Determination of bencycloquidium bromide in rat plasma by liquid chromatography-electrospray ionization-mass spectrometry

被引:11
作者
Xu, Qin [1 ]
Ding, Li [1 ]
Liu, Wenying [1 ]
Bian, Xiaojie [1 ]
Tang, Wenming [1 ]
机构
[1] China Pharmaceut Univ, Dept Pharmaceut Anal, Nanjing 210009, Peoples R China
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2007年 / 846卷 / 1-2期
关键词
bencycloquidium bromide; LC-MS; pharmacokinetics;
D O I
10.1016/j.jchromb.2006.09.028
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS) assay for the determination of bencycloquidium bromide (BCQB) in rat plasma was firstly developed and validated. After addition of 1-ethyl-bencycloquidium bromide as an internal standard (I.S.), the plasma samples were deproteinized with methanol and the supernatant was assayed by LC-ESI-MS. Chromatographic separation was achieved with a Hanbon Lichrospher 5-C18 column. The mobile phase consisted of methanol-40 mM ammonium acetate buffer-formic acid (75:25:0.25, v/v/v) and delivered at the flow rate of 1.0 ml/min. LC-ESI-MS was carried out on a single quadrupole mass spectrometer using electrospray ionization (ESI) and positive selected-ion monitoring (SIM). Target ions were monitored at [M](+) m/z 330.2 for BCQB and [M](+) m/z 344.2 for I.S. Calibration curve was linear over the range of 3-1500 ng/ml. The lower limit of quantification (LLOQ) was 3.0 ng/ml. The intra- and inter-run relative standard deviations (R.S.D.%) of the assay were less than 7.1 and 12.3%, respectively. The accuracy determined at the concentrations of 3.0, 100.0, 500.0 and 1500 ng/ml for BCQB were within +/- 15.0%. The established method has been applied successfully to study the pharmacokinetics of BCQB in rats after intranasal administration. (c) 2006 Elsevier B.V All rights reserved.
引用
收藏
页码:209 / 214
页数:6
相关论文
共 6 条
[1]   Discovery & development of selective M(3) antagonists for clinical use [J].
Alabaster, VA .
LIFE SCIENCES, 1997, 60 (13-14) :1053-1060
[2]  
[Anonymous], 2001, GUID IND BIOAN METH
[3]   MUSCARINIC RECEPTOR SUBTYPES IN AIRWAYS [J].
BARNES, PJ .
LIFE SCIENCES, 1993, 52 (5-6) :521-527
[4]   Pharmacokinetics and tissue distribution of the anticholinergics tiotropium and ipratropium in the rat and dog [J].
Leusch, A ;
Eichhorn, B ;
Müller, G ;
Rominger, KL .
BIOPHARMACEUTICS & DRUG DISPOSITION, 2001, 22 (05) :199-212
[5]  
LI TW, 2004, Patent No. 1532196A
[6]   Stability of baicalin in biological fluids in vitro [J].
Xing, J ;
Chen, XY ;
Zhong, DF .