An EPID based method for efficient and precise asymmetric jaw alignment quality assurance

被引:16
作者
Clews, Luke [1 ,2 ]
Greer, Peter B. [1 ,3 ]
机构
[1] Calvary Mater Newcastle Hosp, Newcastle, NSW 2298, Australia
[2] RMIT Univ, Melbourne, Vic 3000, Australia
[3] Univ Newcastle, Newcastle, NSW 2308, Australia
关键词
asymmetric jaws; junction dose verification; EPID; PORTAL IMAGING DEVICE; MULTILEAF COLLIMATOR; DOSIMETRY; FIELDS; HEAD; NECK; CALIBRATION; ACCELERATOR; ACCURACY; THERAPY;
D O I
10.1118/1.3253463
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this work was to investigate the use of amorphous silicon electronic portal imaging devices (EPIDs) for regular quality assurance of linear accelerator asymmetric jaw junctioning. Methods: The method uses the beam central axis position on the EPID measured to subpixel accuracy found from two EPID images with 180 opposing collimator angles. Individual zero jaw position ("half- beam blocked") images are then acquired and the jaw position precisely determined for each using penumbra interpolation. The accuracy of determining jaw position with the EPID method was measured by translating a block (simulating a jaw) by known distances, using a translation stage, and then measuring each translation distance with the EPID. To establish the utility of EPID based junction dose measurements, radiographic film measurements of junction dose maxima/minima as a function of jaw gap/overlap were made and compared to EPID measurements. Using the method, the long-term stability of zero jaw positioning was assessed for four linear accelerators over a 1-1.5 yr time period. The stability at nonzero gantry angles was assessed over a shorter time period. Results: The accuracy of determining jaw translations with the method was within 0.14 mm found using the translation stage [standard deviation (SD) of 0.037 mm]. The junction doses measured with the EPID were different from film due to the nonwater equivalent EPID scattering properties and hence different penumbra profile. The doses were approximately linear with gap or overlap, and a correction factor was derived to convert EPID measured junction dose to film measured equivalent. Over a 1 yr period, the zero jaw positions at gantry zero position were highly reproducible with an average SD of 0.07 mm for the 16 collimator jaws examined. However, the average jaw positions ranged from -0.7 to 0.9 mm relative to central axis for the different jaws. The zero jaw position was also reproducible at gantry 90 position with 0.1 mm SD variation with the mean jaw position offset from the gantry zero position consistently by 0.3-0.4 mm for the jaws studied. Conclusions: The EPID based method is efficient and yields more precise data on linear accelerator jaw positioning and reproducibility than previous methods. The results highlight that zero jaw positions are highly reproducible to a level much smaller than the displayed jaw resolution and that there is a need for better methods to calibrate the jaw positioning. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3253463]
引用
收藏
页码:5488 / 5496
页数:9
相关论文
共 28 条
[1]   RADIATION MYELOPATHY OF CERVICAL SPINAL-CORD - TIME, DOSE AND VOLUME FACTORS [J].
ABBATUCCI, JS ;
DELOZIER, T ;
QUINT, R ;
ROUSSEL, A ;
BRUNE, D .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1978, 4 (3-4) :239-248
[2]  
Abdel-Hakim K, 2003, STRAHLENTHER ONKOL, V179, P312, DOI 10.1007/s00066-003-1024-1
[3]   Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration [J].
Baker, SJK ;
Budgell, GJ ;
MacKay, RI .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) :1377-1392
[4]   Daily monitoring of linear accelerator beam parameters using an amorphous silicon EPID [J].
Budgell, G. J. ;
Zhang, R. ;
Mackay, R. I. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (06) :1721-1733
[5]   Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager [J].
Budgell, GJ ;
Zhang, Q ;
Trouncer, RJ ;
Mackay, RI .
MEDICAL PHYSICS, 2005, 32 (11) :3267-3278
[6]   Use of EPID for leaf position accuracy QA of dynamic multi-leaf collimator (DMLC) treatment [J].
Chang, J ;
Obcemea, CH ;
Sillanpaa, J ;
Mechalakos, J ;
Burman, C .
MEDICAL PHYSICS, 2004, 31 (07) :2091-2096
[7]   Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle [J].
Clarke, M. F. ;
Budgell, G. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (02) :473-485
[8]  
Curtin-Savard A, 1997, Med Dosim, V22, P101, DOI 10.1016/S0958-3947(97)00002-2
[9]  
DENITTIS AS, 2004, PRINCIPLES PRACTICE, P1282
[10]   Dosimetry of therapeutic photon beams using an extended dose range film [J].
Esthappan, J ;
Mutic, S ;
Harms, WB ;
Dempsey, JF ;
Low, DA .
MEDICAL PHYSICS, 2002, 29 (10) :2438-2445