Analysis of Retrieved Agility™ Total Ankle Arthroplasty Systems

被引:18
作者
Vaupel, Zachary [1 ]
Baker, Erin A. [1 ]
Baker, Kevin C. [1 ]
Kurdziel, Michael D. [1 ]
Fortin, Paul T. [1 ]
机构
[1] William Beaumont Hosp, Dept Orthopaed Res, Res Inst, Royal Oak, MI 48073 USA
关键词
Total Ankle Arthroplasty; Agility (TM) Total Ankle; Osteolysis; Subsidence; Wear; Damage; COMPONENTS; STRESSES; SURFACES; WEAR;
D O I
10.3113/FAI.2009.0815
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background. First generation total ankle arthroplasty (TAA) systems showed high rates of failure. The Agility (TM) (DePuy, Warsaw, IN) TAA system, a second generation design, had improved outcomes; however, implant failure due to loosening of the metallic components persisted. The purpose of this study was to analyze the damage modes and radiographic mode(s) of failure observed in retrieved Agility (TM) TAA. Materials and Methods: Ten devices were collected and each component was analyzed for common damage modes using microscopy. Clinical damage was analyzed with postoperative implant and preoperative revision procedure radiographs. Results: Analyses revealed damage/wear to retrieved components, including abrasion, dishing, and pitting. These third-body wear particles may be a precursor to wear debris induced osteolysis which could cause component loosening. Seven TAA systems were removed due to loosening or subsidence, suggesting component damage/wear may lead to clinically observed component loosening. Retrieval analysis indicated the polyethylene experiences edge loading, resulting in increased contact stresses to polyethylene in the primary articulation region and wear. Conclusion: Since poor clinical outcomes have been associated with component instability and osteolysis, analyzing retrieved components wear and damage may be an important step toward improving implant design, thereby decreasing wear debris induced osteolysis and improving clinical outcomes.
引用
收藏
页码:815 / 823
页数:9
相关论文
共 19 条
[1]   ANKLE ARTHRODESIS - A LONG-TERM STUDY [J].
ABDO, RV ;
WASILEWSKI, SA .
FOOT & ANKLE, 1992, 13 (06) :307-312
[2]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[3]  
BENJAMIN A, 1980, PLASTIC RECONSTRUCTI, V96, P206
[4]   Kinematic analysis of the agility total ankle during gait [J].
Conti, Stephen ;
Lalonde, Karl-Andre ;
Martin, RobRoy .
FOOT & ANKLE INTERNATIONAL, 2006, 27 (11) :980-984
[5]  
DAVIDSON JA, 1987, J BIOMED MATER RES-A, V21, P261
[6]  
DEORIO JK, 2006, NE FLORIDA MED, V57, P9
[7]  
Easley Mark E, 2002, J Am Acad Orthop Surg, V10, P157
[8]   POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[9]   Challenges in total ankle arthroplasty [J].
Gill, LH .
FOOT & ANKLE INTERNATIONAL, 2004, 25 (04) :195-207
[10]   Ankle arthroplasty with preoperative coronal plane deformity - Short-term results [J].
Haskell, A ;
Mann, RA .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (424) :98-103