Vibrations produced during erbium:yttrium-aluminum-garnet laser irradiation

被引:9
作者
Anic, Ivica [1 ]
Miletic, Ivana [1 ]
Jukic Krmek, Silvana [1 ]
Borcic, Josipa [2 ]
Pezelj-Ribaric, Sonja [3 ]
机构
[1] Univ Zagreb, Dept Endodont & Restorat Dent, Sch Dent Med, Gunduliceva 5, Zagreb 10000, Croatia
[2] Univ Rijeka, Dept Oral & Maxillofacial Surg, Sch Med, Rijeka, Croatia
[3] Univ Rijeka, Dept Oral Pathol, Sch Med, Rijeka, Croatia
关键词
Er:YAG laser; Tooth preparation; Piezoelectric accelerometer; Vibration; ER-YAG-LASER; CAVITY PREPARATION; CARIES TREATMENT; ABLATION;
D O I
10.1007/s10103-008-0596-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this study was to evaluate vibrations induced by an erbium:yttrium-aluminum-garnet (Er:YAG) laser in the non-contact mode and compare the vibrations with different pulse durations and energy parameters. The experiment was conducted on an extracted tooth built up in silicone impression material. The vibrations were measured by piezoelectric accelerometer for a super-short pulse (SSP), a very short pulse (VSP), and a short pulse (SP) at a frequency of 5 Hz for 1 s. For VSP and SP, the energy parameters tested were 200 mJ, 300 mJ, and 400 mJ. Measurements were performed 15 times for each individual irradiation energy level. The highest values of vibrations were measured for SP (0.160 +/- 0.04 m/s(2)), and the lowest were measured for VSP mode at the energy parameter 200 mJ (0.05 +/- 0.02 m/s(2)). There was a statistically significant (P < 0.01) difference between the various laser pulse modes (SSP, VSP, SP) at different energy parameters. At energy levels of 300 mJ and 400 mJ, the least amount of vibration during cavity preparations with the non-contact Er:YAG laser was produced by SSP mode.
引用
收藏
页码:697 / 701
页数:5
相关论文
共 26 条
[1]   Comparison between Er:YAG laser and conventional technique for root caries treatment in vitro [J].
Aoki, A ;
Ishikawa, I ;
Yamada, T ;
Otsuki, M ;
Watanabe, H ;
Tagami, J ;
Ando, Y ;
Yamamoto, H .
JOURNAL OF DENTAL RESEARCH, 1998, 77 (06) :1404-1414
[2]   Bonding to Er-YAG-laser-treated dentin [J].
Ceballos, L ;
Toledano, M ;
Osorio, R ;
Tay, FR ;
Marshall, GW .
JOURNAL OF DENTAL RESEARCH, 2002, 81 (02) :119-122
[3]  
Chinelatti MA, 2004, OPER DENT, V29, P430
[4]   Effect of Er:YAG laser energy on the morphology of enamel/adhesive system interface [J].
Delfino, Carina Sincler ;
Souza-Zaroni, Wanessa Christine ;
Corona, Silmara Aparecida Milori ;
Pecora, Jesus Djalma ;
Palma-Dibb, Regina Guenka .
APPLIED SURFACE SCIENCE, 2006, 252 (24) :8476-8481
[5]  
Dostalova T, 1997, J Clin Laser Med Surg, V15, P117
[6]   Er:YAG laser ablation -: Evaluation after two-year long clinical treatment [J].
Dostalova, T ;
Jelinkova, H ;
Hana, K ;
Krejsa, O ;
Hamal, K ;
Kubelka, J ;
Prochazka, S .
LASERS IN DENTISTRY IV, PROCEEDINGS OF, 1998, 3248 :23-32
[7]   Noncontact Er:YAG laser ablation:: Clinical evaluation [J].
Dostalova, T ;
Jelinkova, H ;
Kucerova, H ;
Krejsa, O ;
Hamal, K ;
Kubelka, J ;
Prochazka, S .
JOURNAL OF CLINICAL LASER MEDICINE & SURGERY, 1998, 16 (05) :273-282
[8]   The use of Er:YAG laser for cavity preparation:: An SEM evaluation [J].
Freitas, Patricia Moreira ;
Navarro, Ricardo Scarparo ;
Barros, Juliana Almeida ;
Eduardo, Carlos De Paula .
MICROSCOPY RESEARCH AND TECHNIQUE, 2007, 70 (09) :803-808
[9]   Intrapulpal temperature during preparation with the Er:YAG laser compared to the conventional burr:: An in vitro study [J].
Glockner, K ;
Rumpler, J ;
Ebeleseder, K ;
Stadtler, P .
JOURNAL OF CLINICAL LASER MEDICINE & SURGERY, 1998, 16 (03) :153-157
[10]  
Griffin MJ., 1990, Handbook of Human Vibration