A device for 24 hour ambulatory monitoring of abdominal girth using inductive plethysmography

被引:22
作者
Reilly, BP
Bolton, MP
Lewis, MJ
Houghton, LA
Whorwell, PJ
机构
[1] Christie Hosp, Manchester M20 4BX, Lancs, England
[2] Univ S Manchester Hosp, Dept Med, Manchester M20 2LR, Lancs, England
关键词
IBS; irritable bowel syndrome; bloating; plethysmography;
D O I
10.1088/0967-3334/23/4/306
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Inductive plethysmography (IP) sensors and oscillator modules were assessed for their potential use in the ambulatory monitoring of abdominal girth in subjects with irritable bowel syndrome (IBS) in order to objectively quantify their bloating symptoms. A dedicated microprocessor data logger was designed to record over 24 h the frequency output of IP oscillators connected to a belt around the subject's lower abdomen. Posture was also recorded via tilt switches (standing, sitting and lying). The system was separately calibrated by placing the belts around a variable rectangular phantom and measuring the frequency of oscillation. A theoretical geometric model was devised to convert measured frequency into circumference and account for changes caused by variations in shape. Using the calibration factors, it was found that the circumference of a circular phantom could be measured accurately (mean difference 1.27 cm and SD 0.25 cm). The system has been tested over 24 h with 20 volunteers. Movement introduced variations in measured girth larger than those found during periods of non-movement during sleep. We conclude that IP promises to be a useful and quantitative tool suitable for ambulatory monitoring of abdominal girth, a hitherto relatively unexplored symptom of IBS.
引用
收藏
页码:661 / 670
页数:10
相关论文
共 15 条
  • [1] [Anonymous], 1962, INDUCTANCE CALCULATI
  • [2] ESTIMATES OF VENTILATION FROM INDUCTANCE PLETHYSMOGRAPHY IN SLEEPING ASTHMATIC-PATIENTS
    BALLARD, RD
    KELLY, PL
    MARTIN, RJ
    [J]. CHEST, 1988, 93 (01) : 128 - 133
  • [3] Abbreviated method for calculating the inductance of irregular plane polygons of round wire (Part I of paper, "On the calculation of closed aerials")
    Bashenoff, VJ
    [J]. PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1927, 15 (12): : 1013 - 1039
  • [4] RELIABILITY OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY FOR MEASURING TIDAL VOLUME DURING EXERCISE
    CARETTI, DM
    PULLEN, PV
    PREMO, LA
    KUHLMANN, WD
    [J]. AMERICAN INDUSTRIAL HYGIENE ASSOCIATION JOURNAL, 1994, 55 (10): : 918 - 923
  • [5] MEASUREMENT OF SEPARATE VOLUME CHANGES OF RIB CAGE AND ABDOMEN DURING BREATHING
    KONNO, K
    MEAD, J
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1967, 22 (03) : 407 - &
  • [6] Symptoms and visceral perception in patients with pain-predominant irritable bowel syndrome
    Lembo, T
    Naliboff, B
    Munakata, J
    Fullerton, S
    Saba, L
    Tung, S
    Schmulson, M
    Mayer, EA
    [J]. AMERICAN JOURNAL OF GASTROENTEROLOGY, 1999, 94 (05) : 1320 - 1326
  • [7] Ambulatory abdominal inductance plethysmography: towards objective assessment of abdominal distension in irritable bowel syndrome
    Lewis, MJV
    Reilly, B
    Houghton, LA
    Whorwell, PJ
    [J]. GUT, 2001, 48 (02) : 216 - 220
  • [8] TOWARDS POSITIVE DIAGNOSIS OF IRRITABLE BOWEL
    MANNING, AP
    THOMPSON, WG
    HEATON, KW
    MORRIS, AF
    [J]. BRITISH MEDICAL JOURNAL, 1978, 2 (6138) : 653 - 654
  • [9] WHAT DOES INDUCTANCE PLETHYSMOGRAPHY REALLY MEASURE
    MARTINOTLAGARDE, P
    SARTENE, R
    MATHIEU, M
    DURAND, G
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (04) : 1749 - 1756
  • [10] ABDOMINAL DISTENSION IN FEMALE-PATIENTS WITH IRRITABLE-BOWEL-SYNDROME - EXPLORATION OF POSSIBLE MECHANISMS
    MAXTON, DG
    MARTIN, DF
    WHORWELL, PJ
    GODFREY, M
    [J]. GUT, 1991, 32 (06) : 662 - 664