摘要
Purpose: Traditional implantation of the AMS Sphincter 800 Urinary Control System (American Medical Systems, Minnetonka, Minnesota) requires 2 incisions. The cuff is placed via a perineal incision, and the pressure regulating balloon and pump are placed through a separate suprapubic incision. We describe a novel implantation of all the artificial urinary sphincter components using a single upper scrotal incision. The scrotal incision allows excellent access to the proximal bulbar urethra and retropubic and subdartos spaces, and leaves the bulbocavernosus muscle intact. Materials and Methods: A total of 37 patients have undergone artificial urinary sphincter implantation using the new operative technique for revisions or reimplantations of a sphincter previously removed for infection/erosion (12) or as an initial procedure (25). In 9 of the 25 patients and 2 of the 12 dual implantation of a 3-piece penile prosthesis through the same incision was performed. Results: All patients are using the devices. Of the patients 66% are completely dry with no pad use and the remainder use 1 pad for accident prevention. Operative time was reduced due to easier exposure of the urethra and a second incision for placement of the pressure regulating balloon was not necessary. Followup at 1 year shows no difference in complication rate with the single incision technique compared to the traditional method. Conclusions: Artificial urinary sphincter implantation through a single scrotal incision is easier and faster than the traditional 2-incision technique. Success in achieving continence is similar to traditional methods. Long-term followup is necessary to ensure that complications remain low.
引用
收藏
页码:261 / 264
页数:4
相关论文
共 12 条
[1]   USE OF A DOUBLE CUFF AMS800 URINARY SPHINCTER FOR SEVERE STRESS-INCONTINENCE [J].
BRITO, CG ;
MULCAHY, JJ ;
MITCHELL, ME ;
ADAMS, MC .
JOURNAL OF UROLOGY, 1993, 149 (02) :283-285
[2]   Revision rate after artificial urinary sphincter implantation for incontinence after radical prostatectomy: Actuarial analysis [J].
Clemens, JQ ;
Schuster, TG ;
Konnak, JW ;
McGuire, EJ ;
Faerber, GJ .
JOURNAL OF UROLOGY, 2001, 166 (04) :1372-1375
[3]  
Diokno AC, 2000, J UROLOGY, V164, P706
[4]   EXPERIENCE WITH THE ARTIFICIAL URINARY SPHINCTER MODEL-AS800 IN 148 PATIENTS [J].
FISHMAN, IJ ;
SHABSIGH, R ;
SCOTT, FB .
JOURNAL OF UROLOGY, 1989, 141 (02) :307-310
[5]   Artificial urinary sphincter for post-radical prostatectomy urinary incontinence: Long-term subjective results [J].
Gousse, AE ;
Madjar, S ;
Lambert, MM ;
Fishman, IJ .
JOURNAL OF UROLOGY, 2001, 166 (05) :1755-1758
[6]   Transcorporal artificial urinary sphincter cuff placement in cases requiring revision for erosion and urethral atrophy [J].
Guralnick, ML ;
Miller, E ;
Toh, KL ;
Webster, GD .
JOURNAL OF UROLOGY, 2002, 167 (05) :2075-2078
[7]   Postprostatectomy urinary incontinence: The case for artificial urinary sphincter implantation [J].
Montague, DK ;
Angermeier, KW .
UROLOGY, 2000, 55 (01) :2-4
[8]  
Montague DK, 2001, J UROLOGY, V166, P1375
[9]   Reservoir herniation as a complication of three-piece penile prosthesis insertion [J].
Sadeghi-Nejad, H ;
Sharma, A ;
Irwin, RJ ;
Wilson, SK ;
Delk, JR .
UROLOGY, 2001, 57 (01) :142-145
[10]  
Stone NN, 2001, J UROLOGY, V166, P1375