A portable multi-wavelength near infrared photon time-of-flight instrument for measuring light scattering

被引:1
作者
Esmonde-White, Francis W. L. [1 ]
Burns, David H. [1 ]
机构
[1] McGill Univ, Montreal, PQ H3A 2K6, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
diffuse reflectance; near infrared spectroscopy; photon time-of-flight; pulsed laser diode; scattering; turbid media; CONSTITUENT MEASUREMENTS; TISSUE; WATER; SPECTROSCOPY; HEMOGLOBIN; MEDIA;
D O I
10.1255/jnirs.847
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
Measured Light scattering properties can be used to improve quantitative spectroscopic analyses of turbid samples. Instruments currently used to measure scattering coefficients are not optimised for portability. A hand-held, dual-wavelength instrument was developed and validated for rapid measurement of reduced scattering coefficients in tandem with near infrared spectra. Tissue simulating phantoms composed of Intralipid and dye were used to model clinically relevant optical properties. Time-dependent intensity profiles of diffusely reflected near infrared pulsed laser tight were collected from phantoms and processed to estimate scattering coefficients. In turbid solutions, optical scattering was measured at 850 nm and 905 nm with coefficients of variation of 14.1% and 11.6% over a clinically-relevant reduced scattering coefficient range of 1 mm(-1) to 6 mm(-1). This dual-wavelength scattering measurement provides a practical method for measuring optical scattering. A 35% precision improvement in quantification of an absorbing dye is shown by incorporating the measured reduced scattering coefficients when processing NIR spectra. We discuss the new instrument, methods for estimating the scattering coefficient from the measured temporal profiles and, finally, how the reduced scattering coefficient is used to correct NIR measurements. Correction of near infrared spectra using optical scattering measurements offers one direction for improving practical non-invasive biomedical quantification techniques.
引用
收藏
页码:167 / 176
页数:10
相关论文
共 30 条
[1]   Myoglobin oxygen saturation measured independently of hemoglobin in scattering media by optical reflectance spectroscopy [J].
Arakaki, LSL ;
Kushmerick, MJ ;
Burns, DH .
APPLIED SPECTROSCOPY, 1996, 50 (06) :697-707
[2]   DIRECT CALCULATION OF THE MOMENTS OF THE DISTRIBUTION OF PHOTON TIME-OF-FLIGHT IN TISSUE WITH A FINITE-ELEMENT METHOD [J].
ARRIDGE, SR ;
SCHWEIGER, M .
APPLIED OPTICS, 1995, 34 (15) :2683-2687
[3]   Development and validation of a method for the analysis of a pharmaceutical preparation by near-infrared diffuse reflectance spectroscopy [J].
Blanco, M ;
Coello, J ;
Eustaquio, A ;
Iturriaga, H ;
Maspoch, S .
JOURNAL OF PHARMACEUTICAL SCIENCES, 1999, 88 (05) :551-556
[4]   Spectroscopy enhances the information content of optical mammography [J].
Cerussi, AE ;
Jakubowski, D ;
Shah, N ;
Bevilacqua, F ;
Lanning, R ;
Berger, AJ ;
Hsiang, D ;
Butler, J ;
Holcombe, RF ;
Tromberg, BJ .
JOURNAL OF BIOMEDICAL OPTICS, 2002, 7 (01) :60-71
[5]   Least-squares support vector machines modelization for time-resolved spectroscopy [J].
Chauchard, F ;
Roussel, S ;
Roger, JM ;
Bellon-Maurel, V ;
Abrahamsson, C ;
Svensson, T ;
Andersson-Engels, S ;
Svanberg, S .
APPLIED OPTICS, 2005, 44 (33) :7091-7097
[6]  
Cope M., 1991, TIME RESOLVED SPECTR, P251
[7]   Quantification in tissue near-infrared spectroscopy [J].
Delpy, DT ;
Cope, M .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY B-BIOLOGICAL SCIENCES, 1997, 352 (1354) :649-659
[8]   ESTIMATION OF OPTICAL PATHLENGTH THROUGH TISSUE FROM DIRECT TIME OF FLIGHT MEASUREMENT [J].
DELPY, DT ;
COPE, M ;
VANDERZEE, P ;
ARRIDGE, S ;
WRAY, S ;
WYATT, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (12) :1433-1442
[9]   Determining water content in human nails with a portable near-infrared spectrometer [J].
Egawa, M ;
Fukuhara, T ;
Takahashi, M ;
Ozaki, Y .
APPLIED SPECTROSCOPY, 2003, 57 (04) :473-478
[10]   Oxygen saturation-dependent absorption and scattering of blood [J].
Faber, DJ ;
Aalders, MCG ;
Mik, EG ;
Hooper, BA ;
van Gemert, MJC ;
van Leeuwen, TG .
PHYSICAL REVIEW LETTERS, 2004, 93 (02) :028102-1