Prospects for quantitative two-dimensional radiochromic film dosimetry for low dose-rate brachytherapy sources

被引:9
作者
Le, Yi
Ali, Imad
Dempsey, James F.
Williamson, Jeffrey F.
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Richmond, VA 23298 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
[3] Univ Florida, Dept Radiat Oncol, Gainesville, FL 32610 USA
关键词
dosimetry; low dose-rate; brachytherapy; Monte Carlo; radiochromic film (RCF);
D O I
10.1118/1.2390546
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiochromic film (RCF) has been shown to be a precise and accurate two-dimensional dosimeter for acute exposure radiation fields. However, "temporal history" mismatch between calibration and brachytherapy films due to RCF dose-rate effects could introduce potentially large uncertainties in low dose-rate (LDR) brachytherapy absolute dose measurement. This article presents a quantitative evaluation of the precision and accuracy of a laser scanner-based RCF-dosimetry system and the effect of the temporal history mismatch in LDR absolute dose measurement. MD-55-2 RCF was used to measure absolute dose for a low dose-rate Cs-137 brachytherapy source using both single-and double-exposure techniques. Dose-measurement accuracy was evaluated by comparing RCF to Monte Carlo photon-transport simulation. The temporal history mismatch effect was investigated by examining dependence of RCF accuracy on irradiation-to-densitometry time interval. The predictions of the empirical cumulative dose superposition model (CDSM) were compared with measurements. For the double-exposure technique, the agreement between measurement and Monte Carlo simulation was better than 4% in the 3-60 Gy dose range with measurement precisions (coverage factor k=1) of < 2% and < 6% for the doses greater or less than 3 Gy, respectively. The overall uncertainty (k=1) of dose rate/air-kerma strength measurements achievable by this dosimeter system for a spatial resolution of 0.1 mm is less than 4% for doses greater than 5 Gy. The measured temporal history mismatch systematic error is about 1.8% for a 48 h postexposure time when using the double exposure technique and agrees with CDSM's prediction qualitatively. This work demonstrates that the model MD-55-2 RCF detector has the potential to support quantitative dose measurements about LDR brachytherapy sources with precision and accuracy better than that of previously described dosimeters. The impacts of this work on the future use of new type of RCF were also discussed. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:4622 / 4634
页数:13
相关论文
共 40 条
  • [1] Dependence of radiochromic film response kinetics on fractionated doses
    Ali, I
    Williamson, JF
    Costescu, C
    Dempsey, JF
    [J]. APPLIED RADIATION AND ISOTOPES, 2005, 62 (04) : 609 - 617
  • [2] Dependence of radiochromic film optical density post-exposure kinetics on dose and dose fractionation
    Ali, I
    Costescu, C
    Vicic, M
    Dempsey, JF
    Williamson, JF
    [J]. MEDICAL PHYSICS, 2003, 30 (08) : 1958 - 1967
  • [3] AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    Almond, PR
    Biggs, PJ
    Coursey, BM
    Hanson, WF
    Huq, MS
    Nath, R
    Rogers, DWO
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1847 - 1870
  • [4] [Anonymous], 1985, HDB RAD MEAS
  • [5] Absorption spectra variations of EBT radiochromic film from radiation exposure
    Butson, MJ
    Cheung, T
    Yu, PKN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (13) : N135 - N140
  • [6] Post-irradiation colouration of Gafchromic EBT radiochromic film
    Cheung, T
    Butson, MJ
    Yu, PKN
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20) : N281 - N285
  • [7] Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies
    Chiu-Tsao, ST
    Ho, Y
    Shankar, R
    Wang, L
    Harrison, LB
    [J]. MEDICAL PHYSICS, 2005, 32 (11) : 3350 - 3354
  • [8] Accuracy of Monte Carlo photon transport simulation in characterizing brachytherapy dosimeter energy-response artefacts
    Das, RK
    Li, Z
    Perera, H
    Williamson, JF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (06) : 995 - 1006
  • [9] Quantitative optical densitometry with scanning-laser film digitizers
    Dempsey, JF
    Low, DA
    Kirov, AS
    Williamson, JF
    [J]. MEDICAL PHYSICS, 1999, 26 (08) : 1721 - 1731
  • [10] Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions
    Dempsey, JF
    Low, DA
    Mutic, S
    Markman, J
    Kirov, AS
    Nussbaum, GH
    Williamson, JF
    [J]. MEDICAL PHYSICS, 2000, 27 (10) : 2462 - 2475