Sensitive and ultra-fast determination of arsenic(III) by gas-diffusion flow injection analysis with chemiluminescence detection

被引:28
作者
Lomonte, Cristina
Currell, Matthew
Morrison, Richard J. S.
McKelvie, Ian D.
Kolev, Spas D. [1 ]
机构
[1] Univ Melbourne, Sch Chem, Melbourne, Vic 3010, Australia
[2] Monash Univ, Sch Chem, Clayton, Vic 3800, Australia
关键词
gas-diffusion flow injection analysis; chemiluminescence detection; arsenic determination; arsine; ozone;
D O I
10.1016/j.aca.2006.09.049
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel chemiluminescence gas-diffusion flow injection system for the determination of arsenic(Ill) in aqueous samples is described. The analytical procedure involves injection of arsenic(III) samples and standards into a 0.3 mol L-1 hydrochloric acid carrier stream which is merged with a reagent stream containing 0.2% (w/v) sodium borohydride and 0.015 mol L-1 sodium hydroxide. Arsine, generated in the combined carrier/reagent donor stream, diffuses across the hydrophobic Teflon membrane of the gas-diffusion cell into an argon acceptor stream and then reacts with ozone in the flow-through chemiluminescence measuring cell of the flow system. Under optimal conditions, the method is characterized by a wide linear calibration range from 0.6 mu g L-1 to 25 mg L-1, a detection limit of 0.6 mu g L-1 and a sample throughput of 300 samples per hour at 25 mg L-1 and 450 samples per hour at 25 p g L-1. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:72 / 77
页数:6
相关论文
共 26 条
[1]  
Anderson KA, 1995, J AOAC INT, V78, P1055
[2]  
[Anonymous], ENV HLTH CRIT
[3]  
[Anonymous], 1974, Fundamental Mechanics of Fluids
[4]   Slurry sampling-microwave assisted leaching prior to hydride generation-pervaporation-atomic fluorescence detection for the determination of extractable arsenic in soil [J].
Caballo-López, A ;
de Castro, MDL .
ANALYTICAL CHEMISTRY, 2003, 75 (09) :2011-2017
[5]   Hydride generation-pervaporation-atomic fluorescence detection prior to speciation analysis of arsenic in dirty samples [J].
Caballo-López, A ;
de Castro, MDL .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (10) :1363-1367
[6]   Determination of arsenic by hydride generation gas diffusion flow injection analysis with electrochemical detection [J].
Farrell, JR ;
Iles, PJ ;
Yuan, YJ .
ANALYTICA CHIMICA ACTA, 1996, 334 (1-2) :193-197
[8]   SPECTROSCOPY AND MECHANISM OF CHEMI-LUMINESCENT REACTIONS BETWEEN GROUP-V HYDRIDES AND OZONE [J].
FRASER, ME ;
STEDMAN, DH .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS I, 1983, 79 :527-542
[9]   GAS-PHASE CHEMI-LUMINESCENCE OF ARSINE MIXED WITH OZONE [J].
FRASER, ME ;
STEDMAN, DH ;
HENDERSON, MJ .
ANALYTICAL CHEMISTRY, 1982, 54 (07) :1200-1201
[10]   SELECTIVE DETERMINATION OF ARSENITE BY FLOW-INJECTION SPECTROPHOTOMETRY [J].
FRENZEL, W ;
TITZENTHALER, F ;
ELBEL, S .
TALANTA, 1994, 41 (11) :1965-1971