Validation and Compliance of a Home Monitoring Device in Pregnancy: Microlife WatchBP Home

被引:56
作者
Chung, Yealin [1 ]
de Greeff, Annemarie [1 ]
Shennan, Andrew [1 ]
机构
[1] Kings Coll London, Maternal & Fetal Res Unit, London SE1 7EH, England
关键词
Blood pressure; Home monitoring; Preeclampsia; Pregnancy; Validation; BLOOD-PRESSURE-MEASUREMENT; EUROPEAN-SOCIETY; HYPERTENSION RECOMMENDATIONS; INTERNATIONAL-PROTOCOL; ACCURATE; PREECLAMPSIA; TRIALS;
D O I
10.1080/10641950802601286
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective. To assess the accuracy and patient compliance in using a novel home blood pressure monitoring device in high-risk pregnancy. Methods. Device accuracy was assessed according to the British Hypertension Society protocol in 45 pregnant women, including 15 with preeclampsia. Twenty-one high-risk pregnant women used the device in addition to their antenatal care. Results. The device achieved a mean difference +/- SD of 0.4 +/- 7.3/-0.4 +/- 5.5 mmHg (pregnancy) and -2.6 +/- 7.0/0.8 +/- 4.4 mmHg (preeclampsia) for systolic/diastolic pressure. Eighty-one percent of women did at least 6 measurements/day and all women did at least 2 measurements/week. Conclusion. The Microlife WatchBP Home is accurate for use in pregnancy and increases surveillance in compliant patients.
引用
收藏
页码:348 / 359
页数:12
相关论文
共 28 条
[21]  
Shennan AH, 2007, PRE-ECLAMPSIA: ETIOLOGY AND CLINICAL PRACTICE, P258
[22]   Self monitoring of blood pressure at home - Is an important adjunct to clinic measurements [J].
Stergiou, G ;
Mengden, T ;
Padfield, PL ;
Parati, G ;
O'Brien, E .
BMJ-BRITISH MEDICAL JOURNAL, 2004, 329 (7471) :870-871
[23]   The optimal schedule for self-monitoring of blood pressure by patients at home [J].
Stergiou, George S. ;
Parati, Gianfranco .
JOURNAL OF HYPERTENSION, 2007, 25 (10) :1992-1997
[24]   Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol [J].
Stergiou, George S. ;
Giovas, Periklis P. ;
Gkino, Charilaos P. ;
Patouras, John D. .
BLOOD PRESSURE MONITORING, 2007, 12 (03) :185-188
[25]   A tool for reliable self-home blood pressure monitoring designed according to the European Society of Hypertension recommendations: The Microlife WatchBP Home monitor [J].
Stergiou, George S. ;
Jaenecke, Bernd ;
Giovas, Periklis R. ;
Chang, Arron ;
Chung-Yueh, Yen ;
Tan, Ty-Minh .
BLOOD PRESSURE MONITORING, 2007, 12 (02) :127-131
[26]   Self-monitoring of blood pressure at home: how many measurements are needed? [J].
Stergiou, GS ;
Skeva, II ;
Zourbaki, AS ;
Mountokalakis, TD .
JOURNAL OF HYPERTENSION, 1998, 16 (06) :725-731
[27]   Reproducibility of home, ambulatory, and clinic blood pressure: Implications for the design of trials for the assessment of antihypertensive drug efficacy [J].
Stergiou, GS ;
Baibas, NM ;
Gantzarou, AP ;
Skeva, II ;
Kalkana, CB ;
Roussias, LG ;
Mountokalakis, TD .
AMERICAN JOURNAL OF HYPERTENSION, 2002, 15 (02) :101-104
[28]   Patient initiated home blood pressure recordings are accurate in hypertensive pregnant women [J].
Waugh, J ;
Habiba, MA ;
Bosio, P ;
Shennan, A ;
Halligan, AWF .
HYPERTENSION IN PREGNANCY, 2003, 22 (01) :93-97