Optimal image reconstruction phase at low and high heart rates in dual-source CT coronary angiography

被引:57
作者
Araoz, Philip A. [1 ]
Kirsch, Jacobo [1 ]
Primak, Andrew N. [1 ]
Braun, Natalie N. [1 ]
Saba, Osama [2 ]
Williamson, Eric E. [1 ]
Harmsen, W. Scott [1 ]
Mandrekar, Jayawant N. [1 ]
McCollough, Cynthia H. [1 ]
机构
[1] Mayo Clin, Rochester, MN 55905 USA
[2] Siemens Med Solut, Forcheim, Germany
关键词
Tomography; X-ray computed; Reproducibility of results; Radiographic image enhancement; Dual-source CT; CT coronary angiography; artifact; Motion artifact; Image quality; ECG gating; SOURCE COMPUTED-TOMOGRAPHY; RATE-VARIABILITY; QUALITY; INTERVALS; WINDOWS; MOTION; SPEED;
D O I
10.1007/s10554-009-9489-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to determine the cardiac phase having the highest coronary sharpness for low and high heart rate patients scanned with dual source CT (DSCT) and to compare coronary image sharpness over different cardiac phases. DSCT coronary CT scans for 30 low heart rate (a parts per thousand currency sign70 beats per minute- bpm) and 30 high heart rate (> 70 bpm) patients were reconstructed into different cardiac phases, starting at 30% and increasing at 5% increments until 70%. A blinded observer graded image sharpness per coronary segment, from which sharpness scores were produced for the right (RCA), left main (LM), left anterior descending (LAD), and circumflex (Cx) coronary arteries. For each coronary artery, the phase with maximal image sharpness was identified with repeated measures analysis of variance. Comparison of coronary sharpness between low and high heart rate patients was made using generalized estimating equations. For low heart rates the highest sharpness scores for all four vessels (RCA, LM, LAD, and Cx) were at the 65 or 70% phase, which are end-diastolic cardiac phases. For high heart rates the highest sharpness scores were between the 35 and 45% phases, which are end-systolic phases. Low heart rate patients had higher coronary sharpness at most cardiac phases; however, patients with high heart rates had higher coronary sharpness in the 45% phase for all four vessels (P < 0.0001). Using DSCT scanning, optimal image sharpness is obtained in end-diastole at low heart rates and in end-systole in high heart rates.
引用
收藏
页码:837 / 845
页数:9
相关论文
共 13 条
  • [1] Austen W G, 1975, Circulation, V51, P5
  • [2] First performance evaluation of a dual-source CT (DSCT) system
    Flohr, TG
    McCollough, CH
    Bruder, H
    Petersilka, M
    Gruber, K
    Süss, C
    Grasruck, M
    Stierstorfer, K
    Krauss, B
    Raupach, R
    Primak, AN
    Küttner, A
    Achenbach, S
    Becker, C
    Kopp, A
    Ohnesorge, BM
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (02) : 256 - 268
  • [3] Usefulness of noninvasive cardiac imaging using dual-source computed tomography in an unselected population with high prevalence of coronary artery disease
    Heuschmid, Martin
    Burgstahler, Christof
    Reimann, Ania
    Brodoefel, Harald
    Mysal, Ines
    Haeberle, Ellen
    Tsiflikas, Ilijas
    Claussen, Claus D.
    Kopp, Andreas F.
    Schroeder, Stephen
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2007, 100 (04) : 587 - 592
  • [4] Coronary artery motion and cardiac phases: Dependency on heart rate - Implications for CT image reconstruction
    Husmann, Lars
    Leschka, Sebastian
    Desbiolles, Lotus
    Schepis, Tiziano
    Gaemperli, Oliver
    Seifert, Burkhardt
    Cattin, Philippe
    Frauenfelder, Thomas
    Flohr, Thomas G.
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    [J]. RADIOLOGY, 2007, 245 (02) : 567 - 576
  • [5] Image quality and reconstruction intervals of dual-source CT coronary angiography - Recommendations for ECG-pulsing windowing
    Leschka, Sebastian
    Scheffel, Hans
    Desbiolles, Lotus
    Plass, Andre
    Gaemperli, Oliver
    Valenta, Ines
    Husmann, Lars
    Flohr, Thomas G.
    Genoni, Michele
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    [J]. INVESTIGATIVE RADIOLOGY, 2007, 42 (08) : 543 - 549
  • [6] Noninvasive coronary angiography with 64-section CT: Effect of average heart rate and heart rate variability on image quality
    Leschka, Sebastian
    Wildermuth, Simon
    Boehm, Thomas
    Desbiolles, Lotus
    Husmann, Lars
    Plass, Andre
    Koepfli, Pascal
    Schepis, Tiziano
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    [J]. RADIOLOGY, 2006, 241 (02) : 378 - 385
  • [7] Optimal image reconstruction intervals for non-invasive coronary angiography with 64-slice CT
    Leschka, Sebastian
    Husmann, Lars
    Desbiolles, Lotus M.
    Gaemperli, Oliver
    Schepis, Tiziano
    Koepfli, Pascal
    Boehm, Thomas
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    [J]. EUROPEAN RADIOLOGY, 2006, 16 (09) : 1964 - 1972
  • [8] Dual-source CT coronary angiography: Image quality, mean heart rate, and heart rate variability
    Matt, David
    Scheffel, Hans
    Leschka, Sebastian
    Flohr, Thomas G.
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (03) : 567 - 573
  • [9] Optimal Systolic and Diastolic reconstruction windows for coronary CT angiography using dual-source CT
    Seifarth, Harald
    Wienbeck, Susanne
    Puesken, Michael
    Juergens, Kai-Uwe
    Maintz, David
    Vahlhaus, Christian
    Heindel, Walter
    Fischbach, Roman
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (06) : 1317 - 1323
  • [10] Radiation dose estimates in dual-source computed tomography coronary angiography
    Stolzmann, Paul
    Scheffel, Hans
    Schertler, Thomas
    Frauenfelder, Thomas
    Leschka, Sebastian
    Husmann, Lars
    Flohr, Thomas G.
    Marincek, Borut
    Kaufmann, Philipp A.
    Alkadhi, Hatem
    [J]. EUROPEAN RADIOLOGY, 2008, 18 (03) : 592 - 599