Non-contact luminescence lifetime cryothermometry for macromolecular crystallography

被引:20
作者
Mykhaylyk, V. B. [1 ]
Wagner, A. [1 ]
Kraus, H. [2 ]
机构
[1] Diamond Light Source, Harwell Campus, Didcot OX11 0DE, Oxon, England
[2] Univ Oxford, Dept Phys, Denys Wilkinson Bldg,Keble Rd, Oxford OX1 3RH, England
关键词
cryogenic thermometry; luminescence lifetime; protein crystalography; bismuth germanate (BGO); THERMOGRAPHIC PHOSPHORS; SCINTILLATION MATERIALS; FLUORESCENCE LIFETIME; THERMAL-CONDUCTIVITY; TEMPERATURE SENSOR; THERMOMETRY;
D O I
10.1107/S1600577517003484
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Temperature is a very important parameter when aiming to minimize radiation damage to biological samples during experiments that utilize intense ionizing radiation. A novel technique for remote, non-contact, in situ monitoring of the protein crystal temperature has been developed for the new I23 beamline at the Diamond Light Source, a facility dedicated to macromolecular crystallography (MX) with long-wavelength X-rays. The temperature is derived from the temperature-dependent decay time constant of luminescence from a minuscule scintillation sensor (<0.05mm(3)) located in very close proximity to the sample under test. In this work the underlying principle of cryogenic luminescence lifetime thermometry is presented, the features of the detection method and the choice of temperature sensor are discussed, and it is demonstrated how the temperature monitoring system was integrated within the viewing system of the endstation used for the visualization of protein crystals. The thermometry system was characterized using a Bi4Ge3O12 crystal scintillator that exhibits good responsivity of the decay time constant as a function of temperature over a wide range (8-270K). The scintillation sensor was calibrated and the uncertainty of the temperature measurements over the primary operation temperature range of the beamline (30-150K) was assessed to be +/- 1.6K. It has been shown that the temperature of the sample holder, measured using the luminescence sensor, agrees well with the expected value. The technique was applied to characterize the thermal performance of different sample mounts that have been used in MX experiments at the I23 beamline. The thickness of the mount is shown to have the greatest impact upon the temperature distribution across the sample mount. Altogether, these tests and findings demonstrate the usefulness of the thermometry system in highlighting the challenges that remain to be addressed for the in-vacuum MX experiment to become a reliable and indispensable tool for structural biology.
引用
收藏
页码:636 / 645
页数:10
相关论文
共 36 条
[1]   Improved measurement precision in decay time-based phosphor thermometry [J].
Abou Nada, F. ;
Knappe, C. ;
Alden, M. ;
Richter, M. .
APPLIED PHYSICS B-LASERS AND OPTICS, 2016, 122 (06)
[2]   Remote thermometry with thermographic phosphors: Instrumentation and applications [J].
Allison, SW ;
Gillies, GT .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (07) :2615-2650
[3]  
Balandin AA, 2011, NAT MATER, V10, P569, DOI [10.1038/nmat3064, 10.1038/NMAT3064]
[4]   Thermal conductivity of Kapton tape [J].
Benford, DJ ;
Powers, TJ ;
Moseley, SH .
CRYOGENICS, 1999, 39 (01) :93-95
[5]   Optical fibre temperature sensor in the cryogenic range [J].
Bertrand, S ;
Jalocha, A ;
Tribillon, G ;
Bouazaoui, M ;
Rouhet, J .
OPTICS AND LASER TECHNOLOGY, 1996, 28 (05) :363-366
[6]   Fluorescence temperature sensing on rotating samples in the cryogenic range [J].
Bresson, F ;
Devillers, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1999, 70 (07) :3046-3051
[7]   On surface temperature measurements with thermographic phosphors: A review [J].
Bruebach, Jan ;
Pflitsch, Christian ;
Dreizler, Andreas ;
Atakan, Burak .
PROGRESS IN ENERGY AND COMBUSTION SCIENCE, 2013, 39 (01) :37-60
[8]   Phosphor thermometry at cryogenic temperatures [J].
Cates, MR ;
Beshears, DL ;
Allison, SW ;
Simmons, GM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (06) :2412-2417
[9]   Doped Oxides for High-Temperature Luminescence and Lifetime Thermometry [J].
Chambers, M. D. ;
Clarke, D. R. .
ANNUAL REVIEW OF MATERIALS RESEARCH, 2009, 39 :325-359
[10]   Temperature measurements during high flux ion beam irradiations [J].
Crespillo, M. L. ;
Graham, J. T. ;
Zhang, Y. ;
Weber, W. J. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (02)