using three-dimensional gamma analysis. The numbers of the true density and uncorrected and SEMAR-corrected CT images in a muscle plug with unilateral inserts were 25.9 HU, - 281.8 HU, and 26.1 HU, respectively. A similar tendency was obtained for other tissue-equivalent materials, and the numbers on CT images were improved with the SEMAR algorithm. In cases involving 1 portal irradiation, 10-MV X-ray, and the Acuros XB algorithm, the pass ratio between the true density and uncorrected images was 89.89%, while that between the true density and SEMAR-corrected images was 95.03%. Improvements in dose distribution were evident using the SEMAR algorithm. Similar trends were found for different irradiation methods and dose calculation algorithms. The SEMAR algorithm can significantly reduce metal artifacts on CT images used for radiation treatment planning. This aspect influenced dosimetry in the region of the artifact and dose distribution was significantly improved with use of the SEMAR-corrected images.
引用
收藏
页码:268 / 276
页数:9
相关论文
共 50 条
  • [21] Metal Artifact Reduction for Radiation Therapy: A Simulation Study
    Jin, Yannan
    Giantsoudi, Drosoula
    Fu, Lin
    Verburg, Joost
    Gjesteby, Lars
    Wang, Ge
    Paganetti, Harald
    De Man, Bruno
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [22] Dual-Energy Computed Tomography: Is There a Penalty in Image Quality and Radiation Dose Compared With Single-Energy Computed Tomography?
    Thomas, Christoph
    Ketelsen, Dominik
    Tsiflikas, Ilias
    Reimann, Anja
    Brodoefel, Harald
    Claussen, Claus D.
    Heuschmid, Martin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2010, 34 (02) : 309 - 315
  • [23] Customized Total Variation Algorithm for Metal Artifact Reduction in Computed Tomography
    Deng, Ziheng
    Zhou, Yufu
    Zhang, Weikang
    Lin, Zefan
    Zhao, Jun
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 3479 - 3482
  • [24] Dual-energy-based metal segmentation for metal artifact reduction in dental computed tomography
    Hegazy, Mohamed A. A.
    Eldib, Mohamed Elsayed
    Hernandez, Daniel
    Cho, Myung Hye
    Cho, Min Hyoung
    Lee, Soo Yeol
    MEDICAL PHYSICS, 2018, 45 (02) : 714 - 724
  • [25] Comparison of dual-energy computed tomography (DECT) polychromatic and monochromatic images with and without iterative metal artifact reduction algorithm in patients with dental implants
    Huang, Shiqi
    Liang, Yuhong
    Yao, Xinqun
    Qin, Xiangyun
    He, Changguang
    Luo, Li'an
    Huang, Lveqin
    Lv, Yaping
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (07) : 4688 - 4702
  • [26] Influence of field of view size and reconstruction methods on single-energy metal artifact reduction: a phantom study
    Tsuboi, Kunihito
    Osaki, Naoko
    Ohtani, Yuki
    Tanikawa, Ken
    Kaneko, Masanori
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (02) : 637 - 642
  • [27] Reduction of metal artifacts from knee tumor prostheses on CT images: value of the single energy metal artifact reduction (SEMAR) algorithm
    Zhang, Fang-Ling
    Li, Ruo-Cheng
    Zhang, Xiao-Ling
    Zhang, Zhao-Hui
    Ma, Ling
    Ding, Lei
    BMC CANCER, 2021, 21 (01)
  • [28] The application of metal artifact reduction methods on computed tomography scans for radiotherapy applications: A literature review
    Puvanasunthararajah, Sathyathas
    Fontanarosa, Davide
    Wille, Marie-Luise
    Camps, Saskia M.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (06): : 198 - 223
  • [29] Frequency split metal artifact reduction (FSMAR) in computed tomography
    Meyer, Esther
    Raupach, Rainer
    Lell, Michael
    Schmidt, Bernhard
    Kachelriess, Marc
    MEDICAL PHYSICS, 2012, 39 (04) : 1904 - 1916
  • [30] Utility of second-generation single-energy metal artifact reduction in helical lung computed tomography for patients with pulmonary arteriovenous malformation after coil embolization
    Yudai Asano
    Akihiro Tada
    Takayoshi Shinya
    Yoshihisa Masaoka
    Toshihiro Iguchi
    Shuhei Sato
    Susumu Kanazawa
    Japanese Journal of Radiology, 2018, 36 : 285 - 294