Verification of air-kerma strength of 125I seed for permanent prostate implants in Japan

被引:6
作者
Sumida, Iori [1 ]
Koizumi, Masahiko [1 ]
Takahashi, Yutaka [1 ]
Ogata, Toshiyuki [1 ]
Akino, Yuichi [1 ]
Isohashi, Fumiaki [1 ]
Konishi, Koji [1 ]
Yoshioka, Yasuo [1 ]
Inoue, Takehiro [1 ]
机构
[1] Osaka Univ, Dept Radiat Oncol, Grad Sch Med, Suita, Osaka 5650871, Japan
关键词
Brachytherapy; Single-seed assay; Permanent implant; Source calibration; BRACHYTHERAPY;
D O I
10.1007/s10147-009-0914-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background To assure the physical quality of brachytherapy, we investigated the difference between measured and manufacturer's stated source strengths in a single model SourceTech Medical (STM)1251 I-125 seed. Methods. A well-type ionization chamber with a single-seed holder was used to measure the source strength of 2412 I-125 seeds before implant in 34 patients. The air-kerma strength was 0.450 U for all cases. The mean source strength for each patient was measured and compared with the manufacturer's stated value. The deviation from the measured value was compared with the tolerance range of the American Association of Physicists in Medicine (AAPM) TG-56 report's recommendation. Results. The measured source strength was higher than the manufacturer's stated value, with a median difference of 1% (range,-2% to 5%). Sixteen of the total of 2412 seeds (0.7%) were more than 5% different from the manufacturer's stated value. The median SD from the mean value was 2.2% (range, 1.1% to 2.5%) for all patients. Conclusion. This is the first report of a single-seed assay performed for the model STM1251 I-125 seed. In this study the manufacturer's stated strength agreed well with the measured value. Nevertheless, the advisability of performing a single-seed assay at every institution should be considered, by referring to the appropriate regulations; for example, those used in the United States.
引用
收藏
页码:525 / 528
页数:4
相关论文
共 19 条
  • [1] *AM COLL RAD, 2006, ACR PRACT GUID, P909
  • [2] Third-party brachytherapy source calibrations and physicist responsibilities: Report of the AAPM Low Energy Brachytherapy Source Calibration Working Group
    Butler, Wayne M.
    Bice, William S., Jr.
    DeWerd, Larry A.
    Hevezi, James M.
    Huq, M. Saiful
    Ibbott, Geoffrey S.
    Palta, Jatinder R.
    Rivard, Mark J.
    Seuntjens, Jan P.
    Thomadsen, Bruce R.
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 3860 - 3865
  • [3] DEWERD LA, 1997, CATEGORICAL COURSE B, P143
  • [4] DEWERD LA, 1997, SYLLABUS CATEGORICAL, P143
  • [5] DEWERD LA, 1995, AM ASS PHYS MED 1994
  • [6] DeWerd LA, 2005, MED PHYS MG, P153
  • [7] Hanson WF, 1991, P AM COLL MED PHYS S, P255
  • [8] Monte Carlo-aided dosimetry of the Source Tech Medical Model STM1251 I-125 interstitial brachytherapy source
    Kirov, AS
    Williamson, JF
    [J]. MEDICAL PHYSICS, 2001, 28 (05) : 764 - 772
  • [9] COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40
    KUTCHER, GJ
    COIA, L
    GILLIN, M
    HANSON, WF
    LEIBEL, S
    MORTON, RJ
    PALTA, JR
    PURDY, JA
    REINSTEIN, LE
    SVENSSON, GK
    WELLER, M
    WINGFIELD, L
    [J]. MEDICAL PHYSICS, 1994, 21 (04) : 581 - 618
  • [10] Lee PC, 1999, RADIAT ONCOL INVESTI, V7, P374, DOI 10.1002/(SICI)1520-6823(1999)7:6<374::AID-ROI8>3.3.CO