SPECT Imaging of 2-D and 3-D Distributed Sources with Near-Field Coded Aperture Collimation: Computer Simulation and Real Data Validation

被引:11
作者
Mu, Zhiping [1 ]
Dobrucki, Lawrence W. [1 ]
Liu, Yi-Hwa [1 ]
机构
[1] Yale Univ, Sect Cardiovasc Med, Dept Internal Med, Sch Med, New Haven, CT 06520 USA
关键词
Near field; Coded aperture; Single-photon emission computerized tomography (SPECT); Maximum likelihood expectation maximization (MLEM); Ordered subset expectation maximization (OSEM); RECONSTRUCTION; ANGIOGENESIS; ALPHA(V)BETA(3); RESOLUTION; REDUCTION; APOPTOSIS; INTEGRIN; ARRAYS;
D O I
10.1007/s40846-016-0111-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The imaging of distributed sources with near-field coded aperture (CA) remains extremely challenging and is broadly considered unsuitable for single-photon emission computerized tomography (SPECT). This study proposes a novel CA SPECT reconstruction approach and evaluates the feasibilities of imaging and reconstructing distributed hot sources and cold lesions using near-field CA collimation and iterative image reconstruction. Computer simulations were designed to compare CA and pinhole collimations in two-dimensional radionuclide imaging. Digital phantoms were created and CA images of the phantoms were reconstructed using maximum likelihood expectation maximization (MLEM). Errors and the contrast-to-noise ratio (CNR) were calculated and image resolution was evaluated. An ex vivo rat heart with myocardial infarction was imaged using a micro-SPECT system equipped with a custom-made CA module and a commercial 5-pinhole collimator. Rat CA images were reconstructed via the three-dimensional (3-D) MLEM algorithm developed for CA SPECT with and without correction for a large projection angle, and 5-pinhole images were reconstructed using the commercial software provided by the SPECT system. Phantom images of CA were markedly improved in terms of image quality, quantitative root-mean-squared error, and CNR, as compared to pinhole images. CA and pinhole images yielded similar image resolution, while CA collimation resulted in fewer noise artifacts. CA and pinhole images of the rat heart were well reconstructed and the myocardial perfusion defects could be clearly discerned from 3-D CA and 5-pinhole SPECT images, whereas 5-pinhole SPECT images suffered from severe noise artifacts. Image contrast of CA SPECT was further improved after correction for the large projection angle used in the rat heart imaging. The computer simulations and small-animal imaging study presented herein indicate that the proposed 3-D CA SPECT imaging and reconstruction approaches worked reasonably well, demonstrating the feasibilities of achieving high sensitivity and high resolution SPECT using near-field CA collimation.
引用
收藏
页码:32 / 43
页数:12
相关论文
共 23 条
  • [1] Accorsi R, 2001, IEEE T NUCL SCI, V48, P2411, DOI 10.1109/23.983251
  • [2] Near-field artifact reduction in planar coded aperture imaging
    Accorsi, R
    Lanza, RC
    [J]. APPLIED OPTICS, 2001, 40 (26) : 4697 - 4705
  • [3] Optimal coded aperture patterns for improved SNR in nuclear medicine imaging
    Accorsi, R
    Gasparini, F
    Lanza, RC
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 474 (03) : 273 - 284
  • [4] REQUIREMENT OF VASCULAR INTEGRIN ALPHA(V)BETA(3) FOR ANGIOGENESIS
    BROOKS, PC
    CLARK, RAF
    CHERESH, DA
    [J]. SCIENCE, 1994, 264 (5158) : 569 - 571
  • [5] TOMOGRAPHICAL IMAGING USING UNIFORMLY REDUNDANT ARRAYS
    CANNON, TM
    FENIMORE, EE
    [J]. APPLIED OPTICS, 1979, 18 (07): : 1052 - 1057
  • [6] Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging - a simulation study
    Cao, ZX
    Bal, G
    Accorsi, R
    Acton, PD
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19) : 4609 - 4624
  • [7] Serial Noninvasive Targeted Imaging of Peripheral Angiogenesis: Validation and Application of a Semiautomated Quantitative Approach
    Dobrucki, Lawrence W.
    Dione, Donald P.
    Kalinowski, Leszek
    Dione, Donna
    Mendizabal, Marivi
    Yu, Jun
    Papademetris, Xenophon
    Sessa, William C.
    Sinusas, Albert J.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (08) : 1356 - 1363
  • [8] UNIFORMLY REDUNDANT ARRAYS - DIGITAL RECONSTRUCTION METHODS
    FENIMORE, EE
    CANNON, TM
    [J]. APPLIED OPTICS, 1981, 20 (10): : 1858 - 1864
  • [9] CODED APERTURE IMAGING - PREDICTED PERFORMANCE OF UNIFORMLY REDUNDANT ARRAYS
    FENIMORE, EE
    [J]. APPLIED OPTICS, 1978, 17 (22): : 3562 - 3570
  • [10] NEW FAMILY OF BINARY ARRAYS FOR CODED APERTURE IMAGING
    GOTTESMAN, SR
    FENIMORE, EE
    [J]. APPLIED OPTICS, 1989, 28 (20): : 4344 - 4352