CVD diamonds as thermoluminescent detectors for medical applications

被引:31
作者
Marczewska, B
Olko, P
Nesladek, M
Waligórski, MPR
Kerremans, Y
机构
[1] Inst Nucl Phys, PL-31342 Krakow, Poland
[2] Limburgs Univ Centrum, Inst Mat Res, Krakow, Poland
[3] Ctr Oncol, Krakow, Poland
[4] Sci & Tech Res Ctr Diamond, B-2500 Lier, Belgium
关键词
D O I
10.1093/oxfordjournals.rpd.a006033
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Diamond is believed to be a promising material for medical dosimetry due to its tissue equivalence, mechanical and radiation hardness, and lack of solubility in water or in disinfecting agents. A number of diamond samples, obtained tinder different growth conditions at Limburg University, using the chemical vapour deposition (CVD) technique, was tested as thermoluminescence dosemeters. Their TL glow curve, TL response after doses of gamma rays. fading. and so on were studied at dose levels and for radiation modalities typical for radiotherapy. The investigated CVD diamonds displayed sensitivity comparable with that of MTS-N (Li:Mg,Ti) detectors, signal stability (reproducibility after several readouts) below 10% (1 SD) and no fading was found four days after irradiation. A dedicated CVD diamond plate was grown, cut into 20 detector chips (3 X 3 X 0.5 mm) and used for measuring the dose-depth distribution at different depths in a water phantom, for Co-60 and six MV X ray radiotherapy beams. Due to the sensitivity of diamond to ambient light, it was difficult to achieve reproducibility comparable with that of standard LiF detectors.
引用
收藏
页码:485 / 488
页数:4
相关论文
共 9 条
[1]  
AVILA O, 1995, RADIAT PROT DOSIM, V58, P61
[2]   THERMOLUMINESCENCE EFFICIENCY OF LIFMG,CU,P (MCP-N) DETECTORS TO PHOTONS, BETA-ELECTRONS, ALPHA-PARTICLES AND THERMAL-NEUTRONS [J].
BILSKI, P ;
OLKO, P ;
BURGKHARDT, B ;
PIESCH, E ;
WALIGORSKI, MPR .
RADIATION PROTECTION DOSIMETRY, 1994, 55 (01) :31-38
[3]   Assessment of CVD diamond as a thermoluminescence dosemeter material [J].
Borchi, E ;
Furetta, C ;
Kitis, G ;
Leroy, C ;
Sussmann, RS ;
Whitehead, AJ .
RADIATION PROTECTION DOSIMETRY, 1996, 65 (1-4) :291-295
[4]  
HOPWOOD PJ, 1994, P DIAM C READ 11 13
[5]  
KHRUNOV VS, 1990, RADIAT PROT DOSIM, V33, P155
[6]  
Marczewska B, 2001, PHYS STATUS SOLIDI A, V185, P183, DOI 10.1002/1521-396X(200105)185:1<183::AID-PSSA183>3.0.CO
[7]  
2-N
[8]   An assessment of radiotherapy dosimeters based on CVD grown diamond [J].
Ramkumar, S ;
Buttar, CM ;
Conway, J ;
Whitehead, AJ ;
Sussman, RS ;
Hill, G ;
Walker, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 460 (2-3) :401-411
[9]   APPLICATION OF A NATURAL DIAMOND DETECTOR FOR THE MEASUREMENT OF RELATIVE DOSE DISTRIBUTIONS IN RADIOTHERAPY [J].
VATNITSKY, S ;
JARVINEN, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (01) :173-184