A clinical comparison of bubble elimination in Quadrox and Polystan oxygenators

被引:9
作者
Jirschik, Mario [1 ]
Keyl, Cornelius [1 ]
Beyersdorf, Friedhelm [1 ]
机构
[1] Univ Herzzentrum Freiburg Bad Krozingen, Dept Cardiovasc Surg, D-79189 Bad Krozingen, Germany
来源
PERFUSION-UK | 2009年 / 24卷 / 06期
关键词
extracorporeal circulation; oxygenator; microbubble; gaseous microemboli; cardiopulmonary bypass; CORONARY-BYPASS SURGERY; PERFUSIONIST INTERVENTIONS; CARDIOPULMONARY BYPASS; MICROEMBOLI;
D O I
10.1177/0267659109358206
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Microbubbles generated during heart surgery on extracorporeal circulation have been implicated as a possible cause of postoperative neurocognitive dysfunction and negative outcome. The main sources of microbubbles in the extracorporeal circuit are air leaking from the venous cannulation site, air delivered by drug and volume administration, during the taking of blood samples, during hemofiltration, and by using vacuum-assisted venous drainage (VAVD). Membrane oxygenators, although not designed for the elimination of gaseous microbubbles, can eliminate much of this air. Aim of the study/hypothesis: The aim of this study was to test the hypothesis that the amount of microbubbles passing through the oxygenator varies depending on the oxygenator design. Methods: one hundred patients undergoing aortic valve replacement with cardiopulmonary bypass (CPB) were included in a retrospective clinical trial assessing the bubble elimination of the two oxygenators routinely used in our clinic. The oxygenators (Quadrox-iAdult/Polystan Safe Maxi) are manufactured with the same hollow-fiber material, but display different designs. Bubbles were detected by a two-channel ultrasound bubble counter. The probes were placed directly at the inlet and the outlet of the oxygenators. The filtration index was calculated by the bubble counter as a measure of the percentage reduction of bubble volume and number. Results: The Quadrox has a significantly higher filtration index (92.3% [ 89.7; 95.1], median [25(th), 75(th) percentile]) than the Polystan (74.9% [64.2; 80.9], p<0.001), indicating the superior ability of the Quadrox to eliminate microbubbles. Conclusion: Our study demonstrates major differences in the capacity of bubble elimination of the oxygenators being investigated, possibly depending on oxygenator design.
引用
收藏
页码:423 / 427
页数:5
相关论文
共 9 条
[1]   Microbubbles - Pathophysiology and clinical implications [J].
Barak, M ;
Katz, Y .
CHEST, 2005, 128 (04) :2918-2932
[2]   A comparison of gaseous emboli release in five membrane oxygenators [J].
Beckley, Philip D. ;
Shinko, Paul D. ;
Sites, Jeffrey P. .
PERFUSION-UK, 1997, 12 (02) :133-141
[3]   Stroke during coronary bypass surgery: principal role of cerebral macroemboli [J].
Borger, MA ;
Ivanov, J ;
Weisel, RD ;
Rao, V ;
Peniston, CM .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 19 (05) :627-632
[4]   Neuropsychologic impairment after coronary bypass surgery: Effect of gaseous microemboli during perfusionist interventions [J].
Borger, MA ;
Peniston, CM ;
Weisel, RD ;
Vasiliou, M ;
Green, REA ;
Feindel, CM .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 121 (04) :743-749
[5]  
De Somer Filip, 2007, J Extra Corpor Technol, V39, P271
[6]   Microemboli detection on extracorporeal bypass circuits [J].
Lynch, J. E. ;
Riley, J. B. .
PERFUSION-UK, 2008, 23 (01) :23-32
[7]   Effect of perfusionist technique on cerebral embolization during cardiopulmonary bypass [J].
Rodriguez, RA ;
Williams, KA ;
Babaev, A ;
Rubens, F ;
Nathan, HJ .
PERFUSION-UK, 2005, 20 (01) :3-10
[8]   Cerebral microemboli during cardiopulmonary bypass: Increased emboli during perfusionist interventions [J].
Taylor, RL ;
Borger, MA ;
Weisel, RD ;
Fedorko, L ;
Feindel, CM .
ANNALS OF THORACIC SURGERY, 1999, 68 (01) :89-93
[9]  
Weitkemper Heinz-H, 2005, J Extra Corpor Technol, V37, P256