Are 3D Printing Templates an Advantage in Upper Thoracic Pedicle Screw Fixation?

被引:5
作者
Kaya, Ismail [1 ]
Cingoz, Ilker Deniz [1 ]
Sahin, Meryem Cansu [2 ]
Atar, Murat [3 ]
Ozyoruk, Safak [4 ]
Sayin, Murat [5 ]
Yuceer, Nurullah [6 ]
机构
[1] Usak Univ, Neurosurg, Usak, Turkey
[2] Kutahya Hlth Sci Univ, Med Phys, Kutahya, Turkey
[3] Abdulhamid Han Res & Training Hosp, Neurosurg, Istanbul, Turkey
[4] Private OEM Antalya Hosp, Neurosurg, Antalya, Turkey
[5] Private Saglik Hosp, Neurosurg, Izmir, Turkey
[6] Izmir Katip Celebi Univ, Neurosurg, Izmir, Turkey
关键词
3d printing; pedicle screw; preoperative plan; upper thoracic trauma; FREE-HAND; PLACEMENT; SPINE; ACCURACY; LUMBAR;
D O I
10.7759/cureus.13989
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background This study aims to compare the clinical results of patients with upper thoracic vertebral fractures treated with pedicle screw and posterior spinal fusion with preoperative surgical planning and 3-dimensional (3D) modeling and patients treated with freehand screws. Methods Fifty patients who underwent pedicle screw placement with a diagnosis of upper thoracic fracture between June 2018 and October 2020 were included in our study. Pedicle screws were used in 25 patients (group 1) after the planning was completed with the help of 3D preoperative printing and modeling. Pedicle screws were applied in 25 patients in the control group (group 2) using the freehand technique. Intraoperative bleeding amount, operation time, and correct screw placement data in both groups were recorded. Results The operation time was 134 +/- 22 minutes for group 1 and 152 +/- 38 minutes for group 2. The difference in operation times was found to be statistically significant (p < 0.05). Based on axial and sagittal reconstruction images, the accuracy rate of pedicle screw placement (grades 0 and 1) in group I was 96.6% compared to 83.6% in group II. The minor perforation rate (grade 1, <2 mm) was 5.8% in group I compared to 11.8% in group II. The moderate perforation rate (grade 2, 2-4 mm) was 3.4% in group I compared to 14% in group II. The severe perforation rate (grade 3, >4 mm) was 2.3% in group 11; however, misplaced screws were not associated with neurological deficits. The difference in overall accuracy rates between the two groups was significant (p < 0.05). Conclusions For 3D models of upper thoracic pellicle screw insertion, guide plates can be produced inexpensively and individually. It provides a new method for the accurate placement of upper thoracic pellicle screws with high accuracy and secure use in screw insertion.
引用
收藏
页数:11
相关论文
共 25 条
  • [1] Pedicle screw fixation for nontraumatic lesions of the cervical spine
    Abumi, K
    Kaneda, K
    [J]. SPINE, 1997, 22 (16) : 1853 - 1863
  • [2] Upper Thoracic Spine Fracture Dislocation in a Motorcyclist
    Anthes, Tara B.
    Muangman, Nisa
    Bulger, Eileen
    Stern, Eric J.
    [J]. CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2012, 41 (04) : 128 - 129
  • [3] Thoracic Pedicle Screw Placement Guide Plate Produced by Three-Dimensional (3-D) Laser Printing
    Chen, Hongliang
    Guo, Kaijing
    Yang, Huilin
    Wu, Dongying
    Yuan, Feng
    [J]. MEDICAL SCIENCE MONITOR, 2016, 22 : 1682 - 1686
  • [4] Investigation and Feasibility of Combined 3D Printed Thermoplastic Filament and Polymeric Foam to Simulate the Cortiocancellous Interface of Human Vertebrae
    Clifton, William
    Pichelmann, Mark
    Vlasak, Alexander
    Damon, Aaron
    ReFaey, Karim
    Nottmeier, Eric
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [5] Complications of thoracic pedicle screws in scoliosis treatment
    Di Silvestre, Mario
    Parisini, Patrizio
    Lolli, Francesco
    Bakaloudis, Georgios
    [J]. SPINE, 2007, 32 (15) : 1655 - 1661
  • [6] Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniques
    Gelalis, Ioannis D.
    Paschos, Nikolaos K.
    Pakos, Emilios E.
    Politis, Angelos N.
    Arnaoutoglou, Christina M.
    Karageorgos, Athanasios C.
    Ploumis, Avraam
    Xenakis, Theodoros A.
    [J]. EUROPEAN SPINE JOURNAL, 2012, 21 (02) : 247 - 255
  • [7] ACCURACY OF PEDICULAR SCREW PLACEMENT INVIVO
    GERTZBEIN, SD
    ROBBINS, SE
    [J]. SPINE, 1990, 15 (01) : 11 - 14
  • [8] Accuracy of pedicle screw placement for upper and middle thoracic pathologies without coronal plane spinal deformity using conventional methods
    Guzey, Feyza Karagoz
    Emel, Erhan
    Seyithanoglu, M. Hakan
    Bas, N. Serdar
    Ozkan, Nezih
    Sel, Baris
    Aycan, Abdurrahman
    Alatas, Ibrahim
    [J]. JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (06): : 436 - 441
  • [9] HANLEY EN, 1989, ORTHOPEDICS, V12, P689
  • [10] Free hand pedicle screw placement in the thoracic spine: Is it safe?
    Kim, YJ
    Lenke, LG
    Bridwell, KH
    Cho, YSS
    Riew, KD
    [J]. SPINE, 2004, 29 (03) : 333 - 342