Comparative Study of Pulsed Versus Continuous High-Intensity Focused Ultrasound Ablation Using in vitro and in vivo Models

被引:8
作者
Wang, Yanhao [1 ,2 ]
Wang, Qi [1 ]
Luo, Yong [1 ]
Jiang, Lu [1 ]
Zeng, Zihuan [1 ]
Gan, Lidan [1 ]
Chen, Jianli [1 ]
Han, Hongfei [1 ]
Zou, Jianzhong [1 ]
机构
[1] Chongqing Med Univ, State Key Lab Ultrasound Engn Med Cofounded Chon, Chongqing Collaborat Innovat Ctr Minimally Invasi, Coll Biomed Engn, Chongqing 400016, Peoples R China
[2] Chongqing Med Univ, Coll Stomatol, Chongqing 401147, Peoples R China
基金
美国国家科学基金会;
关键词
ablation; comparison; duty cycle; high-intensity focused ultrasound; identical doses; CAVITATIONAL ULTRASOUND; NONINVASIVE THERAPY; TISSUE; HIFU; EMBOLIZATION; FEASIBILITY; BACKSCATTER;
D O I
10.1002/jum.15098
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objectives To compare the efficacy of pulsed high-intensity focused ultrasound (PHIFU) versus continuous high-intensity focused ultrasound (CHIFU) ablation at identical doses. Methods Continuous and pulsed HIFU (1200 J) at duty cycles (DCs) of 60% and 20% were examined for their capacity to ablate bovine liver tissue in vitro and rabbit liver tissue in vivo. After ablation, grayscale changes and pathologic characteristics were observed or measured, and the tissue necrosis volume, energy efficiency factor, and average grayscale density were calculated. Results The pulsed mode generated greater liquefaction necrosis. An inconspicuous grayscale change was observed for PHIFU at a DC of 20% in some samples, which appeared as an elliptical cavity. The energy efficiency factor of PHIFU at a DC of 60% was significantly lower than that of CHIFU, as observed both in vitro and in vivo (P < .05). The grayscale value and average grayscale density in response to CHIFU were significantly greater than those in response to PHIFU (60% or 20%; P < .05). Histopathologic analysis revealed liquefaction necrosis in all PHIFU groups. Conclusions At identical doses, compared with CHIFU, a single session of PHIFU can generate liquefaction necrosis and at a higher DC can improve ablation efficiency. This increased efficacy of PHIFU may involve enhancement of tissue destruction by cavitation effects and a reduction in the obstruction effect of endogenous microbubbles through cavitation effects or a more effective diffusion of heat.
引用
收藏
页码:259 / 271
页数:13
相关论文
共 36 条
  • [1] TISSUE ABLATION ACCELERATED BY PERIPHERAL SCANNING MODE WITH HIGH-INTENSITY FOCUSED ULTRASOUND: A STUDY ON ISOLATED PORCINE LIVER PERFUSION
    Bu, Rui
    Yin, Li
    Yang, Han
    Wang, Qi
    Wu, Feng
    Zou, Jian Zhong
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2013, 39 (08) : 1410 - 1419
  • [2] In Vitro and In Vivo Investigation of High-Intensity Focused Ultrasound (HIFU) Hat-Type Ablation Mode
    Dai, Hongya
    Chen, Fei
    Yan, Sijing
    Ding, Xiaoya
    Ma, Dazhao
    Wen, Jing
    Xu, Die
    Zou, Jianzhong
    [J]. MEDICAL SCIENCE MONITOR, 2017, 23 : 3373 - 3382
  • [3] High-intensity focused ultrasound: Current potential and oncologic applications
    Dubinsky, Theodore J.
    Cuevas, Carlos
    Dighe, Manjiri K.
    Kolokythas, Orpheus
    Hwang, Joo Ha
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (01) : 191 - 199
  • [4] High-intensity focused ultrasound for the targeted destruction of uterine tissues: experiences from a pilot study using a mobile HIFU unit
    Fruehauf, Johannes H.
    Back, Walter
    Eiermann, Alexa
    Lang, Michael-Christian
    Pessel, Martin
    Marlinghaus, Ernst
    Melchert, Frank
    Volz-Koester, Stefanie
    Volz, Joachim
    [J]. ARCHIVES OF GYNECOLOGY AND OBSTETRICS, 2008, 277 (02) : 143 - 150
  • [5] A real-time measure of cavitation induced tissue disruption by ultrasound imaging backscatter reduction
    Hall, Timothy L.
    Fowlkes, J. Brian
    Cain, Charles A.
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (03) : 569 - 575
  • [6] Holt RG, 2006, AIP CONF PROC, V829, P13, DOI 10.1063/1.2205429
  • [7] High-intensity focused ultrasound in the treatment of solid tumours
    Kennedy, JE
    [J]. NATURE REVIEWS CANCER, 2005, 5 (04) : 321 - 327
  • [8] Kieran K, 2006, P 6 INT S THER ULTR, P185
  • [9] Extracellular matrix remodeling in vivo for enhancing tumor-targeting efficiency of nanoparticle drug carriers using the pulsed high intensity focused ultrasound
    Lee, Sangmin
    Han, Hyounkoo
    Koo, Heebeom
    Na, Jin Hee
    Yoon, Hong Yeol
    Lee, Kyung Eun
    Lee, Hyukjin
    Kim, Hyuncheol
    Kwon, Ick Chan
    Kim, Kwangmeyung
    [J]. JOURNAL OF CONTROLLED RELEASE, 2017, 263 : 68 - 78
  • [10] [李发琪 Li Faqi], 2005, [中国超声医学杂志, Chinese Journal of Ultrasonic in Medical], V21, P252